Prevalence of Extracochlear Electrodes: Computerized Tomography Scans, Cochlear Implant Maps, and Operative Reports.

Jourdan T Holder1, David M Kessler1, Jack H Noble2, René H Gifford1,3, Robert F Labadie3,2.   

Abstract

OBJECTIVE: To quantify and compare the number of cochlear implant (CI) electrodes found to be extracochlear on postoperative computerized tomography (CT) scans, the number of basal electrodes deactivated during standard CI mapping (without knowledge of the postoperative CT scan), and the extent of electrode insertion noted by the surgeon. STUDY
DESIGN: Retrospective.
SETTING: Academic Medical Center.
METHODS: Two hundred sixty-two patients underwent standard cochlear implantation and postoperative temporal bone CT scanning. Scans were analyzed to determine the number of extracochlear electrodes. Standard CI programming had been completed without knowledge of the extracochlear electrodes identified on the CT. These standard CI maps were reviewed to record the number of deactivated basal electrodes. Lastly, each operative report was reviewed to record the extent of reported electrode insertion.
RESULTS: 13.4% (n = 35) of CIs were found to have at least one electrode outside of the cochlea on the CT scan. Review of CI mapping indicated that audiologists had deactivated extracochlear electrodes in 60% (21) of these cases. Review of operative reports revealed that surgeons correctly indicated the number of extracochlear electrodes in 6% (2) of these cases.
CONCLUSIONS: Extracochlear electrodes were correctly identified audiologically in 60% of cases and in surgical reports in 6% of cases; however, it is possible that at least a portion of these cases involved postoperative electrode migration. Given these findings, postoperative CT scans can provide information regarding basal electrode location, which could help improve programming accuracy, associated frequency allocation, and audibility with appropriate deactivation of extracochlear electrodes.

Entities:  

Mesh:

Year:  2018        PMID: 29738386     DOI: 10.1097/MAO.0000000000001818

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  3 in total

1.  Musical Sound Quality as a Function of the Number of Channels in Modern Cochlear Implant Recipients.

Authors:  Katelyn Berg; Jack Noble; Benoit Dawant; Robert Dwyer; Robert Labadie; Virginia Richards; René Gifford
Journal:  Front Neurosci       Date:  2019-09-24       Impact factor: 4.677

2.  Detection of Extracochlear Electrodes in Cochlear Implants with Electric Field Imaging/Transimpedance Measurements: A Human Cadaver Study.

Authors:  Simone R de Rijk; Yu C Tam; Robert P Carlyon; Manohar L Bance
Journal:  Ear Hear       Date:  2020 Sep/Oct       Impact factor: 3.562

3.  Extracochlear Electrode Insertion in the Petrous Area in Cochlear Implantation.

Authors:  Chia-Hsing Lin; Tai-Lien Chiang; Jon-Kway Huang; Hung-Ching Lin
Journal:  J Int Adv Otol       Date:  2021-05       Impact factor: 1.316

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.