Nyeong Keon Kwon1, Sung Mee Jung1, Sang-Jin Park1, Yonghee Han1.
Abstract
BACKGROUND: Compared to an equivalent sized polyvinyl chloride (PVC) double-lumen tube (DLT), a silicone DLT has a shorter endobronchial segment. The aim of this study was to compare the incidence of clinically significant displacement of left-sided PVC and silicone DLTs after a positional change to a lateral position from a supine position and determine its effect on the need for DLT repositioning for successful lung separation in patients undergoing thoracic surgery.Entities:
Keywords: Bronchoscopy; Displacement; Intratracheal intubation; One-lung ventilation; Posture; Thoracic surgical procedure
Mesh:
Substances:
Year: 2018 PMID: 29921088 PMCID: PMC6369336 DOI: 10.4097/kja.d.18.00012
Source DB: PubMed Journal: Korean J Anesthesiol ISSN: 2005-6419
Fig. 1.(A) A 37 Fr silicone (Human Bronch®) double-lumen tube. (B) A 37 Fr polyvinyl chloride (Broncho-CathTM) double-lumen tube.
Fig. 2.Flow Chart. Group P: polyvinyl chloride double-lumen tube, Group S: silicone double-lumen tube. FOB: fiberoptic bronchoscope.
Demographic, Anesthetic, and Surgical Characteristics of the Patients
| Variable | Group P (n = 48) | Group S (n = 52) | P value |
|---|---|---|---|
| Age (yr) | 56.7 ± 16.1 | 57.8 ± 14.9 | 0.704 |
| Sex (M/F) | 34/14 | 35/17 | 0.703 |
| Weight (kg) | 64.3 ± 10.3 | 60.9 ± 8.7 | 0.074 |
| Height (cm) | 165.5 ± 9.3 | 163.7 ± 7.6 | 0.290 |
| ASA physical status (I/II/III) | 13/28/7 | 10/37/5 | 0.404 |
| Left main bronchus | |||
| Length (mm) | 49.2 ± 5.3 | 47.9 ± 5.0 | 0.209 |
| Diameter (mm) | |||
| Anteroposterior | 13.4 ± 2.0 | 13.2 ± 2.0 | 0.635 |
| Transverse | 14.4 ± 2.1 | 14.6 ± 2.0 | 0.750 |
| Bronchial angle (°) | |||
| Right | 37.3 ± 8.5 | 39.2 ± 9.6 | 0.311 |
| Left | 41.2 ± 7.1 | 41.8 ± 7.0 | 0.721 |
| Size of double-lumen tube (Fr) | 0.703 | ||
| 35 | 14 (29.2%) | 17 (32.7%) | |
| 37 | 34 (70.8%) | 35 (67.3%) | |
| Side of surgery | 0.733 | ||
| Right | 27 (56.3%) | 31 (59.6%) | |
| Left | 21 (43.8%) | 21 (40.4%) | |
| Type of surgery | 0.061 | ||
| Thoracotomy | 32 (66.7%) | 25 (48.1%) | |
| Thoracoscopy | 16 (33.3%) | 27 (51.9%) | |
| Duration of anesthesia (min) | 257.9 ± 97.8 | 227.8 ± 93.2 | 0.118 |
| Duration of operation (min) | 164.2 ± 73.9 | 151.0 ± 86.9 | 0.417 |
Data are presented as mean ± SD or number of patients (%). Group P: polyvinyl chloride double-lumen tube, Group S: silicone double-lumen tube.
Lengths and Incidences of Tracheal and Left Main Bronchial Displacement
| Group P (n = 48) | Group S (n = 52) | P value | |
|---|---|---|---|
| Lateral positioning | |||
| Tracheal displacement (mm) | 6.4 ± 10.0 (−10.6, 23.0) | 3.1 ± 10.4 (−18.8, 19.8) | 0.104 |
| Bronchial displacement (mm) | 4.1 ± 9.4 (−12.6, 21.5) | 6.2 ± 10.1 (−16.6, 25.4) | 0.289 |
| Clinically significant displacement | 17 (35.4%) | 18 (34.6%) | 0.933 |
| Critical malposition[ | 4 (8.3%) | 4 (7.9%) | 0.906 |
| Surgical manipulation | |||
| Tracheal displacement (mm) | 1.3 ± 8.3 (−15.2, 18.3) | −2.0 ± 9.0 (−15.2, 13.8) | 0.070 |
| Bronchial displacement (mm) | −0.9 ± 5.1 (−9.0, 9.0) | −2.0 ± 9.4 (−19.8, 14.0) | 0.499 |
| Clinically significant displacement | 10 (21.7%) | 16 (31.4%) | 0.285 |
| Critical malposition[ | 0 | 0 |
Data are presented as mean ± SD (range) or the number of patients (%). The displacement of the double-lumen tube was determined by changes in tracheal and bronchial distances, obtained by subtracting supine measurements from lateral measurements and subtracting measurements at the start of surgery from measurements at the end of surgery, respectively. The direction of displacement is presented as positive values for proximal movement and negative values for distal movement, respectively. Clinically significant displacement was defined when the double-lumen tube was deviated by more than 10 mm from the initial correct position, regardless of the direction of displacement.
Requiring reposition of the double-lumen tube for successful one-lung ventilation.
Group P: polyvinyl chloride double-lumen tube, Group S: silicone double-lumen tube.
Fig. 3.Direction of clinically significant displacement of the left-sided double-lumen tube. Group P: polyvinyl chloride double-lumen tube, Group S: silicone double-lumen tube.