Ordinary Cannulated Compression Screws or Headless Cannulated Compression Screws? A Synthetic Bone Biomechanical Research in the Internal Fixation of Vertical Femoral Neck Fracture.

Baokun Zhang1, Jingwen Liu1, Wei Zhang1.   

Abstract

PURPOSE: The purpose of this study is to verify whether the headless cannulated compression screw (HCCS) has higher biomechanical stability than the ordinary cannulated compression screw (OCCS) in the treatment of vertical femoral neck fractures.
MATERIALS AND METHODS: 30 synthetic femur models were equally divided into 2 groups, with 50°, 60°, and 70° Pauwels angle of femoral neck fracture, under 3D printed guiding plates and C-arm fluoroscopic guidance. The femur molds were fixed with three parallel OCCSs as OCCS group and three parallel HCCSs as HCCS group. All specimens were tested for compressive strength and maximum load to failure with a loading rate of 2 mm/min.
RESULTS: The result showed that there was no significant difference with the compressive strength in the Pauwels angle of 50° and 60°. However, we observed that the maximum load to failure with the Pauwels angle of 50°, 60°, and 70° and the compressive strength with 70° of HCCS group showed better performance than the OCCS group.
CONCLUSION: HCCS performs with better biomechanical stability than OCCS in the treatment of vertical femoral neck fracture, especially with the Pauwels angle of 70°.

Entities:  

Mesh:

Year:  2018        PMID: 29850523      PMCID: PMC5925079          DOI: 10.1155/2018/4898301

Source DB:  PubMed          Journal:  Biomed Res Int            Impact factor:   3.411


1. Introduction

Femoral neck fracture in young adults is usually the result of polytrauma and high-energy injuries, for which accurate reduction and stable fixation are necessary [1, 2]. Orthopedic surgeons should choose the most effective and affordable implant to deal with this kind of fracture. Comparing with other internal fixation implants, cannulated compression screws (CCSs) have been reported for their particular superiority with the treatment of femoral neck fracture, such as less damage of soft tissue, less blood loss, and being easy to operate, making them become one of the most common fixation devices [3]. However, with the increase of Pauwels angle, high rates of fixation failure while fixing with CCS, including femoral neck shortening, loose fixation, varus deformity, and fracture displacement, have been reported. Several factors might affect the stability between fracture fragments while using the CCS in treating femoral neck fracture such as the direction of the screws, the number of the screws, the position of the screws, the configuration of the screws, and the type of the screws. It has been proved that there is no difference in compressive strength among screws placement angle [4]. There are many researches trying to figure out the best choice of the number of screws but it still remains controversial. Several studies have shown evidence about the relationship between the position of cannulated screw and the effect [5-7], and the triangular or inverted triangular configuration performed with better strength and stability of fixation [8]. Then does the type of CCS have effect on the stability of femoral neck fracture? There are several literatures that can be searched but only one literature focused on the length of thread, and no biomechanical research was conducted [9]. HCCS has been introduced for the treatment of femoral neck fractures in recent years, functions of which are via whole thread and continuous compression with proximal lateral femoral cortex to femoral head. The purpose of this study is to verify the assumption that the HCCS has better biomechanical stability.

2. Materials and Methods

2.1. Specimen Preparation

30 same shaped left side synthetic femur models (ENOVO, China) were equally divided into two groups, and each group was further equally divided by three different Pauwels angles (50°, 60°, and 70°). To ensure the site of each screw at exact the same position, the first screw was driven up to the subchondral bone just beneath articular side of the femoral head; the second screw was placed beneath the first screw near to anterior cortex and the third one was placed near to posterior cortex; three screws form the configuration of standard triangle. We designed a guiding plate and made it with 3D printing technology. Fracture line was made with a medical pendulum saw just from the upper side of the lesser trochanter proximal to the superior femoral neck with the assistance of 3D printing guide plates, to avoid the appearance operation error (Figure 1). Firstly, under the C-arm fluoroscopic guidance and guiding plate, 3 parallel guide pins were placed into molds first. Then we removed the guiding plate and predrilled 3 insertion holes with 3.0 mm width drill bit through the guide pin, and after the accurate reduction of fracture, three 6.5 mm cannulated screws were inserted. Finally, the fracture was repaired with three CCSs. Group A (OCCS group): three parallel OCCSs (Stryker Co.) of 6.5 mm width placed with the configuration of triangle; Group B (CCCS group): three parallel HCCSs (Acumed Co.) placed with the configuration of triangle (Figure 2).
Figure 1

The design of guiding plates made with 3D printing technology.

