Literature DB >> 29902926

A report of 3 cases of Exeter V40 Stem fracture and explanation of possible causes.

Kanai Garala1, Thomas Laios2, Trevor Lawrence3.   

Abstract

BACKGROUND: The Exeter V40 Orthinox stem is one of the most commonly used femoral components in total hip arthroplasty. Its design, material and finish has evolved significantly from the original design established by Ling and Lee over 3 decades ago. An early problem reported with the original Exeter stem was a high rate of stem fractures. It was thought that with the new Orthinox steel used in the V40 system, the rate of stem fractures would reduce. There are reports of the Exeter V40 stem failing in a variety of locations. CASES: This paper highlights the 3 locations where the Exeter stem may fail and estimates a rate of 0.262% for stem fractures due to the constant use of the Exeter stem throughout the career of the senior surgeon.

Entities:  

Keywords:  Exeter stem; fracture; incidence; total hip replacement

Mesh:

Year:  2018        PMID: 29902926     DOI: 10.1177/1120700018780887

Source DB:  PubMed          Journal:  Hip Int        ISSN: 1120-7000            Impact factor:   2.135


  2 in total

1.  History, treatment and analysis of a rare form of Exeter stem fracture.

Authors:  Darren Patrick Moloney; Robert J Hurley; James Harty; Shane Guerin
Journal:  BMJ Case Rep       Date:  2019-12-30

2.  Analysis of the Exeter V40 femoral stem prosthesis fracture : systematic review and single centre case series.

Authors:  Joshua W Thompson; James Corbett; Daniel Bye; Adrian Jones; Elizabeth K Tissingh; John Nolan
Journal:  Bone Jt Open       Date:  2021-06
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.