Image quality assessment with dose reduction using high kVp and additional filtration for abdominal digital radiography.

Ji Sung Jang1, Hyung Jin Yang2, Hyun Jung Koo3, Sung Ho Kim2, Chan Rok Park2, Suk Hwan Yoon2, So Youn Shin4, Kyung-Hyun Do5.   

Abstract

PURPOSE: Dose reduction using additional filters with high kilovoltage peak (kVp) for abdominal digital radiography has received much attention recently. We evaluated image quality with dose reduction in abdominal digital radiography by using high kVp and additional copper filters at a tertiary hospital.
METHODS: Between June 2016 and July 2016, 82 patients underwent abdominal digital radiography using 80 kVp in X-ray room 1 and 82 were imaged using 92 kVp with 0.1-mm copper filtration in X-ray room 2. The effective dose was calculated using a PC-based Monte Carlo program. Image quality of the abdominal radiography acquired in the two rooms was evaluated using a five-point ordinal scale, as well as the signal-to-noise and contrast-to-noise ratios.
RESULTS: The mean effective dose decreased by 25.8% and 25.7% for the supine and standing positions, respectively, when abdominal digital radiography using 92 kVp with 0.1-mm copper filtration was performed. In the 20 patients who performed abdominal digital radiography twice in each room, visual grading scores for visualisation of psoas outlines and kidney outlines are higher in room 1. However, there was no statistical significant difference of visual grading scores among the 124 patients who underwent only one abdominal radiography in the room 1 or 2 (P > 0.05).
CONCLUSIONS: Dose reduction for abdominal digital radiography can be achieved with comparable image quality by performing abdominal digital radiography using 92 kVp with 0.1-mm copper filtration, despite the higher AEC dose.
Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Abdominal radiography; Copper filter; Dose reduction; Effective dose

Mesh:

Year:  2018        PMID: 29891093     DOI: 10.1016/j.ejmp.2018.05.007

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

1.  Effect of Silicone Rubber-Lead (SR-Pb) Thickness on Dose Reduction and Image Quality as Gonad Shield.

Authors:  Zahroh F; Anam C; Sutanto H; Irdawati Y; Arifin Z; Kartikasari Y
Journal:  J Biomed Phys Eng       Date:  2020-12-01

2.  Intelligent Image Diagnosis of Pneumoconiosis Based on Wavelet Transform-Derived Texture Features.

Authors:  Zichen Wang; Maoneng Hu; Min Zeng; Guoliang Wang
Journal:  Comput Math Methods Med       Date:  2022-03-17       Impact factor: 2.238

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.