Figure 2

The two types of compression cannulated screw. (a) Headless compression cannulated screw. (b) Ordinary compression cannulated screw.

2.2. Biomechanical Testing

All tests were performed with axial compressive loading with an Instron test system (Instron, Norwood, MA, USA) which included a base, a pressure applicator, and a data analyzer. The distal femur was fixed with shaft adduction angle of 7° using dental powder to imitate the femur form in normal walking. A vertical force was put on the top of the femoral head at the loading rate of 2 mm/min. The failure load was defined as the marked decrease followed by the maximum load or the fragments displaced with 5 mm [10]. Two magnets were placed on the proximal and distal fragments, respectively, to record the displacement of the two fragments (Figure 3).
Figure 3

The preparation of the models. (a, b) The guiding plates were used to obtain the accurate angle of fracture line and ideal placement of screws. (c) Two magnets were placed on the proximal and distal fragments, respectively, to record the displacement of the two fragments.

2.3. Statistical Analysis

The analysis was performed with the use of SPSS software (SPSS Version 20; SPSS Inc., Chicago, IL, USA). To detect the differences of compressive strength and failure load, the two groups were performed using t-test with significance set at P < 0.05.

3. Results

The testing results of compressive strength are showed in Table 1. We found that in our vertical fracture models the compressive strength of HCCS group performed better than the OCCS group with the Pauwels angle of 70° (109.03 ± 7.89 versus 128.58 ± 12.24, P = 0.019), but there is no statistical significance between two groups with 50° (177.58 ± 25.74 versus 214.08 ± 18.62, P = 0.133) and 60° (137.54 ± 32.57 versus 135.96 ± 43.52, P = 0.721). And the maximum failure load of the two groups is performed in Table 2. The results clearly indicate that the maximum load to failure in the HCCS group performed significantly better than that in OCCS group (Figure 4).
Table 1

The results of compressive strength in two groups.

Pauwels angle506070
Compressive strength (N)

Group A (OCCS)
Mean177.58137.54109.03
Minimum135.00114.4391.14
Maximum195.30193.84120.35
SD25.7432.577.89
Group B (HCCS)
Mean214.08135.96128.58
Minimum192.0977.89113.40
Maximum240.47199.69153.21
SD18.6243.5212.24
P 0.1330.7210.019
Table 2

The results of maximum load to failure in two groups.

Pauwels angle506070
Maximum failure load (N)

Group A (OCCS)
Mean691.56437.05312.06
Minimum567.81391.27246.02
Maximum750.39530.00467.81
SD72.0255.9789.64
Group B (HCCS)
Mean1001.80660.05468.83
Minimum809.77508.98324.34
Maximum1202.56772.21521.18
SD151.88104.1682.02
P 0.0220.0260.018
Figure 4

The results of the biomechanical study with statistical significance. (a) The compressive strength with Pauwels angle of 70°. (b, c, d) The maximum load to failure with Pauwels angle of 50°, 60°, and 70°.

