Abstract
An accurate, sensitive, precise and simple method was developed utilizing Gas Chromatography for the quantitative analysis of benzyl isothicyanate in Siwak extract and dental care herbal products claimed to contain Siwak. Rtx (30.0 m × 0.25 mm ID, 25 µm thickness) column was used and helium as carrier gas at a flow rate of 0.74 mL/min. The retention time of standard benzyl isothicyanate was 13.470 min under the described conditions. Linear regression data analysis indicated a good linear relationship between peak height measurement and concentration of benzyl isothiocyanate in the range of 10-50 µg/ml (R2 = 0.9971). The regression equation was y = 11,471x. The developed GC method was subjected to validation requirements set by the ICH for precision, accuracy, and robustness. The entitled GC analyses expected to be valuable for the determination of benzyl isothiocyanate in Siwak extracts and other formulations containing Siwak extract. The amount of benzyl isothiocyanate reflects the efficacy of the products.Entities:
Keywords: Benzyl isothiocyanate; GC; ICH guidelines; Quantitative; S. persica
Year: 2018 PMID: 29844716 PMCID: PMC5961752 DOI: 10.1016/j.jsps.2018.02.013
Source DB: PubMed Journal: Saudi Pharm J ISSN: 1319-0164 Impact factor: 4.330
Fig. 1Structure of benzyl isothiocyanate.
Test Results for calibration curve.
| Concentration (µg/ml) | Peak area |
|---|---|
| 10 | 130,047 |
| 20 | 232,873 |
| 30 | 349,368 |
| 40 | 449,821 |
| 50 | 573,173 |
Fig. 2GC calibration plot of standard benzyl isothiocyanate.
Fig. 3GC chromatogram of standard benzyl isothiocyanate.
Fig. 4GC chromatogram of S. persica roots ethanol extract.
Linear regression data for the calibration curve of benzyl isothiocyanate (n = 6).
| Linearity range (µg/spot) | 10–50 µg/ml Rang |
|---|---|
| Regression equation | Y = 11,472x |
| Correlation coefficient | 0.9983 |
| Slope ± SD | 11,472 ± 455.53 |
| Intercept ± SD | 0 ± 455.53 |
| LOQ | 2.03 µg/ml |
| LOD | 0.67 µg/ml |
LOQ = 10 /S.
LOD = 3.3 /S.
Accuracy of the proposed method (n = 6).
| Excess drug added to analyte (%) | Theoretical content (µg) | Conc. found (µg) ± SD | % Recovery | % RSD |
|---|---|---|---|---|
| 50 | 15 | 14.47 ± 0.17 | 99.35 | 0.28 |
| 100 | 20 | 19.13 ± 0.12 | 99.43 | 0.10 |
| 150 | 35 | 35.53 ± 0.05 | 100.15 | 0.01 |
Accuracy = Mean overall recovery = 100.23 ± 0.13.
Precision of the proposed method.
| Conc. (µg) | Repeatability (Intraday precision) | Intermediate precision (Interday) | ||
|---|---|---|---|---|
| Benzyl isothiocyabnate found (n = 6) | % RSD | Benzyl isothiocyabnate found (n = 6) | % RSD | |
| 12 | 11.82 | 2.14 | 12.46 | 2.20 |
| 17 | 18.08 | 1.19 | 17.79 | 1.14 |
| 23 | 22.81 | 0.85 | 24.08 | 0.65 |
| 32 | 32.98 | 0.17 | 31.89 | 0.27 |
| 43 | 43.11 | 0.32 | 43.34 | 0.31 |
| Av = 0.93% | Av = 0.89% | |||
Robustness of the proposed GC method.
| Conc. (µg) | Variations | Results | |||
|---|---|---|---|---|---|
| 40 | Pressure | Initial T | Flow rate | Resolution | RSD (%) |
| 41.5 | 80 | 0.74 | 13.47 | 0.06 | |
| 46.5 | 90 | 0.76 | 11.55 | 0.21 | |
| 36.5 | 70 | 0.72 | 15.44 | 0.34 | |