Ultrasound probe use and reprocessing: Results from a national survey among U.S. infection preventionists.

Ruth M Carrico1, Stephen Furmanek2, Connor English2.   

Abstract

BACKGROUND: Improper infection prevention practice associated with ultrasound probe use has been linked to increased infection risk, outbreaks, and death. Although guidelines for reprocessing and use of probes exist, it is unclear how extensively these have been adopted in practice.
METHODS: Infection preventionists from U.S. health care facilities were surveyed (N = 358). The anonymous survey had 31 multiple choice, sliding scale, and text response questions. The survey was developed and deployed and the data were stored in the REDCap system.
RESULTS: A high degree of noncompliance with U.S. guidelines was identified. Surface probes used in invasive procedures were not high-level disinfected or sterilized 15% (intraoperative) to 78% (peripheral line placements) of the time. Of invasive procedures, 5%-47% did not use sterile gel (same procedures, respectively). Of the participants, 20% were aware of instances where an ultrasound probe was used but was not correctly reprocessed. Extensive breaches of infection control guidelines were reported. The rapid expansion in use of ultrasound has brought clinical benefit but may be exposing patients to preventable infection risk.
CONCLUSIONS: Infection preventionists are well placed to act as major drivers of change based on their expertise and experience in the management of infection risk across facilities and health systems. They, along with clinicians responsible for probe use and reprocessing, should review practices relating to ultrasound in their facilities. Where practice does not comply with guidelines, policy and training should be updated to ensure patient safety.
Copyright © 2018 The Authors. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Disinfection; Doppler; Endocavitary; Interventional ultrasound; Probe; Reprocessing; Ultrasound gel

Mesh:

Year:  2018        PMID: 29866632     DOI: 10.1016/j.ajic.2018.03.025

Source DB:  PubMed          Journal:  Am J Infect Control        ISSN: 0196-6553            Impact factor:   2.918


  2 in total

Review 1.  Operation of ultrasonography services in a dedicated paediatric hospital and a university hospital in Greece under the COVID-19 pandemic.

Authors:  Maria Raissaki; Marina Vakaki; Alexandros Kotziamanis; Efthymia Alexopoulou; Chrisoula Koumanidou; Apostolos Karantanas
Journal:  Pediatr Radiol       Date:  2020-07-07

2.  Outbreak of Acinetobacter baumannii associated with extrinsic contamination of ultrasound gel in a tertiary centre burn unit.

Authors:  Kruti J Yagnik; Gautam Kalyatanda; Anthony P Cannella; Lennox K Archibald
Journal:  Infect Prev Pract       Date:  2019-06-27