Significance of the outer diameter of an endotracheal tube: a lesser-known parameter.

Bhavna Gupta1, Lalit Gupta1.   

Abstract

Entities:  

Mesh:

Year:  2018        PMID: 29843505      PMCID: PMC6369347          DOI: 10.4097/kja.d.18.00056

Source DB:  PubMed          Journal:  Korean J Anesthesiol        ISSN: 2005-6419


× No keyword cloud information.
The diameter of catheters and tubings are typically measured by the French gauge system. Improper interpretation of these standards may lead to confusion between the users and catheter manufacturers. The outer diameter (OD) and inner diameter (ID) are generally measured in inches or millimeters. The size of various tracheal tubes is usually expressed as ID, unless specified otherwise. Though the ID measurements may be constant, both OD and ID may show a difference of 2–4 mm depending on different manufacturers. For selecting an appropriate tracheal tube, the American society for testing and materials does not include any recommendations for OD measurement, as opposed to the recommendations available for ID measurements [1]. The OD and ID measurements of endotracheal tubes are summarized in Table 1.
Table 1.

Recommended ID and OD of Endotracheal Tubes

AgeWeightEndotracheal tube diameter (ID) (mm)Endotracheal tube diameter (OD) (mm)
Neonates0.7–1.5 kg2.53.4
1.6–3.2 kg34.2
Infants0–3 months3.54.8
3–6 months45.4
6–12 months4.56.2
Children2–3 yr56.8
4–5 yr5.57.4
6–7 yr68.2
8–9 yr6.58.8
10–11 yr79.6
12–13 yr7.510.2
14–15 yr811
AdultsFemale7.510.2
811
Male8.511.6
912.2
9.513

ID: inner diameter, OD: outer diameter.

Armored or flexo-metallic endotracheal tubes have a slightly thicker wall. The difference between OD and a standard ID is greater in an armored tube as compared to a regular endotracheal tube (0.2–0.3 mm). The size of a double lumen tube is decided by the size of the bronchus; this corresponds to the external diameter equivalent of a French catheter gauge, defined as 3 times that of the ID, hence the OD here is more important. The difference in the ODs of Univent and double lumen tubes as compared to that of a regular endotracheal tube is 2–3 mm. In a study published by Al-Mazrou et al. [2], the internal tracheal diameter (ITD) was measured at the level of the cricoid, using magnetic resonance images, and compared to the OD of a utilized endotracheal tube (ETT). The OD of the best fit ETT was less than the ITD measured at the level of the cricoid by 0.1–1.7 mm, and it was concluded that the correlation of OD of the best fit ETT was strongerirrespective of weight or height. In another study by Bernet et al. [3], it was observed that the external diameter of pediatric tracheal tube cuffs could easily expand in vitro, to produce ODs of more than twice the age-corresponding ITD, when over-inflated with air. It has been observed that if the ID is too small, it may result in a smaller OD, which increases the cuff pressure required to create a seal in the trachea. A smaller ID tube also increases the resistance through the tube, making the airway clearance more difficult. However, a tube with a larger OD would be more difficult to pass through a stoma. The OD may be implicated in post extubation edema, sore throat, sub-glottis inflammation, long term stenosis, and other complications; however, research studies may be required to record this observation. The OD is also important to allow smooth passage of ETT through the supra-glottic airway devices. Additionally, attention should be paid to the OD when using an endotracheal tube of different material such as polyvinylchloride (PVC) or silicon, although the most commonly used ones are made up of PVC. Further, it becomes imperative to check the compatibility between the supra-glottic device and endotracheal tube in emergency intubation of the trachea. The OD of a tracheostomy tube should be about 2/3–3/4 of the tracheal diameter. A tube should not be wider than 10 mm in adult females and 11 mm in adult males, to minimize the trauma to the tracheal wall and avoid long term complications [4]. The ID influences the physiology of breathing and course of weaning off from the ventilator. In conclusion, one must ascertain both the ID as well as the OD of tracheal tubes and accordingly decide the size of tube especially in long lasting surgeries and in cases presenting higher chances of post-operative vocal cord edema.
  3 in total

1.  Outer diameter and shape of paediatric tracheal tube cuffs at higher inflation pressures.

Authors:  V Bernet; A Dullenkopf; P Maino; M Weiss
Journal:  Anaesthesia       Date:  2005-11       Impact factor: 6.955

2.  Comparison of tracheal tube cuff diameters with internal transverse diameters of the trachea in children.

Authors:  A Dullenkopf; O Kretschmar; W Knirsch; M Tomaske; M Hug; K Stutz; F Berger; M Weiss
Journal:  Acta Anaesthesiol Scand       Date:  2006-02       Impact factor: 2.105

3.  Comparison of the outer diameter of the 'best-fit' endotracheal tube with MRI-measured airway diameter at the cricoid level.

Authors:  Khalid A Al-Mazrou; Khalid M Abdullah; Riaz A Ansari; Mohamed E Abdelmeguid; Ahmed Turkistani
Journal:  Eur J Anaesthesiol       Date:  2009-09       Impact factor: 4.330

  3 in total
  3 in total

1.  Steroid eluting biocompatible stent for subglottic stenosis.

Authors:  Brian Manzi; Rohan Shirwaiker; Anil Gungor; Paul Weinberger
Journal:  Eur Arch Otorhinolaryngol       Date:  2021-01-02       Impact factor: 2.503

2.  Lung isolation for lobectomy in an elderly, post radiation fibrosis of a difficult airway-pediatric double lumen tube and pediatric ureteroscope as rescue devices.

Authors:  Bhavna Gupta; Atif Khan; Deyashinee Ghosh
Journal:  Saudi J Anaesth       Date:  2020-03-05

Review 3.  Challenge of Lung Isolation in Patients with Vocal Cord Implants.

Authors:  Islam Mohammad Shehata; Amir Elhassan; Behrooz Zaman; Omar Viswanath
Journal:  Anesth Pain Med       Date:  2022-03-22
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.