| Literature DB >> 29803103 |
Fang Zhang1, Yupeng He1, Kena Fu1, Li Fu1, Bin Zhang2, Huaisheng Wang3, Guizheng Zou4.
Abstract
A dual-wavebands-resolved electrochemiluminescence (ECL) multiplexing immunoassay (MIA) was developed for simultaneously detecting alpha fetoprotein antigen (AFP) in greenish waveband with CdSe550 (λmax = 550 nm) nanocrystals (NCs) and carbohydrate antigen 125 (CA125) in near-infrared waveband with CdTe790 (λmax = 790 nm) NCs via one-pot ECL reaction, in which dichroic mirror works as a key part to reflect ECL from CdSe550 to one photomultiplier-tube (PMT) and transmit ECL from CdTe790 to the other PMT for dual-wavebands-resolved assay. The proposed ECL-MIA strategy was capable of simultaneously determining AFP with linearly response from 5 pg/mL to 5 ng/mL and limit of detection at 1 pg/mL, and CA125 with linearly response from 5 mU/mL to 1 U/mL and limit of detection at 1 mU/mL, with desired specificity and without obvious energy-transfer between ECL tags. The dichroic mirror assistant ECL setup is easy-to-assemble and convenient for the popularization of color-resolved multiplexing ECL assay.Entities:
Keywords: Dichroic mirror; Dual-wavebands-resolved; Electrochemiluminescence; Multiplexing immunoassay; Photomultiplier-tube
Mesh:
Substances:
Year: 2018 PMID: 29803103 DOI: 10.1016/j.bios.2018.05.006
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618