| Literature DB >> 29845100 |
Yi-Ching Kuo1, Chih-Kung Lee1,2,3, Chih-Ting Lin3.
Abstract
This article presents a new sensitivity-improved electrochemical measurement architecture for cardiovascular disease (CVD) diagnosis by detecting CVD biomarkers, S100 beta protein and C-reactive protein (CRP). The new architecture includes a design for a new electrochemical measurement set-up, which improves the reaction conditions of chemical and biological molecules and incorporates a newly biochip design. With the new architecture, electrochemical measurement experiments were undertaken. The results obtained are related to the research article entitled "Improving sensitivity of a miniaturized label-free electrochemical biosensor using zigzag electrodes" [1].Entities:
Year: 2018 PMID: 29845100 PMCID: PMC5966522 DOI: 10.1016/j.dib.2018.01.096
Source DB: PubMed Journal: Data Brief ISSN: 2352-3409
Fig. 1Impedance data of the linker/glutaraldehyde/antibody/ETA&BSA modified electrode in a 1× PBS solution with 1 mM Fe(CN)63−/4− after addition of the target protein at different concentrations: (a) 4-ATP, antiS100, and S100 beta protein in Nyquist form of interdigitate biochip, (b) 4-ATP, antiS100, and S100 beta protein in Nyquist form of interdigitate-semicircle biochip, (c) 4-ATP, antiS100, and S100 beta protein in Nyquist form of interdigitate-zigzag biochip, (d) 4-ATP, antiCRP, and CRP in Nyquist form of interdigitate-zigzag biochip, (e) cysteamine, antiS100, and S100 beta protein in Nyquist form of interdigitate-zigzag biochip, and (f) cysteamine, antiCRP, and CRP in Nyquist form of interdigitate-zigzag biochip.
Fig. 2Electrochemical measurement set-up used in this article.
Fig. 3SPR data for antiS100-S100 interaction: (a) during activation, (b) during immobilization, (c) during blocking, (d) during regeneration (Note: red dashed line represents the signal after antigen binding and the blue dotted line represents the signal after regeneration), and (e) RU during regeneration in item (d).
Fig. 4SPR measurement data for the antiCRP-CRP interaction: (a) during activation, (b) during immobilization, (c) during blocking, (d) during regeneration (Note: red dashed line represents the signal after antigen binding and the blue dotted line represents the signal after regeneration), and (e) RU during regeneration in item (d).
| Subject area | Biosensors |
| More specific subject area | Electrochemistry |
| Type of data | Graphs, figures, tables, text files |
| How data was acquired | Electrochemical analyzer (CHI614D, CH Instrument), Biacore T200 (GE Healthcare) |
| Data format | Raw |
| Experimental factors | The proteins at different concentrations were captured by the antibodies based on a 4-ATP or cysteamine modified sensing surface. |
| Experimental features | Development of the process for antibody-antigen interactions. Detection of two kinds of CVD biomarkers, S100 beta proteins and C-reactive proteins, taken as electrochemical impedance measurements. |
| Data source location | National Taiwan University in Taipei, Taiwan |
| Data accessibility | Data is with this article |
| Related research article | Y.-C. Kuo, et al., "Improving sensitivity of a miniaturized label-free electrochemical biosensor using zigzag electrodes," Biosensors and Bioelectronics, 2018. |