Misdiagnosis of pulmonary tuberculosis and associated factors in peripheral laboratories: a retrospective study, Addis Ababa, Ethiopia.

Daniel Melese Desalegn1,2, Kumera Terfa Kitila3,4, Hanna Mekonnen Balcha3, Chalachew Sisay Gebeyehu3, Yohannes W Kidan3, Kassayenew Amare3, Daniel Dejene3, Merone Seifu3, Addis Zewdie3, Abiyot Tenna3, Tinsae Kidanemariam Hailu3, Boja Dufera Taddese3, Abrham Tesfaye Bika3.   

Abstract

OBJECTIVE: Sputum smear microscopy reading errors are likely to result in failure to detect persons with infectious TB. This study was intended to review misdiagnosis of pulmonary TB and associated factors in peripheral laboratories.
RESULTS: During the study period 1033 (10.5%) sputum smear positive and 8783 (89.5%) smear negative slides were reported by peripheral laboratories. The slides were re-read by the central referral laboratories (CRLs) as the reference standard reading. Of 1033 positive slides reported by peripheral laboratories, 25 (2.4%) were false positive. Out of 8783 smear negative slides reported by peripheral laboratories, 35 (0.4%) were false negative. The sensitivity, specificity, positive predictive value and negative predictive value of peripheral laboratories were 96.64, 99.72, 97.58, and 99.61% respectively. The peripheral laboratories and CRLs have an observed agreement (Po) of 0.9939. Of 135 peripheral laboratories, 93 (68.9%) read negative and positive slides correctly, 49 (36.3%) did not have lens cleaning tissue papers, 11 (8.1%) lacked frosted slides, and 14 (10.4%) had shortage of reagents. As conclusions, the peripheral laboratories and CRLs had high agreement for sputum smear microscopy reading. However, a few TB cases were misdiagnosed despite having the disease; these individuals might continue to spread the infection in the community.

Entities:  

Keywords:  Addis Ababa; External quality assessment; ZN sputum smear microscopy

Mesh:

Year:  2018        PMID: 29751778      PMCID: PMC5948669          DOI: 10.1186/s13104-018-3414-6

Source DB:  PubMed          Journal:  BMC Res Notes        ISSN: 1756-0500


Introduction

Laboratory services continue playing critical role in diagnosing TB and treatment monitoring [1, 2]. However, it requires strong quality assurance (QA) systems be in place [3-6]. In contrary, reading errors in ZN-stained sputum smear microscopy are likely to result in failure to detect persons with infectious TB, consequently continues to spread infection in the community, or may unnecessary treated for non TB cases [5-7]. Moreover, in adequately training, erratic reagent, supplies, and poor equipment maintenance inflate the problems. Accurate and reliable TB microscopy result relies on external quality assessment (EQA) programs that support, train, and monitor testing performance of individual laboratories [3, 8]. External quality assessment comprises proficiency testing, blind-rechecking of samples and on-site evaluation [7-9]. In study setting, blind-rechecking programs is crucial EQA method in assessing sputum smear microscopy reading performance of the laboratory [7, 8]. It provides an opportunity to assess quality performance elements including specimen quality, smear size and thickness, and quality of staining. All these information may be very useful to assessing possible reasons for false positive or false negative results, and implementing plans for retraining and corrective action [3, 9, 10]. Information concerning the performance of EQA in private and public laboratories in Addis Ababa is very limited; hence we reviewed 135 public and private health facilities’ ZN sputum smear microscopy EQA performance, to determine misdiagnosis of pulmonary TB cases and associated factors in private and public health facilities in Addis Ababa, Ethiopia.

Main text

Methods

Retrospective review of EQA records in 135 public and private laboratories in Addis Ababa Ethiopia was conducted, from October, 2014 to March, 2016. These peripheral laboratories provide TB diagnostic services and they participate in ZN smear microscopy Regional External Quality Assurance Scheme (REQAS). This study compared the results of peripheral laboratories’ ZN-stained sputum smear microscopy reading performances with CRLs as a gold standard.

