Towards offline PET monitoring of proton therapy at MedAustron.

Heide Meißner1, Hermann Fuchs2, Albert Hirtl1, Christian Reschl3, Markus Stock3.   

Abstract

The characteristic depth-dose profile of protons traveling through material is the main advantage of proton therapy over conventional radiotherapy with photons or electrons. However, uncertainties regarding the range of the protons in human tissue prevent to exploit the full potential of proton therapy. Therefore, a non-invasive in-vivo dose monitoring is desired. At the ion beam center MedAustron in Wiener Neustadt/Austria, patient treatment with proton beams started in December 2016. A PET/CT is available in close vicinity of the treatment rooms, exclusively dedicated to offline PET monitoring directly after the therapeutic irradiation. Preparations for a patient study include workflow tests under realistic clinical conditions using two different phantoms, irradiated with protons prior to the scan in the PET/CT. GATE simulations of the C-11 production are used as basis for the prediction of the PET measurement. We present results from the workflow tests in comparison with simulation results, and by this, we demonstrate the applicability of the PET monitoring at the MedAustron facility.
Copyright © 2018. Published by Elsevier GmbH.

Entities:  

Keywords:  Dose verification; GATE simulations; Positron emission tomography; Proton therapy

Mesh:

Year:  2018        PMID: 29858131     DOI: 10.1016/j.zemedi.2018.05.003

Source DB:  PubMed          Journal:  Z Med Phys        ISSN: 0939-3889            Impact factor:   4.820


  2 in total

1.  The effects of Compton camera data acquisition and readout timing on PG imaging for proton range verification.

Authors:  Jerimy C Polf; Paul Maggi; Rajesh Panthi; Stephen Peterson; Dennis Mackin; Sam Beddar
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2021-02-05

2.  Secondary Particle Interactions in a Compton Camera Designed for in vivo Range Verification of Proton Therapy.

Authors:  Rajesh Panthi; Paul Maggi; Stephen Peterson; Dennis Mackin; Jerimy Polf; Sam Beddar
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2020-10-12
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.