| Literature DB >> 29927027 |
Amitpal Singh Saini1,2, Catherine S Hwang2, Matthew C Biagioli2, Indra J Das3.
Abstract
PURPOSE: To compare doses to organs at risk (OARs) for left-sided whole-breast radiation therapy with comparable planning target volume (PTV) coverage using three techniques: free breathing in a supine position (SFB), deep inspirational breath-hold in a supine position (SDIBH), and free breathing in prone position (PFB).Entities:
Keywords: LAD dose; Lt breast radiation; cardiac toxicity; deep inspiration breath-hold; heart dose; left ventricle dose; lung dose; prone position; supine position
Mesh:
Year: 2018 PMID: 29927027 PMCID: PMC6036360 DOI: 10.1002/acm2.12382
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Table for recording SDIBH measurements
| Voluntary breath‐hold details |
|---|
| Max breath‐hold achieved (sec) |
| Anterior FB bb and BH bb, distance (mm) |
| Lateral FB bb and BH bb, distance (mm) |
Figure 1Clockwise, axial, beam's eye view, sagittal, and coronal images. Green line indicates PTV, red indicates 95% iodose line, orange indicates MLC field shape, and yellow indicates beam outline. Heart and lung volumes are also shown. (a) Supine free breathing (SFB), (b) supine deep inspirational breath‐hold (SDIBH) and (c) prone‐free breathing (PFB). Note the heart and lung positions in three techniques with respect to beam geometry.
Figure 2Box‐whisker plot of dose with error bars in supine free breathing (SFB), supine deep inspiration breath‐hold (SDIBH), and prone free‐breathing (PFB) techniques. Outlier data are shown if they existed for (a) heart, (b) left ventricle (LV), (c) left anterior descending artery (LAD), and (d) lung.
Dosimetry parameters (median values and quartiles) in supine free breathing (SFB), supine deep inspiration breath‐hold (SDIBH), and prone free‐breathing (PFB) techniques. (a) Heart, (b) left ventricle (LV), (c) left anterior descending artery (LAD), and (d) left lung
| SFB | SDIBH | PFB | |
|---|---|---|---|
| (a) Heart dose (Gy) | |||
| Mean dose | 1.92 (1.42–2.76) | 1.08 (0.84–1.36) | 0.98 (0.83–1.15) |
| V2.5 | 14.60 (9.27–22.34) | 7.20 (3.98–11.46) | 6.30 (4.47–8.89) |
| V5 | 4.81 (2.73–7.35) | 0.90 (0.22–1.92) | 0.80 (0.275–1.61) |
| V10 | 2.66 (1.43–4.58) | 0.11 (0.00–0.86) | 0.10 (0.01–0.47) |
| V20 | 1.74 (0.79–3.21) | 0.00 (0.00–0.20) | 0.00 (0.00–0.12) |
| V30 | 1.15 (0.23–2.34) | 0.00 (0.00–0.17) | 0.00 (0.00–0.00) |
| (b) Left ventricle dose (Gy) | |||
| Mean dose | 3.19 (2.25–4.24) | 1.50 (1.15–1.80) | 1.34 (1.13–1.54) |
| V2.5 | 30.00 (18.84–39.00) | 11.54 (7.46–19.08) | 8.92 (6.01–12) |
| V5 | 9.23 (5.24–14.23) | 1.05 (0.38–3.29) | 0.92 (0.43–1.88) |
| V10 | 5.01 (2.34–9.65) | 0.00 (0.00–1.05) | 0.30 (0.00–0.55) |
| V20 | 3.10 (1.29–6.27) | 0.00 (0.00–0.19) | 0.00 (0.00–0.15) |
| V30 | 2.04 (0.27–4.61) | 0.00 (0.00–0.00) | 0.00 (0.00–0.00) |
| (c) LAD dose (Gy) | |||
| Mean dose | 21.73 (8.55–28.5) | 6.30 (3.51–9.31) | 6.57 (3.99–9.49) |
| V2.5 | 95.70 (85.32–99.37) | 84.62 (68.90–90.65) | 87.50 (74.83–93.93) |
| V5 | 74.91 (55.86–93.16) | 39.52 (11.12–61.87) | 54.46 (27.45–66.70) |
| V10 | 61.50 (27.79–81.50) | 12.66 (0.03–40.45) | 19.50 (4.28–35.25) |
| V20 | 48.90 (13.30–73.27) | 0.00 (0.00–9.02) | 0.96 (0.00–9.49) |
| V30 | 36.33 (0.91–58.80) | 0.00 (0.00–0.19) | 0.00 (0.00–0.23) |
| (d) Left lung dose (Gy) | |||
