The Impact of Suture Caliber and Looped Configurations on the Suture-Tendon Interface in Zone II Flexor Tendon Repair.

Derek T Bernstein1, Jamie J Alexander2, Nancy J Petersen3, Bradley S Lambert1, Philip C Noble2, David T Netscher4.   

Abstract

PURPOSE: To evaluate the impact of suture caliber and looped configurations on the integrity of 4-strand modified Kessler zone II flexor tendon repairs during progressive cyclic loading.
METHODS: Seventy-two flexor digitorum profundus tendons from 18 fresh human cadaver hands were divided into 4 repair groups. Thirty-six matched tendons underwent repair using either a 4-0 looped or 4-0 single-stranded suture, and an additional 36 tendons underwent 3-0 looped or 3-0 single-stranded repairs. Repair strength was tested by progressive cyclic loading. The force generating 2-mm gap formation, ultimate failure, and the mechanism of failure were recorded for each test. The impact of looped versus single-stranded configurations and the effect of tendon cross-sectional area on repair integrity were analyzed for each suture caliber.
RESULTS: There was no statistically significant difference between groups regarding the force to 2-mm gap formation or ultimate failure, and all values exceeded the minimum threshold of 27 N required to withstand an early active range of motion rehabilitation protocol. The use of a 3-0 caliber suture resulted in a significantly higher proportion of repairs failing by suture pullout through the tendon substance, including 63.5% of looped and 38.9% of single-stranded core sutures. By comparison, this occurred in 11.1% of 4-0 looped and 0% of 4-0 single-stranded sutures. Larger tendon cross-sectional areas were associated with more robust repairs, particularly in the 3-0 looped group.
CONCLUSIONS: In a human cadaver flexor tendon repair model, there was no significant difference in the mean force to failure between all 4 flexor tendon repair constructs under progressive cyclic loading. However, the 3-0 caliber suture failed more frequently by suture pullout, particularly with the use of a looped suture. CLINICAL RELEVANCE: Four-strand flexor tendon repairs using a 3-0 caliber suture are more prone to early failure by suture pullout under progressive cyclic loading compared with a 4-0 caliber suture.
Copyright © 2019 American Society for Surgery of the Hand. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Flexor tendon repair; caliber; looped; suture; suture-tendon interface; zone II

Year:  2018        PMID: 29891271     DOI: 10.1016/j.jhsa.2018.04.029

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  1 in total

1.  Beyond the Core Suture: A New Approach to Tendon Repair.

Authors:  Weifeng Zeng; Nicholas J Albano; Ruston J Sanchez; Ronald Mccabe; Ray Vanderby; Samuel O Poore; Aaron M Dingle
Journal:  Plast Reconstr Surg Glob Open       Date:  2020-12-17
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.