Similar radiological results with accelerometer-based navigation versus conventional technique in total knee arthroplasty.

Ing How Moo1, Jerry Yong Qiang Chen1, Dickson Hong Him Chau1, See Wei Tan1, Adrian Cheng Kiang Lau1, Yee Sze Teo1.   

Abstract

BACKGROUND: The Zimmer iASSIST system is an accelerometer-based, portable navigation device for total knee arthroplasty (TKA) that does not require the use of a large console for alignment feedback as required in computer-assisted surgery. The purpose of this study was to determine the accuracy of the accelerometer-based system in component positioning and overall mechanical alignment.
METHODS: Two groups of 30 patients each with primary osteoarthritis underwent TKA using either conventional method or Zimmer iASSIST navigation in 2013 was retrospectively studied. Patients were matched according to body mass index (BMI), gender and age. A senior arthroplasty surgeon performed all the operation using the same surgical approach. Perioperative and post-operative regimens were the same. All patients had standardized radiographs performed post-operatively to determine the lower limb mechanical alignment and component placement.
RESULTS: There was no difference between the two groups for age, BMI, gender, side of operated knee and preoperative mechanical axis ( p > 0.05). There was no difference in the proportion of outliers for mechanical axis ( p = 0.38), coronal femoral angle ( p = 0.50), coronal tibia angle ( p = 0.11), sagittal femoral angle ( p = 0.28) and sagittal tibia angle ( p = 0.33). The duration of surgery, post-operative drop in haemoglobin level and transfusion incidence did not show statistically significant differences between the two groups ( p > 0.05).
CONCLUSIONS: Our article showed that iASSIST was safe and remains a useful tool to restore mechanical axis. However, our data demonstrated no difference in lower limb alignment and component placement between the TKA that used accelerometer-based system and those that underwent conventional method.

Entities:  

Keywords:  accelerometer-based navigation; computer navigation; iASSIST; knee arthroplasty; total knee replacement

Mesh:

Year:  2018        PMID: 29747561     DOI: 10.1177/2309499018772374

Source DB:  PubMed          Journal:  J Orthop Surg (Hong Kong)        ISSN: 1022-5536            Impact factor:   1.118


  3 in total

1.  Total knee arthroplasty in the next decade: is navigation necessary?

Authors:  Graham S Goh; Ming Han Lincoln Liow
Journal:  Ann Transl Med       Date:  2020-06

2.  Usefulness of an accelerometer-based navigation system in bilateral one-stage total knee arthroplasty.

Authors:  Artit Laoruengthana; Piti Rattanaprichavej; Thanawat Tantimethanon; Watcharapong Eiamjumras; Passakorn Teekaweerakit; Krit Pongpirul
Journal:  BMC Musculoskelet Disord       Date:  2021-02-10       Impact factor: 2.362

3.  Comparison of iASSIST Navigation System with Conventional Techniques in Total Knee Arthroplasty: A Systematic Review and Meta-Analysis of Radiographic and Clinical Outcomes.

Authors:  Jun-Tan Li; Xiang Gao; Xu Li
Journal:  Orthop Surg       Date:  2019-11-22       Impact factor: 2.071

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.