Study subjects

One hundred thirty five peripheral laboratories (45 privates and 90 public) providing TB diagnostic service to patients in Addis Ababa were considered for the current study.

Data sources and sampling

Data were collected from 135 health facilities in Addis Ababa. The health facilities’ ZN-stained sputum smear microscopy EQA performances and onsite EQA records were used as data source.

Inclusion and exclusion criteria

All peripheral laboratories having complete EQA performance records in the CRLs data storage, from the years 2014–2016 were included. However, TB laboratory results performed by other than ZN technique were excluded.

Data management and statistical analysis

The collected data were checked for completeness and consistency and entered into using Epi-info software. The data were coded and analysed using SPSS version 20.0 (SPSS Inc. Chicago, USA) software. ZN-stained sputum smear microscopy reading agreement between the peripheral microscopy centers and the CRLs were measured by using Kappa (K). Inter-observer variability was assessed on the basis of κ-values, with kappa coefficient: 0.81–1 = almost perfect, 0.61–0.8 = substantial, 0.41–0.60 = moderate, 0.21–0.40 = fair, 0.01–0.20 = slight and ≤ 0 = poor agreement between slide readers [11]. Sensitivity, specificity, PPV and NPV of the peripheral laboratories were calculated against the final readings of the CRLs as a gold standard. Bivariate and Multivariate logistic regression model using odds ratio (OR) at 95% confidence interval (CI) were calculated. P value less than 0.05 were taken as statistically significant.

Data quality assurances

Before extracting data from records, data collectors were adequately trained and the collected data were checked for completeness by data collectors and authors. Data collection process was supervised and monitored by the principal investigator.

Operational definition

Central referral laboratories (CRLs): Regional and intermediate level (Hospital) laboratories serving as ZN smear microscopy a quality assurance (REQAS) centers for peripheral laboratories. Peripheral laboratories: Laboratories located at a Health Center or private Hospital and different level of Clinics, providing ZN smear microscopy diagnostic services. Discordant slides: Positive slides read as negative or vice versa.

Result

ZN-stained sputum smears microscopy reading agreement

During the study period 135 peripheral laboratories were involved in REQAS. Among these, 45 (33.3%) were private and the remaining 90 (66.7%) public health facilities. Of 135 peripheral laboratories, 93 (68.9%) correctly read negative and positive sputum smear slides. The remaining 42 (31.1%) laboratories misread at least one positive sputum smear slides as negative or vice versa. Out of 42 health peripheral laboratories reported false reading results, 23 (54.8%) were public and the remaining 19 (45.2%) were private’ laboratories. However, there was no statistically significant difference in sputum smear microscopy false reading results between public and private laboratories (P value > 0.05). Of 9816 ZN-stained sputum smear slides were collected from the peripheral laboratories for re-checking, 1033 (10.5%) were reported as positive, and the reaming 8783 (89.5%) were reported as negative by peripheral laboratories. The slides were re-read by the CRLs, which were considered as the reference standard reading. Of 1033 sputum smear slides reported as positive by peripheral laboratories, 1008 (97.6%) were true positive, the remaining 25 (2.4%) were false positive. Out of 8783 sputum smear slides reported as negative by peripheral laboratories, 8748 (99.6%) were true negative, the remaining 35 (0.4%) were false negative. The overall discordant rate between peripheral laboratories and CRLs was 60 (0.61%) (Table 1).
Table 1

Sputum smears microscopy reading agreement between peripheral laboratories and central referral laboratories in Addis Ababa, Ethiopia

Peripheral laboratories smear results (test value)Central referral laboratories results (as true value)Total
PositiveNegative
Positive1008251033
Negative3587488783
Total104387739816
Sputum smears microscopy reading agreement between peripheral laboratories and central referral laboratories in Addis Ababa, Ethiopia The study has shown an observed agreement (Po) of 0.9939, expected agreement (Pe) of 0.8109 and calculated kappa value was 0.97, which indicated almost perfect agreement. The sensitivity, specificity, PPV and NPV of peripheral laboratories in ZN-stained sputum reading was 96.64, 99.72, 97.58 and 99.6% respectively.

