Literature DB >> 29863635

Wave Intensity Analysis of Right Ventricular Function during Pulsed Operation of Rotary Left Ventricular Assist Devices.

J Christopher Bouwmeester1,2, Jiheum Park2, John Valdovinos2,3, Pramod Bonde4.   

Abstract

Changing the speed of left ventricular assist devices (LVADs) cyclically may be useful to restore aortic pulsatility; however, the effects of this pulsation on right ventricular (RV) function are unknown. This study investigates the effects of direct ventricular interaction by quantifying the amount of wave energy created by RV contraction when axial and centrifugal LVADs are used to assist the left ventricle. In 4 anesthetized pigs, pressure and flow were measured in the main pulmonary artery and wave intensity analysis was used to identify and quantify the energy of waves created by the RV. The axial pump depressed the intensity of waves created by RV contraction compared with the centrifugal pump. In both pump designs, there were only minor and variable differences between the continuous and pulsed operation on RV function. The axial pump causes the RV to contract with less energy compared with a centrifugal design. Diminishing the ability of the RV to produce less energy translates to less pressure and flow produced, which may lead to LVAD-induced RV failure. The effects of pulsed LVAD operation on the RV appear to be minimal during acute observation of healthy hearts. Further study is necessary to uncover the effects of other modes of speed modulation with healthy and unhealthy hearts to determine if pulsed operation will benefit patients by reducing LVAD complications.

Entities:  

Mesh:

Year:  2019        PMID: 29863635      PMCID: PMC6335194          DOI: 10.1097/MAT.0000000000000835

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  31 in total

1.  Is severe right ventricular failure in left ventricular assist device recipients a risk factor for unsuccessful bridging to transplant and post-transplant mortality.

Authors:  Jeffrey A Morgan; Ranjit John; Brian J Lee; Mehmet C Oz; Yoshifumi Naka
Journal:  Ann Thorac Surg       Date:  2004-03       Impact factor: 4.330

2.  Variable open-end wave reflection in the pulmonary arteries of anesthetized sheep.

Authors:  Nathan Dwyer; Ah Chot Yong; David Kilpatrick
Journal:  J Physiol Sci       Date:  2011-11-20       Impact factor: 2.781

3.  Right heart failure after left ventricular assist device implantation in patients with chronic congestive heart failure.

Authors:  Nicholas C Dang; Veli K Topkara; Michelle Mercando; Joy Kay; Kurt H Kruger; Michael S Aboodi; Mehmet C Oz; Yoshifumi Naka
Journal:  J Heart Lung Transplant       Date:  2005-12-09       Impact factor: 10.247

4.  Novel control system to prevent right ventricular failure induced by rotary blood pump.

Authors:  Mamoru Arakawa; Takashi Nishimura; Yoshiaki Takewa; Akihide Umeki; Masahiko Ando; Yuichiro Kishimoto; Yutaka Fujii; Shunei Kyo; Hideo Adachi; Eisuke Tatsumi
Journal:  J Artif Organs       Date:  2014-02-07       Impact factor: 1.731

5.  Electrocardiogram-synchronized rotational speed change mode in rotary pumps could improve pulsatility.

Authors:  Masahiko Ando; Takashi Nishimura; Yoshiaki Takewa; Kenji Yamazaki; Shunei Kyo; Minoru Ono; Tomonori Tsukiya; Toshihide Mizuno; Yoshiyuki Taenaka; Eisuke Tatsumi
Journal:  Artif Organs       Date:  2011-05-25       Impact factor: 3.094

6.  What stops the flow of blood from the heart?

Authors:  K H Parker; C J Jones; J R Dawson; D G Gibson
Journal:  Heart Vessels       Date:  1988       Impact factor: 2.037

7.  Chronic effect of left ventricular assist pumping on right ventricular function.

Authors:  S Fukuda; H Takano; Y Taenaka; T Nakatani; H Noda; M Kinoshita; T Akutsu; A Takeuchi
Journal:  ASAIO Trans       Date:  1988 Jul-Sep

8.  Prognosis of right ventricular failure in patients with left ventricular assist device based on decision tree with SMOTE.

Authors:  Yajuan Wang; Marc Simon; Pramod Bonde; Bronwyn U Harris; Jeffrey J Teuteberg; Robert L Kormos; James F Antaki
Journal:  IEEE Trans Inf Technol Biomed       Date:  2012-02-10

9.  Model for end-stage liver disease score predicts adverse events related to ventricular assist device therapy.

Authors:  Pramod Bonde; Natalie C Ku; Elizabeth A Genovese; Christian A Bermudez; Jay K Bhama; Maria M Ciarleglio; Xiangyu Cong; Jeffrey J Teuteberg; Robert L Kormos
Journal:  Ann Thorac Surg       Date:  2012-04-04       Impact factor: 4.330

Review 10.  Axial and centrifugal continuous-flow rotary pumps: a translation from pump mechanics to clinical practice.

Authors:  Nader Moazami; Kiyotaka Fukamachi; Mariko Kobayashi; Nicholas G Smedira; Katherine J Hoercher; Alex Massiello; Sangjin Lee; David J Horvath; Randall C Starling
Journal:  J Heart Lung Transplant       Date:  2013-01       Impact factor: 10.247

View more
  1 in total

1.  Electrical power to run ventricular assist devices using the Free-range Resonant Electrical Energy Delivery system.

Authors:  Benjamin H Waters; Jiheum Park; J Christopher Bouwmeester; John Valdovinos; Arnar Geirsson; Alanson P Sample; Joshua R Smith; Pramod Bonde
Journal:  J Heart Lung Transplant       Date:  2018-08-11       Impact factor: 10.247

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.