Literature DB >> 29909212

Flow Diversion via LVIS Blue Stent within Enterprise Stent in Patients with Vertebral Artery Dissecting Aneurysm.

Yong Cheol Lim1, Yong Sam Shin2, Joonho Chung3.   

Abstract

A new Low-profile Visualized Intraluminal Support device (LVIS Blue) is a braided stent that provides a higher degree of metal coverage. This stent may be beneficial for complete obliteration of an aneurysm due to not only its support of a higher occlusion rate using coils inside of the aneurysm but also its flow-diverting effect. We introduce and describe a new flow-diversion technique using an LVIS Blue stent within an Enterprise stent for patients with vertebral artery (VA) dissecting aneurysms. A 6-Fr guiding catheter was positioned in the ipsilateral VA. A microcatheter with an inner diameter of 0.021 inches was navigated and positioned across the dissecting segment of the VA. An Enterprise stent was then deployed and used to cover the dissecting segment by pushing the pusher wire of the stent while simultaneously unsheathing the microcatheter. The microcatheter was navigated along the remaining pusher wire to the distal VA through the Enterprise stent. An LVIS Blue stent was introduced and positioned such that it overlapped the previous stent in order to divert blood flow away from the dissecting segment. We present a case in which this technique was used to successfully obliterate a dissecting aneurysm while maintaining the VA patency with vessel wall remodeling. The LVIS Blue stent-within-an-Enterprise stent technique may be an effective treatment option for patients with VA dissecting aneurysms as an alternative technique to flow diverters.
Copyright © 2018 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Dissecting aneurysm; Enterprise; Flow diverters; LVIS Blue; Vertebral artery dissection

Mesh:

Year:  2018        PMID: 29909212     DOI: 10.1016/j.wneu.2018.06.029

Source DB:  PubMed          Journal:  World Neurosurg        ISSN: 1878-8750            Impact factor:   2.104


  7 in total

1.  Blister-like aneurysm of the anterior communicating artery treated with only Low-profile Visualized Intraluminal Support Junior stent.

Authors:  Katsuyoshi Miyashita; Kosuke Nambu; Yu Shimizu; Yasuo Tohma
Journal:  Surg Neurol Int       Date:  2021-11-16

2.  Overlapping Stents-Assisted Coiling for Vertebral Artery Dissecting Aneurysm : LVIS Stent within Neuroform EZ Stent.

Authors:  Xing-Long Liu; Bin Wang; Lin-Bo Zhao; Zhen-Yu Jia; Hai-Bin Shi; Sheng Liu
Journal:  J Korean Neurosurg Soc       Date:  2022-04-18

3.  Low-profile visualized intraluminal support Blue stenting within a Neuroform Atlas stent for a large wide-necked aneurysm: A case report and a bench-top experiment.

Authors:  Tomoko Hayashi; Hiroyuki Ikeda; Ryota Ishibashi; Toshio Fujiwara; Ryosuke Kaneko; Minami Uezato; Masanori Kinosada; Yoshitaka Kurosaki; Akira Handa; Masaki Chin
Journal:  Neuroradiol J       Date:  2021-06-27

4.  Detection of Angiographically Occult Ruptured Basilar Sidewall Perforator Aneurysm by Vessel Wall MR Imaging.

Authors:  Mougnyan Cox; Jae W Song; Seyed Ali Nabavizadeh; David Kung; Laurie Loevner; Omar Choudhri
Journal:  Neurohospitalist       Date:  2020-10-05

5.  Endovascular treatment for ruptured vertebral dissecting aneurysms involving PICA: Reconstruction or deconstruction? Experience from 16 patients.

Authors:  Xiangjie Kong; Zeyu Sun; Chenhan Ling; Liang Xu; Cong Qian; Jun Yu; Jing Xu
Journal:  Interv Neuroradiol       Date:  2020-10-28       Impact factor: 1.610

Review 6.  Intracranial Fusiform and Circumferential Aneurysms of the Main Trunk: Therapeutic Dilemmas and Prospects.

Authors:  Yunbao Guo; Ying Song; Kun Hou; Jinlu Yu
Journal:  Front Neurol       Date:  2021-07-09       Impact factor: 4.003

7.  LVIS-within-enterprise double-stent procedure without coiling beneficial as treatment of unruptured vertebral artery dissecting aneurysms.

Authors:  Hyun Ho Choi; Young Dae Cho; Dong Hyun Yoo; Hyun-Seung Kang; Moon Hee Han
Journal:  Interv Neuroradiol       Date:  2021-06-18       Impact factor: 1.764

  7 in total