Saviour of stylet injuries: 'Ryle's tube'.

Rafat Shamim1, Gaurav Sindwani1, Vansh Priya2, Aditi Suri3.   

Abstract

Entities:  

Year:  2018        PMID: 29910504      PMCID: PMC5971635          DOI: 10.4103/ija.IJA_25_18

Source DB:  PubMed          Journal:  Indian J Anaesth        ISSN: 0019-5049


× No keyword cloud information.
Sir, The endotracheal tube (ETT) stylet is commonly used in difficult airway management. Although its use can increase the success rate of difficult tracheal intubation,[1] this comes at the risk of complications such as tracheal perforation, soft palate perforation, tonsillar injury and mucosal injury. Stylets can be rigid or malleable. A malleable stylet like Mallinckrodt's intubating stylet has a soft polyurethane covering over the aluminium wire. Chances of injury are lower with malleable stylet when compared to a rigid stylet. However, they are costly and not meant for reuse, thereby restricting their widespread use in the developing countries like India. Hence, we suggest an alternative cheap method of converting such rigid intubation stylets into a much safer airway adjunct, wherein a nasogastric (Ryle's) tube was mounted over an ETT as a stylet guard. In this technique, a Ryle's tube was cut at an appropriate length for the stylet [Figure 1]. After lubrication jelly application, the Ryle's tube was mounted slowly and carefully over the aluminium stylet, in such a way that the stylet tip was kept well behind the tip of Ryles tube [Figure 2]. Special care should be taken while selecting Ryles tube size. Ryle's tube should fix snuggly over the stylet, such that it does not slide over the stylet itself. In our case, we have used 16 F Ryle's tube for the medium size aluminium stylet. Stylets come in different sizes and the width can vary based on its manufacturer. For adult stylets, either 16 F or 14 F Ryle's tube fits snuggly over the stylet, and similarly, for paediatric stylet, either 8 F or 10 F Ryle's tube fits snuggly over the stylet.
Figure 1

Appropriately cut Ryles tube for the stylet

Figure 2

Mounted Ryles tube over the stylet with the stylet tip well behind the Ryles tube

Appropriately cut Ryles tube for the stylet Mounted Ryles tube over the stylet with the stylet tip well behind the Ryles tube Our technique combines the benefit of malleability and rigidity of aluminium with the protective covering of Ryles tube. Most of the stylet-induced injuries are either caused by its blunt tip or by the brittle broken piece of the stylet. Stylet-induced injuries can occur when resistance is encountered while inserting the ETT, thus exposing the stylet tip to the subglottic structures or while withdrawing the stylet, when the acute hockey stick angle, the ETT can cause injury to the anterior part of the trachea.[2] There are many cases where either a broken stylet piece or torn sheath covering the stylet were found causing the airway obstruction.[34] These may occur due to usage of an old brittle stylet or reusing the same stylet more than the recommended number of times. With our technique, both stylet tip injuries and airway obstruction resulting from the broken brittle stylet can be taken care of as the tip of the stylet is well behind the Ryle's tube tip. Another advantage is that the same stylet can be reused making our approach highly cost-effective. However, we suggest to change the Ryles tube, covering stylet after each use and proper sterilisation of the stylet before mounting a new Ryle's tube. To reduce stylet-induced soft-tissue injuries, in addition to our technique, we also suggest avoiding excessive bending of stylet tip, use of adjustable rubber cock and proper jelly application before introducing the stylet into the ETT.[25] We have been regularly using this technique in our institute. However, we suggest larger trials before this technique can be used widely.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  4 in total

1.  Stylet bend angles and tracheal tube passage using a straight-to-cuff shape.

Authors:  Richard M Levitan; James T Pisaturo; William C Kinkle; Kenneth Butler; Worth W Everett
Journal:  Acad Emerg Med       Date:  2006-11-01       Impact factor: 3.451

2.  An Unnoticed Broken Sheathed Metallic Stylet in an Endotracheal Tube: A case report.

Authors:  Pradeep K Sharma; Rashid M Khan; Naresh Kaul
Journal:  Sultan Qaboos Univ Med J       Date:  2010-04-17

3.  Practice guidelines for management of the difficult airway: an updated report by the American Society of Anesthesiologists Task Force on Management of the Difficult Airway.

Authors:  Jeffrey L Apfelbaum; Carin A Hagberg; Robert A Caplan; Casey D Blitt; Richard T Connis; David G Nickinovich; Carin A Hagberg; Robert A Caplan; Jonathan L Benumof; Frederic A Berry; Casey D Blitt; Robert H Bode; Frederick W Cheney; Richard T Connis; Orin F Guidry; David G Nickinovich; Andranik Ovassapian
Journal:  Anesthesiology       Date:  2013-02       Impact factor: 7.892

4.  A piece of broken metal from intubation stylet retained in tracheobronchial tree: a case report.

Authors:  Mohammad Fathi; Behrooz Farzanegan; Meysam Mojtabaee; Abdolrahim Nikzamir
Journal:  Tanaffos       Date:  2014
  4 in total
  1 in total

1.  Discrimination ability of the endotracheal tube location using real-time palpation during intubation using an endotracheal tube with a preloaded stylet.

Authors:  Go Un Roh; Yun Jeong Chae; Young Bok Lee; Wikwang Wang; Chang Ik Choi; In Kyong Yi
Journal:  Ther Clin Risk Manag       Date:  2018-07-24       Impact factor: 2.423