A E Quimby1, S Kohlert2, L Caulley3, M Bromwich4.
Abstract
BACKGROUND: Flexible nasolaryngoscopy is an essential component of the otolaryngological physical exam. Historically, the ability to create and share video recordings of these endoscopic exams has been limited by poor mobility of fixed endoscopy towers. The advent of smartphone endoscope adapters has allowed physicians to create and share video recordings of endoscopy in a wide variety of locations that would not have previously been feasible. This paper sought to review the literature on the effect of smartphone endoscope adapters on patient care, patient satisfaction, and resident learning.Entities:
Keywords: FNL; Flexible nasolaryngoscopy; Laryngoscopy; Smartphone endoscope adapters
Mesh:
Year: 2018 PMID: 29739440 PMCID: PMC5941791 DOI: 10.1186/s40463-018-0279-6
Source DB: PubMed Journal: J Otolaryngol Head Neck Surg ISSN: 1916-0208
Inclusion and Exclusion Criteria
| Inclusion Criteria | |
| Population | ○ Endoscopy performed by Otolaryngology residents/ physicians |
| Intervention | ○ Use of smartphone adapters for flexible nasolaryngoscopy |
| Comparator | ○ Endoscopy tower video recordings of flexible nasolaryngoscopy, or no recording |
| Outcome | ○ Patient satisfaction |
| Study Design | ○ Randomized and non-randomized comparative studies, retrospective and prospective cohort studies, case series |
| Exclusion Criteria | |
| Population | ○ Non-Otolaryngology-trained residents/ physicians |
| Intervention | ○ Endoscopy other than flexible nasolaryngoscopy |
| Comparator | ○ NA |
| Outcome | ○ No reported outcomes |
| Study Design | ○ Single case reports |
Included study characteristics
| Study name | Year | Type of Study | Patients (N) | Exams recorded with endoscope adapter (N) | Exams recorded with endoscopy video tower (N) | Scope operator trainee level | Model of scope adapter | Model of endoscopy video tower | Model of smartphone | Safety/ Privacy measures |
|---|---|---|---|---|---|---|---|---|---|---|
| Liu H et al.. | 2016 | Prospective cohort | 30 | 30 | 30 | Staff, residents (all levels) | ClearScope (Clearwater Clinical Limited, Ottawa, Canada) | KayPentax (HOYA Corporation, Pentax Lifecare Division, Tokyo, Japan) | iPhone (Apple, Cupertino, CA) | Modica (Clearwater Clinical Limited, Ottawa, Canada) |
| Liu YF et al | 2016 | Prospective cohort | 43 | 43 | 0 | Residents (PGY-1, −2) | ClearScope | NA | NA | NA |
| Lozada et al | 2017 | Prospective cohort | 79 | 79 | 0 | Residents (PGY-1) | Mobile Optyx (MobileOptyx, Philadelphia, PA) | NA | iPhone | Dedicated team iPhone |
Included study outcomes
| Study name | Year | Primary Outcomes of Interest | Secondary Outcomes | Outcome Measures | Findings |
|---|---|---|---|---|---|
| Liu H et al | 2016 | Diagnostic accuracy | NA | Controlled blinded comparison of scope adapters and endoscope tower recorded videos | No significant difference between scope adapter and endoscopy tower videos (mean difference = 1.54%, p = 0.69). |
| Video recording quality | 5-point Likert scale across 7quality variables | No significant difference across 7 categories ( | |||
| Liu YF et al | 2016 | Resident Education | NA | Resident and attending self-ratings of educational value of scope adapter examinations (non-validated 5-point scale) | Residents felt that reviewing examinations recorded with scope adapters enhanced learning in 79% of cases, and that ability to discuss recorded exams with attendings enhanced learning in 88% of cases. Attendings felt discussing recordings enhanced learning in 81% of cases. |
| Lozada et al | 2017 | Diagnostic accuracy | NA | Event rates of discordant diagnoses between staff/ resident based on smartphone adapter recordings; χ2 to compare frequency of discordant diagnoses across diagnostic categories | 11% frequency of discordant exams; No statistically significant difference in number of discordant diagnoses among diagnostic categories |
| Video recording quailty | Event rate of repeated examinations | 1.3% of exams needed to be repeated due to poor recording quailty |
Fig. 1Search strategy and results
Fig. 2Mobile endoscope adapter (ClearScope; Clearwater Clinical Limited, Ottawa, Canada)