The major adrenal steroids were separated by multistep gradient elution with a reversed-phase high-performance liquid chromatography system, employing water and 1-propanol as solvents. With this solvent system, a wide range of 21 5-ene-3 beta-ol and 4-ene-3-one steroids can be resolved in a single chromatogram, which was not possible with previously published gradient solvent systems. In particular, intermediate steroids of the biosynthetic pathway, 17 alpha-hydroxyprogesterone, 17 alpha-hydroxypregnenolone, and dehydroepiandrosterone, were separated with baseline or sufficient resolution to allow accurate quantitation. Using the 1-propanol-water gradient, the separations of 5-ene and 4-ene steroids were compared on different octadecylsilyl packings. Optimum resolution was obtained with a fully covered, spherical particle. The 1-propanol-water gradient was compared to a previously published methanol-water gradient in the analysis of steroidogenesis by adrenocortical cell cultures. HPLC analysis of the steroid production was quantitatively the same with both gradient solvent systems. However, qualitatively, the methanol-water gradient system did not resolve the above-mentioned intermediate steroids.

Entities:  

Mesh:

Substances:

Year:  1985        PMID: 2990253     DOI: 10.1016/0003-2697(85)90285-4

Source DB:  PubMed          Journal:  Anal Biochem        ISSN: 0003-2697            Impact factor:   3.365


  1 in total

1.  Clonal growth and culture life span of bovine adrenocortical cells in a serum-free medium.

Authors:  M H Simonian; M L White; D A Foggia
Journal:  In Vitro Cell Dev Biol       Date:  1987-04
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.