| Literature DB >> 29909801 |
Sreeja Pavithran1, Kothandam Sivakumar1.
Abstract
A neonatal ductal stent deployed in a straight short conical duct on second postnatal day migrated owing to inadequate ductal constriction. It was successfully retrieved using a larger balloon and redeployed in the duct again. Intravenous indomethacin prevented further stent migration. This is the first report of successful transcatheter retrieval and repositioning of a migrated expanded neonatal ductal stent.Entities:
Keywords: Patent arterial duct; drug-eluting stent; indomethacin; prostaglandin E1; stent migration
Mesh:
Year: 2018 PMID: 29909801 DOI: 10.1017/S104795111700292X