Literature DB >> 29852893

High Throughput Quantification of Quaternary Ammonium Cations in Food Simulants by Flow-Injection Mass Spectrometry.

Longjiao Yu1, Sargun Malik1, Timothy V Duncan2, Joseph E Jablonski2.   

Abstract

Background: A flow-injection MS (FI/MS) method was evaluated for the quantitation of quaternary ammonium cations (QACs) in simple food simulants.
Methods: The calibration standard was dimethyldioctadecyl ammonium ion (C18-C18), and the internal standard was benzyldimethylhexadecyl (BDMHD) ammonium ion. Calibration standards based on the C18-C18 ion were prepared in ethanol with a range of 5 to 500 ppb and contained 100 ppb BDMHD. The mobile phase was 90 + 10 (v/v) acetonitrile-5 mM aqueous ammonium acetate and flowed directly into an electrospray source of the mass spectrometer. Detection was accomplished by single ion recording (SIR) in positive mode.
Results: Calibration curves were linear with coefficients of determination above 0.995, and the LOQ was 5 ppb. Recoveries of four QACs derived from Arquad 2HT-75, a commercially available surfactant, were measured in common food simulants: ethanol, water, 10% (v/v) ethanol in water, and 3% (v/v) aqueous acetic acid. A solvent exchange procedure was employed for the three aqueous solvents, which included complete evaporation of the sample followed by reconstitution in ethanol prior to injection. The solvent exchange method minimized losses because of QAC adsorption on glass surfaces. Recoveries ranged from 74.4 ± 4.0 to 106.7 ± 6.6% for the two most abundant Arquad 2HT-75 component cations, dimethyldioctadecyl ammonium and dimethyloctadecyl-hexadecyl ammonium. Conclusions: This method is suitable to quantify trace levels of QACs in food simulants as part of exposure evaluations related to their use in emerging food contact materials.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29852893     DOI: 10.5740/jaoacint.18-0091

Source DB:  PubMed          Journal:  J AOAC Int        ISSN: 1060-3271            Impact factor:   1.913


  2 in total

1.  Migration of Quaternary Ammonium Cations from Exfoliated Clay/Low-Density Polyethylene Nanocomposites into Food Simulants.

Authors:  Joseph E Jablonski; Longjiao Yu; Sargun Malik; Ashutosh Sharma; Akhil Bajaj; SuriyaPrakaash L Balasubramaniam; Reiner Bleher; Rebecca G Weiner; Timothy V Duncan
Journal:  ACS Omega       Date:  2019-08-07

2.  The first detection of quaternary ammonium compounds in breast milk: Implications for early-life exposure.

Authors:  Guomao Zheng; Erika Schreder; Sheela Sathyanarayana; Amina Salamova
Journal:  J Expo Sci Environ Epidemiol       Date:  2022-04-18       Impact factor: 6.371

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.