Smear quality

None of the peripheral laboratories assessed met sputum smear quality standards. Of 9816 collected sputum smear slides from the peripheral laboratories, 3475 (35.4%) specimen quality, 3316 (33.8%) evenness, 3004 (30.6%) smear size and 2817 (28.7%) labeling were found as poor. Moreover, 3846 (39.2%) smears were prepared with inappropriate thickness, and 3029 (30.9%) smears were unacceptable staining background cleanness. Compared to the performance of public and private laboratories for each of the quality indices, of 3168 sputum smear slides collected from the private laboratories, 1163 (36.7%) specimen quality, 1042 (32.9%) evenness, 991 (31.3%) smear size, 916 (28.9%) labeling, 1291 (40.8%) thickness, and 1039 (32.8%) staining background were unacceptable. Of 6648 slides collected from the public laboratories, 2312 (34.8%) specimen quality, 2274 (34.2%) evenness, 2013 (30.3%) smear size and 1901 (28.6%) labeling, 2552 (38.4%) thickness, and 2053 (30.9%) staining background were found as poor.

Tuberculosis laboratory facilities and infrastructures

On-site evaluation of EQA was conducted in 135 peripheral laboratories to assess the status of TB laboratory infrastructures and factors affecting the quality of ZN-stained sputum microscopy diagnosis. The entire assessed peripheral laboratories had functional light microscope, 104 (77.0%) had separate work station for smearing, staining, microscopy examination and recording. Of the total TB laboratories assessed, 79 (58.5%) had clean, separate and ventilated TB laboratory room, 127 (94.1%) had regular water supply and 131 (97.0%) had back up electric power supply (Table 2).
Table 2

TB laboratory diagnostic facilities and infrastructure assessment, in Addis Ababa, Ethiopia

VariablesNumberPercent
Separate TB laboratory room
 Yes7958.5
 No5641.5
Separate area for TB laboratory works
 Yes10477.0
 No3123.0
Regular electric power supply
 Yes13197
 No1914.1
Regular water supply
 Yes12794.1
 No85.9
Runn IQC slides parallel with routine tasks
 Yes3828.1
 No9771.9
SOPs, guideline and manuals
 Yes10779.3
 No2820.7
Functional light microscopes
 Yes135100
 No00.0
ZN-staining reagents
 Yes12189.6
 No1410.4
Lens cleaning tissue paper
 Yes4936.3
 No8663.7
IQC reagents
 Yes2619.3
 No10980.7
Good storage condition for reagents
 Yes8260.7
 No5339.3
Checking quality of sputum
 Yes7656.3
 No5943.7
Always carbol-fuchsin filtered before use
 Yes7857.8
 No5742.2
Having news slides (frosted slide)
 Yes12491.9
 No118.1
Re-used sputum container
 Yes00.0
 No135100
Having personal protective equipment
 Yes135100
 No00.0
Having functioning incinerator
 Yes135100
 No00.0
Staff received training on ZN sputum smears microscopy
 Yes10376.3
 No3223.7
Having standard TB result registration log-book
 Yes10980.7
 No2619.3
Received EQA feedback
 Yes12491.9
 No118.1
TB laboratory diagnostic facilities and infrastructure assessment, in Addis Ababa, Ethiopia Shortage of supplies including all type of reagents (Carbol-fuchsin, methylene blue and Acid alcohol), lens tissue; frosted slides, internal quality control (IQC) reagents were identified in most peripheral laboratories. Of 135 peripheral laboratories, 49 (36.3%) did not have lens cleaning tissue papers, 11 (8.1%) lacked frosted slides, 14 (10.4%) had shortage of ZN-staining reagents, and nearer to half (42.2%) did not filter 1% carbol-fuchsin during staining of sputum smears. However, all peripheral laboratories used the appropriate personal protective equipment (PPE) when performing TB laboratory procedures, and they used incinerator to dispose sputum, and other sputum contaminated solid wastes (Table 2).