4. Discussion

The incidence of femoral neck fracture has increased rapidly in recent years because of the acceleration of ageing with the population process. Different from the old, most of the young patients suffered from high-energy injury directly to the femoral neck, for whom choosing an appropriate device is significant. In recent years, the use of new methods to treat femoral neck fracture has been reported a lot. For example, the study of Samsami et al. [11] on vertical femoral neck fracture of young people compared the CCS with proximal femoral locking plate (PFLP) and the combination of DHS and an antirotation screw. The result revealed that the latter had better resistance of rotation shear force. Further, a new type of femoral neck locking plate (FNLP) consisting of a locking plate and five screws showed satisfactory biomechanical results in Pauwels type III femoral neck fracture. Compared with DHS, the new type of FNLP performed better in the biomechanical stability and it also could reduce the incidence of bone nonunion effectively by conduct force to the femoral neck dispersedly via the multiaperture screw system [10]. In Zhu et al. [12] follow-up study including 74 patients with femoral neck fracture, they introduced a new treatment for femoral neck fracture with percutaneous compression plate. The results showed that 98.5% of patients had good prognosis and could walk unaided again. Only two of them appeared with avascular necrosis of femoral head and delayed union, respectively. Although there are so many choices for treatment, none of them has been proven to be overwhelmingly superior. The use of OCCS in the treatment of femoral neck fracture achieved great success [13]. A 17-item survey showed that for the undisplaced fracture near 80% surgeons prefer using OCCS [14]. In the retrospective study of 59 patients in a single institution, 4/5 of them treated with OCCS showed good results at last [15]. However, for the more vertical fracture, the rate of failure and complication has risen obviously because of the increase of shear force [16, 17]. In our study, the compressive strength of OCCS of the Pauwels angle of 50°, 60°, and 70° is 177.58 ± 25.74 N/mm, 137.54 ± 32.57 N/mm, and 109.03 ± 7.89 N/mm, respectively. And the maximum load to failure is 691.56 ± 72.02 N, 437.05 ± 55.97, and 312.06 ± 89.64 N. It appeared that with the increase of Pauwels angle the axial stiffness and the maximum load to failure decrease obviously. The HCCS has been introduced as a reliable choice of internal fixation in recent years. The diameter of HCCS becomes larger from screw tip to tail, and the screw pitch becomes smaller to make the screw tip goes faster than the tail while going into the bone, and it forms compression between fractures. The design of the screw thread increases the contact between the screw and the bone, and the conical shape makes it possible to obtain greater holding force, pull-out strength, and shear strength. Thereby it can increase the compression in joint and create an immediate stability to help early mobilization. Dodds et al. [18] found that long thread screws provided optimal fixation when used for the scaphoid fracture. In the retrospective study with 41 distal ulna patients with rheumatoid arthritis, the patients with the treatment of the HCCS showed better stability and higher rate of bone union, compared to those treated with OCCS [19]. As for the treatment of metacarpal neck fractures, the HCCS also shows its advantages in earlier mobilization than those fixed with K-wires [20]. Moreover, in the case reported by Borse et al., they introduced a method using two HCCS in the treatment of Hoffa fracture and have achieved satisfying result [21]. And the method gained affirmation and improvement in the prospective study of Li et al. [22]; eight Chinese patients suffering isolated Hoffa fractures were treated with HCCS combined with back buttress plate; the result is inspiring with all fractures healed clinically and the range of motion of the knee joint became better. With all the successful use, the HCCS has become an effective internal fixation device. In addition, the compressive strength with the Pauwels angle of 70° (128.58 ± 12.24 N/mm) also showed satisfactory results in our biomechanical study. And the load of failure showed good results as well; with 50°, 60°, and 70° Pauwels angle, the loads were 1001.80 ± 151.88 N, 660.05 ± 104.16 N, and 468.83 ± 82.02 N, respectively. We strongly proved the advantages of HCCS in treatment of the vertical femoral neck fracture. Especially with the Pauwels angle of 70°, no matter the axial stiffness (128.58 ± 12.24 versus 109.03 ± 7.89, P = 0.019) or the maximum load to failure (468.83 ± 82.02 N versus 312.06 ± 89.64 N, P = 0.018), the superiority of biomechanical stability of HCCS is obvious. However, there are several limitations for this experiment. Firstly, the bone is synthetic femur model rather than cadaveric femur bone. Secondly, limited to implants supply, we have failed to consider the length of thread purchase and screw arm of the OCCS, which might affect the results.

5. Conclusion

In summary, with the vertical femoral neck fracture especially Pauwels angle of 70°, HCCS performs with much better biomechanical stability than OCCS. And HCCS can be introduced a better implant than OCCS while treating the vertical femoral neck fracture. However, further investigation with clinical research is needed in the future.
  22 in total

1.  Optimum configuration of cannulated hip screws for the fixation of intracapsular hip fractures: a biomechanical study.

