Literature DB >> 29731190

Novel driveline route for prevention from driveline infection: Triple tunnel method.

Shuichi Yoshitake1, Osamu Kinoshita2, Kan Nawata2, Yasuhiro Hoshino2, Yoshifumi Itoda2, Mitsutoshi Kimura2, Haruo Yamauchi2, Minoru Ono3.   

Abstract

BACKGROUND: The most prevalent and serious infection related to left ventricular assist devices (LVADs) is driveline infection (DLI). From 2014, we employed a revised surgical technique (triple tunnel method), which deployed a longer subfascial driveline (DL) route. METHODS AND PATIENTS: We retrospectively analyzed 34 patients fitted with either of the two types of axial pumps: HeartMate II (n=23) and Jarvik 2000 (n=11). Prior to 2014, the DL proceeded from the pump pocket just above the posterior sheath of the rectus muscle toward a vertical skin incision at the right lateral border of the rectus muscle. Then, DL was turned leftward into the subcutaneous tissue to redirect its exit to the left side [subcutaneous tissue group (Group S): n=14]. From 2014, we made an additional skin incision below the umbilicus with the aim of lengthening the subfascial DL route [muscle group (Group M): n=20].
RESULTS: DLI occurred in 10 patients (71.4%) in Group S and in 1 patient (5%) in Group M (p<0.05, Chi-square test). The freedom rate from re-admission at 1 year due to DLI was 64% in Group S and 95% in Group M, respectively (p=0.021, log-rank test). Furthermore, logistic regression analysis revealed that DL route was significantly associated with DLI (odds ratio, 10.1; 95% confidence interval, 1.15-275.3).
CONCLUSION: Although a longer follow-up period will be needed, the triple tunnel method may be beneficial in the prevention of DLI.
Copyright © 2018 Japanese College of Cardiology. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Driveline infection; Heart failure; Mechanical circulatory support; Ventricular assisted device

Mesh:

Year:  2018        PMID: 29731190     DOI: 10.1016/j.jjcc.2018.04.003

Source DB:  PubMed          Journal:  J Cardiol        ISSN: 0914-5087            Impact factor:   3.159


  3 in total

Review 1.  Decreasing driveline infections in patients supported on ventricular assist devices: a care pathway approach.

Authors:  Julia Seretny; Tara Pidborochynski; Holger Buchholz; Darren H Freed; Roderick MacArthur; Nicole Dubyk; Laura Cunliffe; Osiris Zelaya; Jennifer Conway
Journal:  BMJ Open Qual       Date:  2022-05

Review 2.  In Full Flow: Left Ventricular Assist Device Infections in the Modern Era.

Authors:  Radoslav Zinoviev; Christopher K Lippincott; Sara C Keller; Nisha A Gilotra
Journal:  Open Forum Infect Dis       Date:  2020-04-17       Impact factor: 3.835

3.  Driveline Features as Risk Factor for Infection in Left Ventricular Assist Devices: Meta-Analysis and Experimental Tests.

Authors:  Melanie Kranzl; Martin Stoiber; Anne-Kristin Schaefer; Julia Riebandt; Dominik Wiedemann; Christiane Marko; Günther Laufer; Daniel Zimpfer; Heinrich Schima; Thomas Schlöglhofer
Journal:  Front Cardiovasc Med       Date:  2021-12-16
  3 in total