Literature DB >> 29754982

Computed Tomography Techniques Help Understand Wear Patterns in Retrieved Total Knee Arthroplasty.

Arianna Cerquiglini1, Johann Henckel1, Harry S Hothi1, Lorenzo Dall'Ava1, Paul Shearing2, Michael T Hirschmann3, Alister J Hart1.   

Abstract

BACKGROUND: Suboptimal total knee arthroplasty (TKA) position of both femoral and tibial components is thought to be linked with poor clinical outcomes, polyethylene wear and the "unexplained" painful knee arthroplasty. The aim of this study was to better understand the effect of implant orientation on knee implant performance.
METHODS: We analyzed 30 retrieved contemporary TKA implants. Implant positioning measurements in the coronal plane were made prior to revision using a diagnostic algorithm, based on 3D computed tomography (CT) images. Each retrieved polyethylene component was imaged using a micro-CT scanner and a high resolution computational 3D model of each component was digitally reconstructed. The difference in thickness between medial and lateral components was calculated. Statistical analysis was performed to investigate the association between component positioning and damage patterns.
RESULTS: We found a significant correlation between both the tibiofemoral and femoral angles and difference in thickness between polyethylene compartments: varus angulations were strongly associated with thinner medial compartments, whilst valgus angulations were associated with thinner lateral compartments. Moreover, suboptimal tibiofemoral orientations and tibial component angulations were associated to greater differences in thickness between polyethylene compartments.
CONCLUSION: Our study is the first to compare accurate 3D CT measurements of prerevision TKA positioning in the coronal plane with postrevision retrieval analysis from innovative, accurate and highly reliable micro-CT-based method. Our results demonstrate the impact of component positioning on polyethylene damage and helps understanding of the in vivo performance of these implants. LEVEL OF EVIDENCE: III.
Copyright © 2018 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  3D computed tomography; implant orientation; malalignment; microcomputed tomography; polyethylene wear; total knee arthroplasty

Mesh:

Substances:

Year:  2018        PMID: 29754982     DOI: 10.1016/j.arth.2018.04.010

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  3 in total

1.  Retrieval analysis of contemporary antioxidant polyethylene: multiple material and design changes may decrease implant performance.

Authors:  Arianna Cerquiglini; Johann Henckel; Harry Hothi; Lukas B Moser; Antti Eskelinen; Michael T Hirschmann; Alister J Hart
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2019-02-10       Impact factor: 4.342

2.  CORR Insights®: Are Serum Metal Ion Levels a Concern at Mid-term Followup of Revision Knee Arthroplasty With a Metal-on-metal Hinge Design?

Authors:  Antti Eskelinen
Journal:  Clin Orthop Relat Res       Date:  2019-09       Impact factor: 4.176

3.  SPECT/CT Assessment of In-Vivo Loading of the Knee Correlates with Polyethylene Deformation in Retrieved Total Knee Arthroplasty.

Authors:  Harry Hothi; Arianna Cerquiglini; Lukas Büel; Johann Henckel; Lukas B Moser; Michael T Hirschmann; Alister Hart
Journal:  Tomography       Date:  2022-01-07