Quality assurance

There were gaps in having valid internal quality control materials in all assessed laboratories. Among these, 38 (28.1%) laboratories prepare smears from known positive and negative sputum as internal quality control to check reagents quality before staining patient sample. About 59 (43.7%) did not check sputum quality before processing. Among 135 visited laboratories, 109 (80.7%) had standard TB laboratory request form and result log book, 107 (79.3%) had TB national guideline; manual, SOPs and other reference materials for smear microscopy technique. More than half (76.3%) of the peripheral laboratories’ staffs members were trained on sputum smear microscopy (Table 2). From those who took training, 73 (70.9%) were from public health facilities.

Factors affecting sputum smear microscopic reading

In binary logistic regression, false reading of sputum smear microscopy had significant association with checking the quality of sputum before smear preparation, having separate TB laboratory room and filtering carbol-fuchsin before staining sputum smear (P value < 0.05). However, in multivariate logistic regression analysis, false reading had no significant association with these factors (P value > 0.05) (Table 3).
Table 3

Factors affecting for ZN sputum smears microscopy reading in peripheral diagnostic facilities, Addis Ababa, Ethiopia

VariablesFalse readingCOR (95% CI)P valueAOR (95% CI)P value
NoYes
Separate TB laboratory room
 Yes31291
 No25502.14 (1.06, 4.30)0.033*0.46 (0.74, 2.89)0.41
Training on ZN microscopy
 Yes14461
 No18570.96 (0.43, 2.14)0.93
Cheek quality of sputum
 Yes32281
 No27482.03 (1.02, 4.06)0.04*0.63 (0.12, 3.28)0.59
Use new slides
 Yes6541
 No5701.56 (0.45, 5.37)0.49
Filter reagent before use
 Yes31291
 No26492.02 (1.02, 4.04)0.048*1.54 (0.14, 4.17)0.73
Prepared and use IQC
 Yes25101
 No84160.47 (0.32, 1.79)0.53
Having lenses tissue
 Yes20151
 No66340.68 (0.29, 1.17)0.13
Good reagent storage condition
 Yes34661
 No19160.43 (0.28, 1.15)0.12
Received feed back
 Yes6691
 No5550.96 (0.28, 3.30)0.94
Owner ship
 Private26191
 Government67230.87 (0.42, 1.80)0.47
Having ZN reagents
 Yes5741
 No63112.49 (0.87, 12.29)0.08

* Statistically significant P < 0.05

Factors affecting for ZN sputum smears microscopy reading in peripheral diagnostic facilities, Addis Ababa, Ethiopia * Statistically significant P < 0.05

