Treatment with a pipeline embolization device (PED) is widely accepted as an excellent option for patients harboring giant, wide neck, saccular intracranial aneurysms. It has been shown to be safe and very effective. We describe a previously unreported case of a large middle cerebral artery aneurysm, related to the carotid tip, that was treated with a PED and coils. The 6 month follow-up angiogram demonstrated complete occlusion of the aneurysm. However, the 9 month angiogram revealed recurrent filling of the aneurysm sac. To our knowledge, this is the first report to describe recanalization of an aneurysm totally occluded on the control angiography after treatment with a PED. Unanswered questions include the mechanism of recanalization and the need for additional control angiograms and MRI following a previous examination revealing total aneurysm occlusion. © Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2019. All rights reserved. No commercial use is permitted unless otherwise expressly granted.Entities:
Keywords: aneurysm; complication; flow diverter
Mesh:
Year: 2018 PMID: 29794160 DOI: 10.1136/neurintsurg-2018-013842.rep