Acoustic features of auditory medical alarms-An experimental study of alarm volume.

Joseph J Schlesinger1, Sarah H Baum Miller2, Katherine Nash3, Marissa Bruce4, Daniel Ashmead5, Matthew S Shotwell6, Judy R Edworthy7, Mark T Wallace8, Matthew B Weinger9.   

Abstract

Audible alarms are a ubiquitous feature of all high-paced, high-risk domains such as aviation and nuclear power where operators control complex systems. In such settings, a missed alarm can have disastrous consequences. It is conventional wisdom that for alarms to be heard, "louder is better," so that alarm levels in operational environments routinely exceed ambient noise levels. Through a robust experimental paradigm in an anechoic environment to study human response to audible alerting stimuli in a cognitively demanding setting, akin to high-tempo and high-risk domains, clinician participants responded to patient crises while concurrently completing an auditory speech intelligibility and visual vigilance distracting task as the level of alarms were varied as a signal-to-noise ratio above and below hospital background noise. There was little difference in performance on the primary task when the alarm sound was -11 dB below background noise as compared with +4 dB above background noise-a typical real-world situation. Concurrent presentation of the secondary auditory speech intelligibility task significantly degraded performance. Operator performance can be maintained with alarms that are softer than background noise. These findings have widespread implications for the design and implementation of alarms across all high-consequence settings.

Entities:  

Mesh:

Year:  2018        PMID: 29960450      PMCID: PMC6910025          DOI: 10.1121/1.5043396

Source DB:  PubMed          Journal:  J Acoust Soc Am        ISSN: 0001-4966            Impact factor:   1.840


  49 in total

1.  Psychometric properties of the coordinate response measure corpus with various types of background interference.

Authors:  David A Eddins; Chang Liu
Journal:  J Acoust Soc Am       Date:  2012-02       Impact factor: 1.840

Review 2.  Risk of burnout in perioperative clinicians: a survey study and literature review.

Authors:  Steve A Hyman; Damon R Michaels; James M Berry; Jonathan S Schildcrout; Nathaniel D Mercaldo; Matthew B Weinger
Journal:  Anesthesiology       Date:  2011-01       Impact factor: 7.892

3.  Auditory filter shapes derived with noise stimuli.

Authors:  R D Patterson
Journal:  J Acoust Soc Am       Date:  1976-03       Impact factor: 1.840

4.  The effects of noise on man.

Authors:  K D KRYTER
Journal:  J Speech Hear Disord Monogr Suppl       Date:  1950-09

5.  Alarm fatigue: a patient safety concern.

Authors:  Sue Sendelbach; Marjorie Funk
Journal:  AACN Adv Crit Care       Date:  2013 Oct-Dec

6.  Benefits of quiet time for neuro-intensive care patients.

Authors:  Christina M Dennis; Robert Lee; Elizabeth Knowles Woodard; Jeffery J Szalaj; Catrice A Walker
Journal:  J Neurosci Nurs       Date:  2010-08       Impact factor: 1.230

7.  Prevalence and risk factors for development of delirium in surgical and trauma intensive care unit patients.

Authors:  Pratik Pandharipande; Bryan A Cotton; Ayumi Shintani; Jennifer Thompson; Brenda Truman Pun; John A Morris; Robert Dittus; E Wesley Ely
Journal:  J Trauma       Date:  2008-07

Review 8.  Identifying clinical and acute psychological risk factors for PTSD after critical care: a systematic review.

Authors:  D Wade; R Hardy; D Howell; M Mythen
Journal:  Minerva Anestesiol       Date:  2013-04-05       Impact factor: 3.051

Review 9.  Systematic Review of Physiologic Monitor Alarm Characteristics and Pragmatic Interventions to Reduce Alarm Frequency.

Authors:  Christine Weirich Paine; Veena V Goel; Elizabeth Ely; Christopher D Stave; Shannon Stemler; Miriam Zander; Christopher P Bonafide
Journal:  J Hosp Med       Date:  2015-12-14       Impact factor: 2.960

10.  The association between noise exposure and blood pressure and ischemic heart disease: a meta-analysis.

Authors:  Elise E M M van Kempen; Hanneke Kruize; Hendriek C Boshuizen; Caroline B Ameling; Brigit A M Staatsen; Augustinus E M de Hollander
Journal:  Environ Health Perspect       Date:  2002-03       Impact factor: 9.031

View more
  2 in total

1.  Letter to the editor.

Authors:  Christopher P Siemer; Tulasi M Iyengar; Christy J Crockett; Joseph J Schlesinger
Journal:  Langenbecks Arch Surg       Date:  2022-07-26       Impact factor: 2.895

2.  Improved Patient Monitoring with a Novel Multisensory Smartwatch Application.

Authors:  Kendall J Burdick; Mohh Gupta; Ayush Sangari; Joseph J Schlesinger
Journal:  J Med Syst       Date:  2022-10-20       Impact factor: 4.920

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.