Addition of minute ventilation to rate-response pacing improves heart rate score more than accelerometer alone.

Mark Richards1, Brian Olshansky2, Arjun D Sharma3, Nicholas Wold4, Paul Jones4, David Perschbacher4, Bruce L Wilkoff5.   

Abstract

BACKGROUND: Heart rate score (HRSc) ≥70%, a novel parameter, predicts risk of mortality in patients with implantable cardioverter-defibrillators and identifies patients who have survival benefit with DDDR vs DDD pacing.
OBJECTIVE: The purpose of this study was to determine if DDDR pacing lowers HRSc, and a blended sensor with minute ventilation (MV) and accelerometer (XL) improves HRSc more than accelerometer (XL) alone in patients requiring pacemakers (PMs).
METHODS: HRSc, the percentage of all beats in the tallest 10-beat/min device histogram bin, was calculated. Data from the Limiting Chronotropic Incompetence for Pacemaker Recipients Study, a prospective randomized trial of PM patients, comparing XL to blended-sensor (XL + MV) rate-responsive pacing, were analyzed retrospectively for HRSc changes from baseline. The relationship of patient activity (sensor-detected from device memory) to HRSc was examined.
RESULTS: Of the 501 randomized patients, 215 (43%) patients had HRSc ≥70% during DDD pacing at baseline. In these patients, HRSc decreased after DDDR programming by 14.2%, while it increased by 0.4% in those with baseline HRSc <70% (n = 286) (HRSc ≥70% vs HRSc <70%; P < .01). No differences were detected between the 2 randomized sensor-based groups at baseline. Blended-sensor (MV + XL) programming reduced HRSc more than the XL sensor alone (MV + XL: 18% vs XL: 10%; P < .001). No correlation was observed between patient activity and HRSc (correlation = -0.14; P = .07).
CONCLUSION: HRSc improved (reduced) with rate-response (DDDR) programming in PM patients with high HRSc during DDD pacing. Blended sensors (MV + XL) improved HRSc more than XL alone. HRSc does not correlate with patient activity levels, suggesting that other patient factors determine this parameter. This programming approach needs to be investigated prospectively in a PM outcomes trial.
Copyright © 2018 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Bradycardia; Heart rate; Heart rate score; Minute ventilation; Pacemaker; Rate-response pacing; Sinus node dysfunction

Mesh:

Year:  2018        PMID: 29908371     DOI: 10.1016/j.hrthm.2018.06.021

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  2 in total

1.  Impact of a poor functional capacity on the clinical outcomes in patients with a pacemaker implantation -Results from the Japanese Heart Rhythm Society Registry.

Authors:  Takanori Arimoto; Eiichi Watanabe; Ritsuko Kohno; Kenji Shimeno; Kan Kikuchi; Atsushi Doi; Kanki Inoue; Takashi Nitta; Akihiko Nogami; Haruhiko Abe; Ken Okumura
Journal:  J Arrhythm       Date:  2020-12-09

2.  Heart rate score, a measure related to chronotropic incompetence in pacemaker patients.

Authors:  Arjun D Sharma; Mark Richards; Brian Olshansky; Nicholas Wold; Paul Jones; David Perschbacher; Bruce L Wilkoff
Journal:  Heart Rhythm O2       Date:  2021-02-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.