Concordance in the measurement of cardiac output. Vigileo vs. Capstesia

Janaí Santiago-López1, Víctor León-Ramírez, Sheila Hernández-Ramírez, Petra Isidora Vásquez-Márquez, Antonio Castellanos-Olivares.   

Abstract

Background: Capstesia is an app for smartphones that can serve as a simple, accessible and affordable alternative to current advanced hemodynamic monitoring devices. The aim of this paper was to establish the degree of concordance between the Vigileo monitor and the Capstesia app in the measurement of cardiac output in patients undergoing cardiac surgery.
Methods: A process study was conducted that included 30 patients whose cardiac output was measured simultaneously with the Capstesia app and the Vigileo monitor. The analysis was done using the Lin's concordance correlation coefficient and Student's t - test. A p < 0.05 was considered significant. The information was processed in SPSS v. 22.0.
Results: 332 paired samples were taken. The cardiac output values obtained were 4.43 ± 1.37 L/min for the Vigileo monitor and 4.45 ± 1.38 L/min for the Capstesia app. Lin's concordance correlation coefficient yielded a value of 0.966 with a confidence range of 0.958-0.973. Conclusions: There is substantial concordance between the cardiac output values yielded by the Vigileo monitor and the Capstesia app in patients undergoing cardiac surgery.

Entities:  

Keywords:  Cardiac Output; Hemodynamic Monitoring; Thoracic Surgery

Mesh:

Year:  2018        PMID: 29901910

Source DB:  PubMed          Journal:  Rev Med Inst Mex Seguro Soc        ISSN: 0443-5117


  2 in total

1.  Estimation of pulse pressure variation and cardiac output in patients having major abdominal surgery: a comparison between a mobile application for snapshot pulse wave analysis and invasive pulse wave analysis.

Authors:  Phillip Hoppe; Fabian Gleibs; Luisa Briesenick; Alexandre Joosten; Bernd Saugel
Journal:  J Clin Monit Comput       Date:  2020-08-04       Impact factor: 2.502

2.  Hemodynamic Monitoring by Smartphone-Preliminary Report from a Comparative Prospective Observational Study.

Authors:  Michał P Pluta; Magdalena Dziech; Mateusz N Zachura; Anna J Szczepańska; Piotr F Czempik; Piotr S Liberski; Łukasz J Krzych
Journal:  J Pers Med       Date:  2022-02-01