Discussion

In the current study, sputum smear microscopy reading agreements between the peripheral laboratories with the CRLs were evaluated, using CRLs as reference standard. The sensitivity, specificity, PPV and NPV of the peripheral laboratories were 96.64, 99.72, 97.58 and 99.61%, respectively, which is comparable to 95, 99.7, 93.3 and 99.7% respectively reported in Ethiopia and 91.3, 98.9, 92.2 and 98.8% respectively reported in Argentine [12, 13]. In contrast, it is higher than others studies [14-16]. In present study overall sputum smear reading agreement between the peripheral diagnostic laboratories and CRLs was 99.4%, which is comparable to 99.5% reported in west Amhara Region [17] and 99.7%, reported Lima, Peru [18]. However, it is higher compared to 96.8% reported in south Ethiopia [19] and 87% reported in Addis Ababa [16] and 89.2% report in Tanzania [20]. Although, the proportion of sputum smear microscopy reading agreement of 99.4% achieved is higher than acceptable performance of > 80% [7, 10, 21], still disagreement of 0.6% may indicate a need to improve the quality of TB microscopic reading. In this study overall discordant (false reading) results of sputum microscopy reported by the peripheral laboratories was 60 (0.6%), which is lower than the previous reports of 7.8% in Argentine, 3.5% in west Amhara Region, 3.2% in southern Ethiopia, 3.3% in India, and 5.5% in eastern part of Ethiopia [13, 17, 19, 22, 23]. It is higher than study reported 0.2% in Ethiopia and 0.3% in Taiwan [12, 24]. These variations might be due to sputum smear microscopy reading is highly dependent on the training, diligence of microscopist and laboratory supplies. The other possible reason, unfiltered fuchsin crystals reagents, weak decoloration of AFB might affect the quality of ZN smear microscopy reading and lead to false reading results [16, 20, 26]. In the present study specimen quality, evenness, smears size, inappropriate labeling, thickness, and smears with unacceptable staining background cleanness were 35.4, 33.8, 30.6, 28.7, 39.2 and 31.5% respectively, which is relatively lower than other studies [24-26]. Specimen container, sample transport condition, specimen quality, labeling, smearing, and staining technique including failure to filter carbol fuchsin are some of the factors that compromise ZN sputum smears quality performance indices [20, 26]. In the current study 79 (58.5%) of the peripheral laboratories having clean, separate and ventilated TB laboratory room, which is higher than 18.2–42% previously reported in other studies [15, 16, 27, 28]. Unventilated TB laboratory room can create unfavorable working environment, which might affect the overall quality of sputum smear microscopy services from pre-analytical to post analytical phases. In addition, this may create a chance of cross contamination [2, 29]. Regarding TB laboratory safety issue, all the peripheral laboratories used the appropriate PPE and disposed sputum and other sputum contaminated solid wastes appropriately. This finding was higher when compared with study conducted in Malawi, which reported most hospitals’ laboratories didn’t wear white coat, face mark, protective apron and soap for washing hand [27]. The difference might be due to availability of PPE, professionals’ attitude towards PPE. A poor laboratory safety practice does not only put the laboratory workers at risk of infection, but also the patients and any other person accessing the laboratory services.

Conclusion

In the present study peripheral laboratories and CRLs had high agreement for ZN sputum smear microscopy reading. However, a few TB cases were misdiagnosed despite having the disease; these individuals might continue to spread the infection in the community. This indicates the needs to improve ZN-stain sputum smear microscopy reading to attain the end TB strategy.

Limitations of the study

Results of this study depend only on the TB-EQA record review. Therefore, not illustrate providers view.
  17 in total

1.  Quality assessment of smear microscopy by stratified lot sampling of treatment follow-up slides.

Authors:  L Otero; A Van Deun; J Agapito; R Ugaz; G Prellwitz; E Gotuzzo; P Van der Stuyft
Journal:  Int J Tuberc Lung Dis       Date:  2011-02       Impact factor: 2.373

2.  Understanding interobserver agreement: the kappa statistic.

Authors:  Anthony J Viera; Joanne M Garrett
Journal:  Fam Med       Date:  2005-05       Impact factor: 1.756

3.  Quality assessment of sputum smear microscopy for detection of acid fast bacilli in peripheral health care facilities in Dar es Salaam, Tanzania.

Authors:  D Basra; M I N Matee; R McNerney
Journal:  East Afr Med J       Date:  2006-06

Review 4.  Optimizing sputum smear microscopy for the diagnosis of pulmonary tuberculosis.

Authors:  Karen R Steingart; Andrew Ramsay; Madhukar Pai
Journal:  Expert Rev Anti Infect Ther       Date:  2007-06       Impact factor: 5.091

5.  External quality assessment of smear microscopy by the National Reference Laboratory in nine states of India.

Authors:  V Kumar; R Raghavan; S Nagamiah; L S Chauhan
Journal:  Int J Tuberc Lung Dis       Date:  2009-09       Impact factor: 2.373

6.  Safety in laboratories carrying out sputum smear microscopy: a dilemma for resource-poor countries.

Authors:  T E Nyirenda; C J Mundy; A D Harries; A Banerjee; F M Salaniponi
Journal:  Int J Tuberc Lung Dis       Date:  1998-08       Impact factor: 2.373

7.  Quality assessment of smear microscopy for acid-fast bacilli in the Argentine tuberculosis laboratory network, 1983-2001.

