Literature DB >> 29806402

[Treatment of unstable pelvic fractures by cannulated screw internal fixation with the assistance of three-dimensional printing insertion template].

Hongqi Yang1, Qing Lei2, Lihong Cai3, Feng Liu1, Weili Zhou1, Song Chen1, Li Chen1, Tangyou Liu1, Minghui Jiang1, Kang Wang1, Sishun Xiao1, Wenqian Liu1.   

Abstract

Objective: To evaluate the effectiveness of unstable pelvic fractures treated by cannulated screw internal fixation with the assistance of three-dimensional (3D) printing insertion template.
Methods: The clinical data of 10 patients who underwent surgical treatment for unstable pelvic fractures by cannulated screw internal fixation with the assistance of 3D printing insertion template between May 2015 and June 2016 were retrospectively analysed. There were 7 males and 3 females with an average age of 37.5 years (range, 20-58 years). The causes of injury included falling from height in 5 cases, crushing from heavy load in 1 case, and traffic accidents in 4 cases. The interval from injury to admission was 1-5 hours (mean, 3.1 hours). The fracture situation included 6 cases of sacral fracture, 1 case of right sacroiliac joint dislocation, and 3 cases of iliac bone fracture. There were 10 cases of superior and inferior pubic rami fracture, including 3 cases on the left side (2 cases of suprapubic fracture adjacent to symphysis pubis), 2 cases on the right side, and 5 cases on the bilateral. All fractures were classified according to the Tile system, there were 4 cases of type B2, 1 of type B3, 4 of type C1, and 1 of type C2. The radiological outcome was evaluated by Matta scale, and the positions of the iliosacral screw and superior pubic ramus screw were evaluated according to 3D reconstruction of CT postoperatively. The functional outcome was evaluated by Majeed function scale.
Results: The average time of each screw implantation was 30 minutes, and the average blood loss per screw incision was 50 mL. The time of implantation of each sacroiliac screw was 24-96 seconds (mean, 62 seconds), and the time of implantation of each suprapubic screw was 42-80 seconds (mean, 63.2 seconds). The hospitalization duration was 17-90 days (mean, 43.7 days). All incisions healed by first intention. All patients were followed up 12-22 months (mean, 15.6 months). The radiological outcome was excellent in 8 cases and good in 2 cases according to Matta scale; and 3D reconstruction of CT demonstrated that all the 9 iliosacral screws were placed as type Ⅰ, and all the 13 suprapubic ramus screws were placed as grade 0 on the first postoperative day. No complication such as neurovascular injury, screw back out or rupture, or secondary fracture displacement was observed during the follow-up. At 6 months after operation, the X-ray films showed good fracture healing in all the 10 patients. The functional outcome was excellent in 9 cases and good in 1 case according to Majeed scale at 1 year after operation. One patient sustained Tile C2 pelvic disruption complicated with L 5 nerve root injury achieved complete nervous functional recovery at last follow-up.
Conclusion: It has advantages of precise screw insertion and lower risk of neurovascular injury to treat unstable pelvic fractures by cannulated screw internal fixation with the assistance of 3D printing insertion template, which can be a good alternative for the treatment of unstable pelvic fractures.

Entities:  

Keywords:  Unstable pelvic fracture; insertion template; internal fixation; three-dimensional printing technology

Mesh:

Year:  2018        PMID: 29806402     DOI: 10.7507/1002-1892.201708059

Source DB:  PubMed          Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi        ISSN: 1002-1892


  5 in total

1.  A novel approach using 3-D printing in the Irish National Centre for pelvic and acetabular surgery.

Authors:  Colum Downey; Cathleen McCarrick; Christopher Fenelon; Evelyn P Murphy; Brendan J O'Daly; Michael Leonard
Journal:  Ir J Med Sci       Date:  2019-07-06       Impact factor: 1.568

2.  [Effectiveness analysis of three-dimensional printing assisted surgery for unstable pelvic fracture].

Authors:  Chao Wu; Jiayan Deng; Lun Tan; Haigang Hu; Dechao Yuan
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2019-04-15

3.  [Application of individualized transiliac crest nail-grafting guide plate in deep pelvic external fixator implantation].

Authors:  Cheng Tang; Bin Liang; Qingqiang Yao; Shuai Liu; Yan Xu; Liming Wang
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2019-09-15

Review 4.  [Research progress on minimally invasive treatment of anterior pelvic ring fracture].

Authors:  Jianchao Wu; Zhengwei Shi; Jianwei Zhou; Yixin Pan; Shaoping Li; Yuliang Wang
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2020-04-15

5.  Comparative study of sacroiliac screw placement guided by 3D-printed template technology and X-ray fluoroscopy.

Authors:  Wu Zhou; Tian Xia; Yi Liu; Faqi Cao; Mengfei Liu; Jing Liu; Bobin Mi; Liangcong Hu; Yuan Xiong; Guohui Liu
Journal:  Arch Orthop Trauma Surg       Date:  2019-05-24       Impact factor: 3.067

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.