Literature DB >> 29920949

Evaluation of a new commercial method for von Willebrand factor multimeric analysis.

R A Crist1, N M Heikal1,2, G M Rodgers1,2,3, D G Grenache1,2, K J Smock1,2.   

Abstract

INTRODUCTION: Evaluation of von Willebrand factor (VWF) multimeric distribution is useful for subclassification of von Willebrand disease (VWD). Multimer analysis has historically been a manual, labor-intensive laboratory-developed test. The first commercial method for multimeric analysis was recently developed that utilizes a single instrument for gel electrophoresis, staining, and densitometry. The current study was undertaken to evaluate the performance characteristics of the new commercial method.
METHODS: Studies performed with the commercial method included evaluation of accuracy (method comparison), reference intervals (establishment of normal migration patterns in normal donor specimens), precision (multimer pattern reproducibility), and analytical sensitivity.
RESULTS: In the method comparison studies, concordant interpretations were obtained in 19 of 24 comparisons, including normal and abnormal specimens. The 5 specimens with discordant interpretations all involved slight differences and none were considered clinically significant. Thirty-eight normal donor specimens demonstrated normal multimer patterns. Multimer pattern reproducibility was demonstrated in normal and abnormal controls tested on each gel. In the sensitivity studies, adequate visualization of multimers was determined to require VWF protein concentrations of approximately 5%-10% of normal.
CONCLUSION: The commercial multimer method is a streamlined test that demonstrates comparable performance characteristics to our current laboratory-developed method and that provides the advantage of both electrophoresis gels and densitometry scans to aid interpretation.
© 2018 John Wiley & Sons Ltd.

Entities:  

Keywords:  multimers; von Willebrand disease; von Willebrand factor

Year:  2018        PMID: 29920949     DOI: 10.1111/ijlh.12869

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  4 in total

1.  Improving diagnosis of von Willebrand disease: Reference ranges for von Willebrand factor multimer distribution.

Authors:  Inge Vangenechten; Alain Gadisseur
Journal:  Res Pract Thromb Haemost       Date:  2020-07-16

2.  Evaluation of a new semi-automated Hydragel 11 von Willebrand factor multimers assay kit for routine use.

Authors:  Marika Pikta; Timea Szanto; Margus Viigimaa; Sandra Lejniece; Dārta Balode; Kadri Saks; Valdas Banys
Journal:  J Med Biochem       Date:  2021-03-12       Impact factor: 3.402

3.  Establishing reference intervals for von Willebrand factor multimers.

Authors:  Marika Pikta; Marc Vasse; Kristi J Smock; Karen A Moser; Dievoet Marie-Astrid van; Sandra Lejniece; Timea Szanto; Hector Bautista; George Nouadje; Valdas Banys
Journal:  J Med Biochem       Date:  2022-02-02       Impact factor: 3.402

4.  Von Willebrand Factor Multimer Densitometric Analysis: Validation of the Clinical Accuracy and Clinical Implications in Von Willebrand Disease.

Authors:  Johan Boender; Ferdows Atiq; Marjon H Cnossen; Johanna G van der Bom; Karin Fijnvandraat; Joke de Meris; Moniek P M de Maat; Karin P M van Galen; Britta A P Laros-van Gorkom; Karina Meijer; Jeroen Eikenboom; Frank W G Leebeek
Journal:  Hemasphere       Date:  2021-02-17
  4 in total