Literature DB >> 29794336

Regional lung ventilation and perfusion by electrical impedance tomography compared to single-photon emission computed tomography.

Benjamin Hentze1, Thomas Muders, Henning Luepschen, Enn Maripuu, Göran Hedenstierna, Christian Putensen, Marian Walter, Steffen Leonhardt.   

Abstract

OBJECTIVE: Electrical impedance tomography (EIT) is a noninvasive imaging modality that allows real-time monitoring of regional lung ventilation ([Formula: see text]) in intensive care patients at bedside. However, for improved guidance of ventilation therapy it would be beneficial to obtain regional ventilation-to-perfusion ratio ([Formula: see text]) by EIT. APPROACH: In order to further explore the feasibility, we first evaluate a model-based approach, based on semi-negative matrix factorization and a gamma-variate model, to extract regional lung perfusion ([Formula: see text]) from EIT measurements. Subsequently, a combined validation of both [Formula: see text] and [Formula: see text] measured by EIT against single-photon emission computed tomography (SPECT) is performed on data acquired as part of a porcine animal trial. Four pigs were ventilated at two different levels of positive end-expiratory pressure (PEEP 0 and 15 cm H2O, respectively) in randomized order. Repeated injections of an EIT contrast agent (NaCl 10%) and simultaneous SPECT measurements of [Formula: see text] (81mKr gas) and [Formula: see text] (99mTc-labeled albumin) were performed. MAIN
RESULTS: Both [Formula: see text] and [Formula: see text] from EIT and SPECT were compared by correlation analysis. Very strong (r 2  =  0.94 to 0.95) correlations were found for [Formula: see text] and [Formula: see text] in the dorsal-ventral direction at both PEEP levels. Moderate (r 2  =  0.36 to 0.46) and moderate to strong (r 2  =  0.61 to 0.82) correlations resulted for [Formula: see text] and [Formula: see text] in the right-left direction, respectively. SIGNIFICANCE: The results of combined validation indicate that monitoring of [Formula: see text] and [Formula: see text] by EIT is possible. However, care should be taken when trying to quantify [Formula: see text] by EIT, as imaging artefacts and model bias may void necessary spatial matching.

Entities:  

Mesh:

Year:  2018        PMID: 29794336     DOI: 10.1088/1361-6579/aac7ae

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  5 in total

1.  What's new in electrical impedance tomography.

Authors:  Tommaso Mauri; Alain Mercat; Giacomo Grasselli
Journal:  Intensive Care Med       Date:  2018-10-04       Impact factor: 17.440

Review 2.  Lung aeration, ventilation, and perfusion imaging.

Authors:  Lorenzo Ball; Gaetano Scaramuzzo; Jacob Herrmann; Maurizio Cereda
Journal:  Curr Opin Crit Care       Date:  2022-06-01       Impact factor: 3.359

3.  A Rotational Invariant Neural Network for Electrical Impedance Tomography Imaging without Reference Voltage: RF-REIM-NET.

Authors:  Jöran Rixen; Benedikt Eliasson; Benjamin Hentze; Thomas Muders; Christian Putensen; Steffen Leonhardt; Chuong Ngo
Journal:  Diagnostics (Basel)       Date:  2022-03-22

Review 4.  Electrical Impedance Tomography for Cardio-Pulmonary Monitoring.

Authors:  Christian Putensen; Benjamin Hentze; Stefan Muenster; Thomas Muders
Journal:  J Clin Med       Date:  2019-08-07       Impact factor: 4.241

Review 5.  Personalized mechanical ventilation in acute respiratory distress syndrome.

Authors:  Paolo Pelosi; Lorenzo Ball; Carmen S V Barbas; Rinaldo Bellomo; Karen E A Burns; Sharon Einav; Luciano Gattinoni; John G Laffey; John J Marini; Sheila N Myatra; Marcus J Schultz; Jean Louis Teboul; Patricia R M Rocco
Journal:  Crit Care       Date:  2021-07-16       Impact factor: 9.097

  5 in total