Literature DB >> 29739569

Rapid determination of cadmium in rice using an all-solid RGO-enhanced light addressable potentiometric sensor.

Wen Zhang1, Yiwei Xu2, Xiaobo Zou3.   

Abstract

Herein, an all-solid light addressable potentiometric sensor (LAPS) is presented for determination of cadmium (Cd) in rice. On the working surface of the LAPS, reduced graphene oxide (RGO) is introduced as a part of ion-to-electron transducer to improve ionophore behaviors. The composite modification of RGO and ionophore is validated with scanning electron microscopy. The as-fabricated sensor presents a rapid response in less than 10 s to target Cd. Meanwhile, it shows lower noise (0.23 mV) and better limit of detection (0.002 mg L-1) than LAPS (control) without RGO modification (0.37 mV; 0.008 mg L-1). With the proposed method, satisfactory precision, accuracy and selectivity are also established. This method is adopted in an extensive survey for 25 rice samples from 5 regions in China. The results are in very good agreement with those obtained using inductively-coupled plasma-mass spectrometry.
Copyright © 2018 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Heavy metal contamination; Ion-selective layer; Light addressable potentiometric sensor

Mesh:

Substances:

Year:  2018        PMID: 29739569     DOI: 10.1016/j.foodchem.2018.04.022

Source DB:  PubMed          Journal:  Food Chem        ISSN: 0308-8146            Impact factor:   7.514


  1 in total

Review 1.  Light-Addressable Potentiometric Sensors in Microfluidics.

Authors:  Xueliang Li; Shibin Liu; Jie Tan; Chunsheng Wu
Journal:  Front Bioeng Biotechnol       Date:  2022-02-21
  1 in total