An investigation into the INTRABEAM miniature x-ray source dosimetry using ionization chamber and radiochromic film measurements.

Peter G F Watson1, Hamed Bekerat2, Pavlos Papaconstadopoulos2, Stephen Davis1, Jan Seuntjens1.   

Abstract

PURPOSE: Intraoperative radiotherapy using The INTRABEAM System (Carl Zeiss Meditec AG, Jena, Germany), a miniature low-energy x-ray source, has proven to be an effective modality in the treatment of breast cancer. However, some uncertainties remain in its dosimetry. In this work, we investigated the INTRABEAM system dosimetry by performing ionization chamber and radiochromic film measurements of absorbed dose in a water phantom.
METHODS: Ionization chamber measurements were performed with a PTW 34013 parallel-plate soft x-ray chamber at source to detector distances of 5 to 30 mm calculated using (a) the dose formula consistent with the TARGIT breast protocol (TARGIT), (b) the formula recommended by the manufacturer (Zeiss), and (c) the recently proposed CQ formalism of Watson et al. (Physics in Medicine & Biology, 2018;63:015016) EBT3 Gafchromic film measurements were made at the same depths in water. To account for the energy dependence of EBT3 film, multiple dose response calibration curves were employed across a range of photon beam qualities relevant to the INTRABEAM spectrum in water.
RESULTS: At all depths investigated, the TARGIT dose was significantly lower than that measured by the Zeiss and CQ methods, as well as film. These dose differences ranged from 14% to as large as 80%. In general, the doses measured by film, and the Zeiss and CQ methods were in good agreement to within measurement uncertainties (5-6%).
CONCLUSIONS: These results suggest that the TARGIT dose underestimates the physical dose to water from the INTRABEAM source. Understanding the correlation between the TARGIT and physical dose is important for any studies wishing to make dosimetric comparisons between the INTRABEAM and other radiation emitting devices.
© 2018 American Association of Physicists in Medicine.

Entities:  

Keywords:  electronic brachytherapy; gafchromic film; kV dosimetry

Year:  2018        PMID: 29935088     DOI: 10.1002/mp.13059

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Dosimetric characteristics of the INTRABEAM ® system with spherical applicators in the presence of air gaps and tissue heterogeneities.

Authors:  Eyachew Misganew Tegaw; Somayeh Gholami; Gilnaz Omyan; Ghazale Geraily
Journal:  Radiat Environ Biophys       Date:  2020-03-31       Impact factor: 1.925

2.  Determination of paramagnetic ferrous gel sensitivity in low energy x-ray beam produced by a miniature accelerator.

Authors:  Yassir Ben Ahmed; Jérémy Coulaud; Soléakhena Ken; Laure Parent
Journal:  PLoS One       Date:  2020-05-04       Impact factor: 3.240

3.  Dosimetric evaluation of incorporating the revised V4.0 calibration protocol for breast intraoperative radiotherapy with the INTRABEAM system.

Authors:  Mubin Y Shaikh; Jay Burmeister; Robin Scott; Lalith K Kumaraswamy; Adrian Nalichowski; Michael C Joiner
Journal:  J Appl Clin Med Phys       Date:  2020-02-10       Impact factor: 2.102

4.  Toward 3D dose verification of an electronic brachytherapy source with a plastic scintillation detector.

Authors:  Peter Georgi; Gustavo Kertzscher; Lars Nyvang; Jaroslav Šolc; Thorsten Schneider; Kari Tanderup; Jacob Graversen Johansen
Journal:  Med Phys       Date:  2022-03-03       Impact factor: 4.506

5.  Evaluation of the dosimetric impact of manufacturing variations for the INTRABEAM x-ray source.

Authors:  Mubin Y Shaikh; Michael C Joiner; Adrian Nalichowski; Lalith K Kumaraswamy; Jay Burmeister
Journal:  J Appl Clin Med Phys       Date:  2020-01-24       Impact factor: 2.102

6.  First Intraoperative Radiation Therapy Center in Africa: First 2 Years in Operation, Including COVID-19 Experiences.

Authors:  Yastira Ramdas; Carol-Ann Benn; Michelle van Heerden
Journal:  JCO Glob Oncol       Date:  2020-11
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.