Abstract
A simple, inexpensive, versatile, and environment-friendly extraction method, using low-temperature partitioning extraction (LTPE), was validated to quantify pharmaceutical-active compounds (PhACs) in surface water samples by high-performance liquid chromatography coupled to electrospray ionization tandem mass spectrometry (HPLC-ESI-MS/MS). The PhACs analyzed were acetaminophen, bezafibrate, diclofenac, diltiazem, fluconazole, linezolid, miconazole, ondansetron hydrochloride, and trimethoprim. The detection and quantification limits ranged from 0.15 to 12.30 ng L-1 and 0.43 to 40.60 ng L-1, respectively. Recovery rates ranged from 46 to 135%, and relative standard deviation (RSD%) varied between 0.49 and 6.13%. This method was applied to monitor water contamination by PhACs in the Paraopeba River Basin (PRB), Minas Gerais state, Brazil. All PhACs, except linezolid which was not detected, were found in PRB water samples in concentrations that ranged from 2.6 ng L-1 to 2.62 μg L-1.Entities:
Keywords: HPLC-ESI-MS/MS; Low-temperature partitioning extraction; Pharmaceuticals; Surface water
Mesh:
Substances:
Year: 2018 PMID: 29744776 DOI: 10.1007/s11356-018-2123-y
Source DB: PubMed Journal: Environ Sci Pollut Res Int ISSN: 0944-1344 Impact factor: 4.223