| Literature DB >> 29752010 |
Anwar Sheed Khan1, Sajid Ali2, Muhammad Tahir Khan3, Sajjad Ahmed4, Yasir Khattak5, Muhammad Irfan6, Wasim Sajjad7.
Abstract
GeneXpert is one of the recent technological instruments used to diagnose tuberculosis in a short span of time. In this study, the performance of GeneXpert MTB/RIF assay for the diagnosis of extrapulmonary tuberculosis (EPTB) was compared with light-emitting diode Fluorescent Microscopy (LED-FM) in Khyber Pakhtunkhwa, Pakistan. A total of 737 EPTB samples were collected from tuberculosis (TB) suspected patients. Out of these samples, male to female ratio was 53% (n=390) to 47% (n=347) respectively. The sensitivity and specificity was 73% and 100% for GeneXpert, while 40% and 100% for LED-FM microscopy. This shows that the sensitivity of GeneXpert is 40-50%, higher than LED-FM microscopy. GeneXpert also detected low number of bacilli as compared to LED-FM microscopy.Entities:
Keywords: GeneXpert; LED-FM microscopy; Sensitivity; Specificity
Mesh:
Year: 2018 PMID: 29752010 PMCID: PMC6175715 DOI: 10.1016/j.bjm.2018.02.011
Source DB: PubMed Journal: Braz J Microbiol ISSN: 1517-8382 Impact factor: 2.476
Sample wise comparison of LED-FM, GeneXpert and culture.
| Types | Total sample 2015 | 2014 | Total sample | Total number of processed sample | Positive | ||
|---|---|---|---|---|---|---|---|
| LED FM | GeneXpert | Culture | |||||
| Pleural fluid | 85 | 45 | 129 | 259 | 22 (8%) | 30 (12%) | 52 (20%) |
| BAL | 1 | 1 | 23 | 25 | 3 (12%) | 7 (28%) | 8 (32%) |
| CSF | 18 | 15 | 26 | 59 | 5 (8%) | 10 (17%) | 12 (20%) |
| Pus | 31 | 20 | 31 | 82 | 4 (5%) | 10 (12%) | 12 (15%) |
| Acsetic fluid | 48 | 20 | 57 | 125 | 8 (6%) | 14 (11%) | 14 (11%) |
| Synovail fluid | 1 | 1 | 11 | 13 | 2 (15%) | 6 (46%) | 7 (54%) |
| Urine | 15 | 10 | 22 | 47 | 4 (9%) | 4 (9%) | 7 (15%) |
| Pericardial fluid | 17 | 10 | 29 | 56 | 2 (4%) | 6 (11%) | 7 (12%) |
| Bone marrow | 20 | 15 | 1 | 36 | 0 | 3 (8%) | 6 (17%) |
| Tissues | 16 | 8 | 11 | 35 | 2 (8%) | 5 (14%) | 5 (14%) |
| Total | 252 | 145 | 340 | 737 | 52 (7%) | 95 (13%) | 130 (18%) |
| M = 390 | 53% | ||||||
| F = 347 | 47% | ||||||
Comparison between LED-FM, GeneXpert sensitivity and specificity.
| Total | Pleural fluid | BAL | CSF | Pus | Ascetic fluid | Synovail fluid | Urine | Pericardial Fluid | Bone Marrow | Tissues. | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| GX | LED-FM | GX | LED-FM | GX | LED-FM | GX | LED-FM | GX | LED-FM | GX | LED-FM | GX | LED-FM | GX | LED-FM | GX | LED-FM | GX | LED-FM | GX | LED-FM | |
| Sensitivity % | 73 | 40 | 58 | 42 | 88 | 38 | 83 | 42 | 83 | 33 | 100 | 57 | 86 | 29 | 57 | 57 | 86 | 29 | 50 | 0 | 100 | 40 |
| Specificity % | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
| PPV | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
| NPV | 95 | 89 | 90 | 87 | 94 | 77 | 96 | 87 | 98 | 90 | 100 | 95 | 86 | 55 | 93 | 93 | 98 | 91 | 91 | 83 | 100 | 91 |
PPV, positive predictive value; NPV, negative predictive value; GX, GeneXpert; CSF, cerebrospinal fluid; LED-FM, light emitting diode fluorescent microscopy.