Literature DB >> 29762187

Reliability of the Osstell Implant Stability Quotient and Penguin Resonance Frequency Analysis to Evaluate Implant Stability.

Gul Buyukguclu1, Zeynep Ozkurt-Kayahan2, Ender Kazazoglu3.   

Abstract

PURPOSE: To evaluate the reliability of the Osstell implant stability quotient (ISQ) and Penguin resonance frequency analysis (RFA) devices in measuring implant stability.
MATERIALS AND METHODS: Forty implants were embedded in self-curing acrylic resin, soft-lining material, polyvinyl siloxane impression material, and polycarboxylate cement (n = 10). After the initial setting times were completed, the stability of each implant was measured with Osstell ISQ and Penguin RFA with 3 repeated measurements. The "intraclass correlation coefficient" evaluated the correspondence between the measurements (P < 0.05).
RESULTS: Polyvinyl siloxane impression material had lower ISQ values than soft-lining material, self-curing acrylic resin, and polycarboxylate cement in both devices (P < 0.05). The intraclass correlation was 1.00 in self-curing acrylic resin and 0.48 in polycarboxylate cement (P < 0.05) for Osstell. This value was 0.95 in self-curing acrylic resin and 0.38 in polycarboxylate cement (P < 0.05) for Penguin. There was no correlation between the repeated measurements in soft-lining material and polyvinyl siloxane impression material for both devices (P > 0.05). The repeatability was 0.90 for Osstell and 0.60 for Penguin (P < 0.05).
CONCLUSIONS: Osstell ISQ and Penguin RFA are reliable only when the implants are embedded in stiff materials. Osstell ISQ is more reliable than Penguin RFA.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29762187     DOI: 10.1097/ID.0000000000000766

Source DB:  PubMed          Journal:  Implant Dent        ISSN: 1056-6163            Impact factor:   2.454


  3 in total

1.  Intra- and inter-operator concordance of the resonance frequency analysis. A cross-sectional and prospective clinical study.

Authors:  Raquel Antequera-Diaz; Maria-Pilar Quesada-García; Cristina Vallecillo; Marta Vallecillo-Rivas; Esther Muñoz-Soto; Maria-Victoria Olmedo-Gaya
Journal:  Clin Oral Investig       Date:  2022-07-08       Impact factor: 3.573

2.  Influence of cortical bone and implant design in the primary stability of dental implants measured by two different devices of resonance frequency analysis: An in vitro study.

Authors:  David Chávarri-Prado; Aritza Brizuela-Velasco; Markel Diéguez-Pereira; Esteban Pérez-Pevida; Antonio Jiménez-Garrudo; Iratxe Viteri-Agustín; Alejandro Estrada-Martínez; Oier Montalbán-Vadillo
Journal:  J Clin Exp Dent       Date:  2020-03-01

3.  Prospective, Clinical Pilot Study with Eleven 4-Mm Extra-Short Implants Splinted to Longer Implants for Posterior Maxilla Rehabilitation.

Authors:  Daniel Torassa; Pablo Naldini; José Luis Calvo-Guirado; Enrique Fernández-Bodereau
Journal:  J Clin Med       Date:  2020-01-28       Impact factor: 4.241

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.