Authors:  V T Selvan; M J Oakley; A Rangan; M K Al-Lami
Journal:  Injury       Date:  2004-02       Impact factor: 2.586

2.  Treatment of femoral neck fracture with percutaneous compression plate: preliminary results in 74 patients.

Authors:  Feng Zhu; Gang Liu; Hong-guo Shao; Yi-jun Wang; Rong-qun Li; Hui-lin Yang; De-chun Geng; Yao-zeng Xu
Journal:  Orthop Surg       Date:  2015-05       Impact factor: 2.071

Review 3.  Biomechanical rationale for implant choices in femoral neck fracture fixation in the non-elderly.

Authors:  Michalis Panteli; Paul Rodham; Peter V Giannoudis
Journal:  Injury       Date:  2015-01-03       Impact factor: 2.586

4.  Headless Compression Screw Versus Kirschner Wire Fixation for Metacarpal Neck Fractures: A Biomechanical Study.

Authors:  Daniel M Avery; Stephen Klinge; Felix Dyrna; Leo Pauzenberger; David Lam; Mark Cote; Jessica DiVenere; Elifho Obopilwe; Augustus Mazzocca; Craig Rodner
Journal:  J Hand Surg Am       Date:  2017-03-27       Impact factor: 2.230

5.  Sauvé-Kapandji procedure with headless compression screw in patients with rheumatoid arthritis.

Authors:  Atsushi Maeda; Taku Suzuki; Masaki Hasegawa; Takashi Kuroiwa; Kanae Shizu; Katsuhiko Hayakawa; Takashi Tsuji; Katsuji Suzuki; Harumoto Yamada
Journal:  Mod Rheumatol       Date:  2017-04-11       Impact factor: 3.023

6.  Treatment of femoral neck fractures in young adults.

Authors:  Thuan V Ly; Marc F Swiontkowski
Journal:  Instr Course Lect       Date:  2009

7.  Hoffa fracture: fixation using headless compression screws.

Authors:  V Borse; J Hahnel; A Cohen
Journal:  Eur J Trauma Emerg Surg       Date:  2010-04-27       Impact factor: 3.693

8.  Internal fixation with headless compression screws and back buttress plate for treatment of old Hoffa fracture.

Authors:  Li Min; Chong-Qi Tu; Guang-Lin Wang; Yue Fang; Hong Duan; Lei Liu; Hui Zhang
Journal:  Chin J Traumatol       Date:  2014-04-01

9.  Quality of reduction and cortical screw support in femoral neck fractures. An analysis of 72 fractures with a new computerized measuring method.

Authors:  S Lindequist; H Törnkvist
Journal:  J Orthop Trauma       Date:  1995-06       Impact factor: 2.512

10.  The vertical hip fracture - a treatment challenge. A cohort study with an up to 9 year follow-up of 137 consecutive hips treated with sliding hip screw and antirotation screw.

Authors:  Anders Enocson; Lasse J Lapidus
Journal:  BMC Musculoskelet Disord       Date:  2012-09-13       Impact factor: 2.362

View more
  3 in total

Review 1.  Surgical treatment options for femoral neck fractures in the elderly.

Authors:  Boris A Zelle; Luis M Salazar; Shain L Howard; Kisan Parikh; Hans-Christoph Pape
Journal:  Int Orthop       Date:  2022-02-21       Impact factor: 3.075

2.  Comparison of Ordinary Cannulated Compression Screw and Double-Head Cannulated Compression Screw Fixation in Vertical Femoral Neck Fractures.

Authors:  Yuelei Zhang; Chao Yan; Lecheng Zhang; Wei Zhang; Gang Wang
Journal:  Biomed Res Int       Date:  2020-12-30       Impact factor: 3.411

3.  Comparison between headless cannulated screws and partially threaded screws in femoral neck fracture treatment: a retrospective cohort study.

Authors:  Yilin Wang; Na Han; Dianying Zhang; Peixun Zhang; Baoguo Jiang
Journal:  Sci Rep       Date:  2022-02-02       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.