Radiation Exposure to Surgeon and Assistant During Flat Panel Mini C-Arm Fluoroscopy in Hand and Wrist Surgical Procedures.

Juliaan R M van Rappard1, Tijmen de Jong2, Willy A Hummel3, Marco J P F Ritt4, Chantal M Mouës2.   

Abstract

PURPOSE: During mini C-arm fluoroscopy, both the patient and surgical team are exposed to scatter radiation. The objective of this study was to quantify body, thyroid, and hand radiation exposure to surgeon and assistant during intraoperative use of flat panel mini C-arm fluoroscopy in hand and wrist surgical procedures.
METHODS: Over 5 months, the surgeon's and assistant's radiation exposure was recorded during all osseous hand and wrist surgical procedures. Whole-body and thyroid radiation exposure were measured with 2 types of dosimeters: a photon thermoluminescence detector and a RaySafe i2 real-time dosimeter. Ring dosimeters were used to quantify hand radiation exposure.
RESULTS: Mini C-arm fluoroscopy was used in 94 surgical procedures. Total fluoroscopy time was 1,996 seconds and varied between surgical procedures (range, 1-152 seconds; median, 11 seconds). No thermoluminescence detector photon dosimeter exceeded the threshold limit of 0.1 mSv. The RaySafe i2 real-time dosimeters recorded a cumulated dose of 0.029 mSv for the body and 0.012 mSv for the thyroid position of the surgeon. The assistant received a cumulated dose of 0.011 mSv for the body and 0.011 mSv for the thyroid position. The ring dosimeters showed a cumulated dosage of 1.28 mSv for the surgeon and 0.20 mSv for the assistant.
CONCLUSIONS: Our results show that the surgeon's and assistant's body, thyroid, and hands were exposed to acceptable levels of scatter radiation during intraoperative use of the flat panel mini C-arm. The surgeon received the highest radiation exposure: 2.9% of the yearly radiation limits for the body, 0.05% for the thyroid position, and 2.56% for the hands. The assistant was exposed to less scatter radiation: 1.1% for the body, 0.04% for the thyroid, and 0.4% for the hands. CLINICAL RELEVANCE: This study quantified radiation levels to which surgeon and assistant are exposed during mini C-arm fluoroscopy in hand and wrist surgical procedures.
Copyright © 2019 American Society for Surgery of the Hand. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Hand; mini C-arm; radiation; surgeon

Year:  2018        PMID: 29934087     DOI: 10.1016/j.jhsa.2018.05.010

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  4 in total

1.  Does nonexistent of your hands on the screen guarantee no radiation exposure to your body? - Study on exposure of the practitioner's hands to radiation during C-arm fluoroscopy-guided injections and effectiveness of a new shielding device.

Authors:  Jae-Eun Lee; Ju-Hyun Kim; Su-Jin Lee; Mingeon Kim; Tae-Woo Nam; Chul Hyun Kim; Tae-Du Jung; Jong-Moon Hwang
Journal:  Medicine (Baltimore)       Date:  2019-11       Impact factor: 1.817

2.  Confirming Proper Button Deployment of Suspensory Fixation During ACL Reconstruction.

Authors:  Daniel F O'Brien; Lilah Fones; Victoria Stoj; Cory Edgar; Katherine Coyner; Robert A Arciero
Journal:  Orthop J Sports Med       Date:  2021-01-19

3.  Artifacts in fluoroscopy and changes in radiation dose caused by heating blankets and insulating covers during simulated endovascular treatment.

Authors:  Paweł Podsiadło; Grzegorz Liszka; Tadeusz Popiela; Tomasz Sanak; Sylweriusz Kosiński; Tomasz Darocha
Journal:  Emerg Radiol       Date:  2020-05-30

4.  Results from a real-time dosimetry study during left atrial ablations performed with ultra-low dose radiation settings.

Authors:  T Schreiber; N Kähler; S Biewener; V Tscholl; P Nagel; P Attanasio; U Landmesser; M Huemer
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2021-05-11
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.