Literature DB >> 29784162

Disinfection of the Goldman applanation tonometer: a systematic review.

Alex Ragan1, Stephanie L Cote2, John T Huang3.   

Abstract

OBJECTIVES: The Goldmann applanation tonometer is the gold standard for measuring intraocular pressure and a possible vector for the transmission of infectious diseases. The purpose of this study is to consolidate the evidence pertaining to the disinfection of Goldmann tonometers in an effort to facilitate an informed discussion regarding public policy in this important area.
DESIGN: Systematic review.
METHODS: An exhaustive literature review was undertaken to identify primary-level research that assesses the effectiveness of different agents used in the disinfection of Goldmann applanation tonometer prisms. Seven discrete databases were reviewed by 2 independent researchers, and a symmetrical screening process was used to identify and review all pertinent studies.
RESULTS: Our review identified only 19 primary-level studies that relate specifically to the disinfection of Goldmann tonometer prisms. These studies are largely heterogeneous with regard to the pathogens and disinfectants that were tested and the experimental protocols that were employed. Accordingly, definitive conclusions as to the optimal agent cannot be made on this basis. Furthermore, the results of our review reinforce previous suggestions that only adenovirus has been transmitted between patients via the Goldmann tonometer.
CONCLUSIONS: The present state of the literature does not permit a definitive conclusion regarding optimal disinfection agent for Goldmann applanation tonometer prisms. Further well-constructed studies are required to better delineate the effectiveness of disinfectants in the specific context of tonometer prisms.
Copyright © 2018 Canadian Ophthalmological Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 29784162     DOI: 10.1016/j.jcjo.2017.09.013

Source DB:  PubMed          Journal:  Can J Ophthalmol        ISSN: 0008-4182            Impact factor:   1.882


  1 in total

1.  The effectiveness of a three-step sterilization method for Goldmann tonometer prism: A cross-sectional study.

Authors:  Jinping Hu; Jun Liu; Jingting Xia; Jingyu Yan; Zheng Liu; Dan Ma
Journal:  Indian J Ophthalmol       Date:  2022-01       Impact factor: 1.848