Ventilation Parameters under Adaptive Servo Ventilation: A Comparison on Behalf of CSA-Pattern, BNP/NT-pro-BNP, and Ejection Fraction.

Michael Westhoff1,2, Patric Litterst1.   

Abstract

BACKGROUND: The SERVE-HF study has raised questions concerning the higher mortality under adaptive servoventilation. The ventilatory mode was discussed as a possible aggravating factor.
OBJECTIVES: We wondered if the data recorded by the adaptive servo-ventilation (ASV)-devices in heart failure patients with CSA-CSR ± OSA are different in terms of respiratory parameters and therapeutic pressures compared to patients with CPAP-resistant/emergent-CSA with normal BNP/NT-pro-BNP.
METHODS: Patients were included, if ASV had normalized respiratory disturbance index in the first night of application and after at least 6 weeks. ASV-device data were analyzed in terms of respiratory rate (RR), min ventilation (MV), endexpiratory (EEP), peak inspiratory pressure (Ppeak) and median pressure.
RESULTS: Compared to CPAP-resistant/emergent-CSA with normal BNP/NT-pro-BNP (n = 25), CSA-CSR- (n = 13) CSA-CSR+OSA-patients (n = 32) with elevated BNP/NT-pro-BNP had higher RR (p < 0.01) in the first night of ASV therapy and during follow-up (15.3 ± 1.3 vs. 17.3 ± 2.4/min) with similar MV (6.5 ± 1.3 vs. 6.6 ± 1.3 L), resulting in significantly lower tidal volumes. EEP (5.6 ± 1.1 vs. 5.5 ± 1.1 hPa), Pmedian and Ppeak (9.8 ± 1.5 vs. 9.7 ± 1.2 hPa) were comparable. Ventilatory parameters were not different between LVEF < 40, 40-49, and ≥50%, neither within the whole group nor the group of CSA-CSR ± OSA and heart failure.
CONCLUSION: Patients with heart failure and CSA-CSR ± OSA have higher RRs but similar MV under ASV-therapy than patients with CSA and normal BNP. This indicates higher dead space ventilation. EF was not found to have an influence on the ventilatory parameters.
© 2018 S. Karger AG, Basel.

Entities:  

Keywords:  Adaptive servoventilation; Central sleep apnoea; Cheyne-stokes-respiration; Ejection fraction; Therapeutic pressures; Ventilatory parameters

Mesh:

Substances:

Year:  2018        PMID: 29940592     DOI: 10.1159/000489228

Source DB:  PubMed          Journal:  Respiration        ISSN: 0025-7931            Impact factor:   3.580


  2 in total

1.  Changes of blood gas and serum indexes in patients with acute heart failure complicated with respiratory failure treated by noninvasive ventilator.

Authors:  Yuan Wu; Xin Hong; Ni Yang; Ru Zhang; Yu Shi
Journal:  Pak J Med Sci       Date:  2022 Sep-Oct       Impact factor: 2.340

2.  Patterns of adaptive servo-ventilation settings in a real-life multicenter study: pay attention to volume! : Adaptive servo-ventilation settings in real-life conditions.

Authors:  Dany Jaffuel; Claudio Rabec; Carole Philippe; Jean-Pierre Mallet; Marjolaine Georges; Stefania Redolfi; Alain Palot; Carey M Suehs; Erika Nogue; Nicolas Molinari; Arnaud Bourdin
Journal:  Respir Res       Date:  2020-09-21
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.