Literature DB >> 29774550

Efficacy of vaporized hydrogen peroxide for repeated sterilization of a single-use single-incision laparoscopic surgery port.

Valery F Scharf1, Brian Dent1, Megan E Jacob1, Brandon Moore1.   

Abstract

OBJECTIVE: To determine the ability of vaporized hydrogen peroxide (VHP) to sterilize a single-use single-incision laparoscopic surgery port and its associated components after repeated simulated uses. STUDY
DESIGN: Prospective in vitro experimental study. SAMPLE POPULATION: Six single-use single-incision laparoscopic surgery ports with associated cannulas and insufflation tubing.
METHODS: Ports, cannulas, and tubing were subjected to 10 cycles of simulated use, bacterial inoculation with Staphylococcus pseudintermedius and Escherichia coli, decontamination and sterilization, and testing via culture based on their treatment group designation. Bacteriological scores were compared among the negative control, positive control, and 4 treated ports and components.
RESULTS: There was no difference in bacteriological scores between treated ports, cannulas, and insufflation tubing and the negative control port and components. Bacteriological scores of ports and components undergoing 6-10 cycles were not significantly different from scores of ports and components undergoing 5 or fewer sterilization cycles. No difference was found in detection of bacteria from treated ports by biopsy of the foam versus sampling via wash.
CONCLUSION: This study suggests that a single-use single-incision laparoscopic port and its associated components can be effectively sterilized after multiple simulated uses by using VHP. CLINICAL SIGNIFICANCE: Reuse of a single-use single-incision laparoscopic port is a safe and effective method of cost reduction in veterinary patients.
© 2018 The American College of Veterinary Surgeons.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29774550     DOI: 10.1111/vsu.12905

Source DB:  PubMed          Journal:  Vet Surg        ISSN: 0161-3499            Impact factor:   1.495


  3 in total

1.  Examination of Material Compatibilities with Ionized and Vaporized Hydrogen Peroxide Decontamination.

Authors:  Tohru Kimura; Hiroyuki Yahata; Yoshimichi Uchiyama
Journal:  J Am Assoc Lab Anim Sci       Date:  2020-09-17       Impact factor: 1.232

2.  Transanal access port (TrAAP) technique: the use of a single incision laparoscopic surgical port during canine colonoscopy (a cadaveric study).

Authors:  James Howard; Judith Bertran; Valerie Parker; Jenessa Winston; Adam J Rudinsky
Journal:  BMC Vet Res       Date:  2021-01-21       Impact factor: 2.741

3.  Maneuverability of the Scope and Instruments within Three Different Single-Incision Laparoscopic Ports: An Experimental Pilot Study.

Authors:  Georg Haider; Ursula Schulz; Nikola Katic; Christian Peham; Gilles Dupré
Journal:  Animals (Basel)       Date:  2021-04-26       Impact factor: 2.752

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.