Literature DB >> 29952252

Stent fractures after common femoral artery bail-out stenting due to suture device failure in TAVR.

Verena Veulemans1,2, Shazia Afzal1,2, Paul Ledwig1, Christian Heiss1, Lucas Busch1, Roberto Sansone1, Dagmar B Soetemann1, Oliver Maier1, Laura Kleinebrecht1, Malte Kelm1,3, Tobias Zeus1, Katharina Hellhammer1.   

Abstract

BACKGROUND: Vascular access site-related complications are frequent in the context of transfemoral transcatheter aortic valve replacement (TAVR). The implantation of a covered stent graft is an effective treatment option for bleeding control. However, the external iliac and common femoral arteries are exposed to flexion of the hip joint. Therefore, stent compression and stent/strut fractures may occur, facilitating stent occlusion. PATIENTS AND METHODS: In all 389 patients who received transfemoral TAVR from 2013-2015 at the Düsseldorf Heart Centre, we monitored the management of vascular access site-related complications. Our analyses focused on immediate technical success and bleeding control, primary patency, and the occurrence of stent/strut fractures after six to 12 months of follow-up.
RESULTS: Vascular access site-related complications occurred in 13 % (n = 51), whereof in 10 patients, the bleeding was successfully managed by prolonged compression. In 40 out of 51 patients, a covered stent graft was implanted in the common femoral artery, leading to 100 % immediate bleeding control. After a mean follow-up of 334 ± 188 days, 28 stents out of 29 patients with completed follow-up (excluding e. g. death) were without flow-limiting stenosis (primary patency 97 %) or relevant stent compression (diameter pre/post 8.6/8.1 mm, p = 0.048, late lumen loss 1.1 ± 0.2 mm, mean flow velocity 92 ± 34 cm/s). In four asymptomatic patients, stent/strut fractures were detected (14 %) without flow-limiting stenosis.
CONCLUSIONS: The implantation of a covered stent graft is highly effective and safe to control vascular access site-related complications after TAVR. Stent/strut fractures in the flexible segment of the common femoral artery may occur, as consequently verified by X-ray visualization, but show no impairment on flow or clinical parameters after six to 12 months.

Entities:  

Keywords:  Percutaneous valve; interventional cardiology; risk factors

Mesh:

Year:  2018        PMID: 29952252     DOI: 10.1024/0301-1526/a000712

Source DB:  PubMed          Journal:  Vasa        ISSN: 0301-1526            Impact factor:   1.961


  1 in total

1.  Endovascular snare retrieval of an Angio-Seal causing acute limb ischemia.

Authors:  Lucas Busch; Manuel Stern; Georg Wolff; Malte Kelm
Journal:  Clin Case Rep       Date:  2021-12-05
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.