Authors:  G F Kusznierz; O A Latini; M D Sequeira
Journal:  Int J Tuberc Lung Dis       Date:  2004-10       Impact factor: 2.373

8.  External quality assessment for acid fast bacilli smear microscopy in eastern part of Ethiopia.

Authors:  Desalegn Admassu Ayana; Zelalem Teklemariam Kidanemariam; Habtamu Mitiku Tesfaye; Fitsum Weldegebreal Milashu
Journal:  BMC Res Notes       Date:  2015-10-05

9.  The quality of sputum smear microscopy in public-private mix directly observed treatment laboratories in West Amhara region, Ethiopia.

Authors:  Almaw Manalebh; Meaza Demissie; Daniel Mekonnen; Bayeh Abera
Journal:  PLoS One       Date:  2015-04-07       Impact factor: 3.240

10.  External quality assessment of AFB smear microscopy performances and its associated factors in selected private health facilities in Addis Ababa, Ethiopia.

Authors:  Lemi Mosissa; Abebaw Kebede; Tedla Mindaye; Muluwork Getahun; Sisay Tulu; Kassu Desta
Journal:  Pan Afr Med J       Date:  2016-06-09
View more
  6 in total

1.  Evaluation of risk factors associated with the development of MDR- and XDR-TB in a tertiary care hospital: a retrospective cohort study.

Authors:  Amna Saifullah; Tauqeer Hussain Mallhi; Yusra Habib Khan; Muhammad Shahid Iqbal; Nasser Hadal Alotaibi; Abdulaziz Ibrahim Alzarea; Maria Rasheed
Journal:  PeerJ       Date:  2021-03-18       Impact factor: 2.984

2.  Comparison of different bacteriological testing strategies and factors for bacteriological confirmation among pulmonary TB patients: a retrospective study in Tianjin, China, 2017-2018.

Authors:  Guoqin Zhang; Yuhua Zhang; Mingting Chen; Fan Zhang
Journal:  BMC Infect Dis       Date:  2020-07-29       Impact factor: 3.090

3.  Sputum Smear Positive Pulmonary Tuberculosis Diagnostic Dropout Rate in Public Health Facilities, Addis Ababa, Ethiopia.

Authors:  Daniel Melese Desalegn; Kumera Terfa Kitila; Boja Dufera Taddese; Tinsae Kidanemariam Hailu; Tariku Takle Dinku; Kassahun Demisse Asferie; Hanna Mekonnen Balcha; Chalachew Sisay Gebeyehu; Girmay Medhin
Journal:  Pulm Med       Date:  2019-03-20

4.  A Low-Cost Humidity Control System to Protect Microscopes in a Tropical Climate.

Authors:  Anders J Asp; Christina M Webber; Evan N Nicolai; Gabriel Martínez-Gálvez; Victoria S Marks; Ephraim I Ben-Abraham; John W Wilson; J Luis Lujan
Journal:  Ann Glob Health       Date:  2020-02-13       Impact factor: 2.462

5.  Barriers in the access, diagnosis and treatment completion for tuberculosis patients in central and western Nepal: A qualitative study among patients, community members and health care workers.

Authors:  Sujan Babu Marahatta; Rajesh Kumar Yadav; Deena Giri; Sarina Lama; Komal Raj Rijal; Shiva Raj Mishra; Ashish Shrestha; Pramod Raj Bhattrai; Roshan Kumar Mahato; Bipin Adhikari
Journal:  PLoS One       Date:  2020-01-15       Impact factor: 3.240

6.  Quality of Same-Day Sputum Smears Microscopy and Presumptive Tuberculosis Patients Drop-out at Health Facilities of Addis Ababa, Ethiopia.

Authors:  Boja Dufera Taddese; Abay Sisay Misganaw
Journal:  Tuberc Respir Dis (Seoul)       Date:  2020-01
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.