Literature DB >> 29881199

Long-term results of metal on metal total hip arthroplasty in younger patients (<55yrs).

Muhammad Umar1, Noman Jahangir2, Qasim Malik1, Steven Kershaw1, Keith Barnes1, Syam Morapudi1.   

Abstract

INTRODUCTION: Metal on Metal (MoM) hip arthroplasty saw a new era of popularity with development of its second generation bearing surfaces, in the first decade of this century. However, by the end of that decade, concerns had been raised due to metal debris related complications.We aimed to determine the survival of MoM stemmed hip replacement in younger population. We also studied the rate of revision related to adverse reaction to metal debris (ARMD) along with reviewing the clinical and radiological progress of MoM hip arthroplasty in younger age (<55 yrs) group. PATIENTS &
METHODS: This is a retrospective cohort study of patients 55 yrs old or younger, who had metal on metal (MoM) hip arthroplasty for osteoarthritis. We had 109 procedures performed on 90 patients with a mean follow up of 10 years. All patients were reviewed as per MHRA guidelines in planned follow-up clinics. Data analyses were performed using SPSS.
RESULTS: Survival of implant in our younger cohort was 88.1% at a mean age if 10 years, with revision for any cause as an endpoint. Most of the patients were happy with the outcome of their hip replacement as they were able to perform activities of daily living and work without compromise. Mean Oxford hip score was 43.Altogether, there were 12 revisions, 7 of these were for metallosis and associated symptoms. Average time to revision was 7 years. Other analysis revealed mean acetabular cup inclination angle to be 49 degrees but no significant correlation was found between this angle and serum metal ion levels. Serum Chromium and Cobalt levels were significantly higher in revision group. DISCUSSION: Metal on metal hip arthroplasty prime popularity time has gone. In younger population, although revision rates are higher, the surviving implants give a very good outcome in terms of patient satisfaction. Most of the patients report a desired outcome of 'forgotten hip'.

Entities:  

Keywords:  Hip arthroplasty; Metal ions; Metal on metal; Oxford hip score; Younger patients

Year:  2018        PMID: 29881199      PMCID: PMC5990322          DOI: 10.1016/j.jor.2018.05.041

Source DB:  PubMed          Journal:  J Orthop        ISSN: 0972-978X


  23 in total

Review 1.  Metal-on-metal bearings: the evidence so far.

Authors:  F S Haddad; R R Thakrar; A J Hart; J A Skinner; A V F Nargol; J F Nolan; H S Gill; D W Murray; A W Blom; C P Case
Journal:  J Bone Joint Surg Br       Date:  2011-05

2.  The Metasul metal-on-metal articulation in primary total hip replacement: clinical and radiological results at ten years.

Authors:  V Eswaramoorthy; P Moonot; Y Kalairajah; L C Biant; R E Field
Journal:  J Bone Joint Surg Br       Date:  2008-10

3.  The use of the Oxford hip and knee scores.

Authors:  D W Murray; R Fitzpatrick; K Rogers; H Pandit; D J Beard; A J Carr; J Dawson
Journal:  J Bone Joint Surg Br       Date:  2007-08

4.  Birmingham hip resurfacing at a mean of ten years: results from an independent centre.

Authors:  G Coulter; D A Young; R E Dalziel; A J Shimmin
Journal:  J Bone Joint Surg Br       Date:  2012-03

5.  Long-term clinical results of the Metasul metal-on-metal total hip arthroplasty: 12.6 years follow-up of 128 primary total hip replacements.

Authors:  Hendrik A Zuiderbaan; Dennis Visser; Inger N Sierevelt; Janine Penders; Jeanette Verhart; Diederik A Vergroesen
Journal:  Hip Int       Date:  2017-10-14       Impact factor: 2.135

6.  Cementless metal-on-metal versus ceramic-on-polyethylene hip arthroplasty in patients less than fifty years of age: a comparative study with twelve to fourteen-year follow-up.

Authors:  Henri Migaud; Sophie Putman; Nicolas Krantz; Laurent Vasseur; Julien Girard
Journal:  J Bone Joint Surg Am       Date:  2011-05       Impact factor: 5.284

7.  Long-term (20- to 25-year) results of an uncemented tapered titanium femoral component and factors affecting survivorship.

Authors:  Marcus R Streit; Moritz M Innmann; Christian Merle; Thomas Bruckner; Peter R Aldinger; Tobias Gotterbarm
Journal:  Clin Orthop Relat Res       Date:  2013-05-14       Impact factor: 4.176

8.  Pseudotumours associated with metal-on-metal hip resurfacings.

Authors:  H Pandit; S Glyn-Jones; P McLardy-Smith; R Gundle; D Whitwell; C L M Gibbons; S Ostlere; N Athanasou; H S Gill; D W Murray
Journal:  J Bone Joint Surg Br       Date:  2008-07

9.  Outcomes of a metal-on-metal total hip replacement system.

Authors:  G S Matharu; K Theivendran; P B Pynsent; L Jeys; A M Pearson; D J Dunlop
Journal:  Ann R Coll Surg Engl       Date:  2014-10       Impact factor: 1.891

10.  Independent predictors of failure up to 7.5 years after 35 386 single-brand cementless total hip replacements: a retrospective cohort study using National Joint Registry data.

Authors:  S S Jameson; P N Baker; J Mason; M Rymaszewska; P J Gregg; D J Deehan; M R Reed
Journal:  Bone Joint J       Date:  2013-06       Impact factor: 5.082

View more
  2 in total

1.  CORR Insights®: Partially Melted Ti6Al4V Particles Increase Bacterial Adhesion and Inhibit Osteogenic Activity on 3D-printed Implants: An In Vitro Study.

Authors:  Jonathan P Garino
Journal:  Clin Orthop Relat Res       Date:  2019-12       Impact factor: 4.176

2.  Long-term results of the M2A-38-mm metal-on-metal articulation.

Authors:  Carlo Trevisan; Stefano Piscitello; Raymond Klumpp; Tonino Mascitti
Journal:  J Orthop Traumatol       Date:  2018-12-07
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.