| Mean dose | 5.63 (4.23–6.86) | 5.54 (4.29–6.42) | 0.61 (0.47–0.80) |
| V2.5 | 30.60 (25.75–38.3) | 34.90 (28.04–39.21) | 2.52 (1.85–4.49) |
| V5 | 19.99 (15.97–25.00) | 21.23 (16.30–25.25) | 0.95 (0.34–1.61) |
| V10 | 13.27 (9.76–17.16) | 13.14 (9.84–16.37) | 0.38 (0.05–0.865) |
| V20 | 9.84 (6.39–12.98) | 9.34 (6.79–11.73) | 0.10 (0.0–0.32) |
| V30 | 7.54 (4.78–10.26) | 7.15 (4.81–8.79) | 0.01 (0.0–0.13) |
P value between PFB and SDIBH, PFB and SFB, PFB and SDIBH for all the dosimetric parameters of heart, LV, LAD, and lung. Please note that P values for heart, LV, LAD are statistically significant between SDIBH and SFB, and PFB and SFB. P values ≤ 0.05 were considered statistically significant
| Dose/volume | Technique | Heart | LV | LAD | Lung |
|---|---|---|---|---|---|
| Mean | SDIBH and SFB | 0.000 | 0.000 | 0.000 | 0.964 |
| PFB and SFB | 0.000 | 0.000 | 0.000 | 0.000 | |
| PFB and SDIBH | 0.114 | 0.137 | 0.122 | 0.000 | |
| V2.5 | SDIBH and SFB | 0.000 | 0.000 | 0.000 | 0.0080 |
| PFB and SFB | 0.000 | 0.000 | 0.034 | 0.0000 | |
| PFB and SDIBH | 0.242 | 0.055 | 0.211 | 0.0000 | |
| V5 | SDIBH and SFB | 0.000 | 0.000 | 0.000 | 0.0560 |
| PFB and SFB | 0.000 | 0.000 | 0.000 | 0.0000 | |
| PFB and SDIBH | 0.936 | 0.335 | 0.007 | 0.0000 | |
| V10 | SDIBH and SFB | 0.000 | 0.000 | 0.000 | 0.7791 |
| PFB and SFB | 0.000 | 0.000 | 0.000 | 0.0000 | |
| PFB and SDIBH | 0.765 | 0.746 | 0.153 | 0.0000 | |
| V20 | SDIBH and SFB | 0.000 | 0.000 | 0.000 | 0.6739 |
| PFB and SFB | 0.000 | 0.000 | 0.000 | 0.0000 | |
| PFB and SDIBH | 0.932 | 0.935 | 0.627 | 0.0000 | |
| V30 | SDIBH and SFB | 0.000 | 0.000 | 0.000 | 0.5143 |
| PFB and SFB | 0.000 | 0.000 | 0.000 | 0.0000 | |
| PFB and SDIBH | 0.569 | 0.311 | 0.955 | 0.0000 |
Dosimetric parameters (median values and quartiles) of OARs in SFB, SDIBH, and PFB based on breast PTV volume <750 cm3 and >=750 cm3. Please note than PFB has the lowest mean values for heart, LV, LAD, and lung for breast PTV volume >=750 cm3
| Breast PTV volume <750 cm3 | SFB | SDIBH | PFB |
|---|---|---|---|
| Mean heart dose (Gy) | 1.65 (1.12–2.32) | 0.87 (0.71–1.21) | 0.90 (0.81–1.10) |
| Mean LV dose (Gy) | 2.93 (1.85–4.04) | 1.30 (1.01–1.70) | 1.32 (1.13–1.50) |
| Mean LAD dose (Gy) | 19.86 (7.85–25.1) | 5.97 (3.01–8.53) | 6.5 (3.58–9.16) |
| Mean lung dose (Gy) | 5.48 (3.93–6.52) | 5.06 (4.09–6.38) | 0.61 (0.48–0.97) |
| Breast PTV volume >=750 cm3 | |||
| Mean heart dose (Gy) | 2.59 (1.87–4.06) | 1.36 (0.97–1.62) | 1.07 (0.87–1.31) |
| Mean LV dose (Gy) | 3.61 (3.02–5.77) | 1.72 (1.40–2.11) | 1.2 (1.11–1.58) |
| Mean LAD dose (Gy) | 24.74 (10.22–36.75) | 7.05 (3.27–12.99) | 6.7 (4.51–9.93) |
| Mean lung dose (Gy) | 5.69 (4.77–7.08) | 5.7 (5.23–7.06) | 0.57 (0.36–0.68) |
Dosimetric differences of median values between each technique, that is, SFB‐SDIBH, SFB‐PFB, SDIBH‐PFB, based on breast PTV volume
| Breast PTV volume <750 cm3 | SFB‐SDIBH | SFB‐PFB | SDIBH‐PFB |
|---|---|---|---|
| Mean heart dose (Gy) | 0.78 | 0.75 | −0.03 |
| Mean LV dose (Gy) | 1.63 | 1.61 | −0.02 |
| Mean LAD dose (Gy) | 13.96 | 13.36 | −0.6 |
| Mean lung dose (Gy) | 0.42 | 4.87 | 4.45 |
| Breast PTV volume >=750 cm3 | |||
| Mean heart dose (Gy) | 1.235 | 1.525 | 0.29 |
| Mean LV dose (Gy) | 1.89 | 2.415 | 0.525 |
| Mean LAD dose (Gy) | 17.695 | 18.045 | 0.35 |
| Mean lung dose (Gy) | −0.005 | 5.12 | 5.125 |