Improved stent expansion with prolonged compared with short balloon inflation: A meta-analysis.

Marwan Saad1,2, Mahesh Bavineni3, Barry F Uretsky1,4, Srikanth Vallurupalli1,4.   

Abstract

BACKGROUND: Despite evidence from individual studies suggesting that prolonged inflation improves coronary stent expansion, relatively shorter inflation times are commonly employed in clinical practice.
METHODS: We performed an electronic search of PubMed, Web of Science, Cochrane, and CINAHL databases to retrieve outcome studies comparing prolonged versus short inflation times during stent deployment. Outcomes studied included minimal stent diameter (MSD) and minimal stent area (MSA). Standardized mean difference (SMD) was used to estimate the effect sizes for these continuous variables.
RESULTS: Seven studies with a total of 341 patients (415 lesions; mean age 67.6 years; 82% male) were included. Drug-eluting stents were used in 89 ± 15% of coronary lesions. Prolonged stent inflation was associated with increased minimal stent diameter (2.93 ± 0.34 mm vs. 2.72 ± 0.28 mm; SMD = 0.42; 95% CI 0.25-0.59; P < 0.001) and minimal stent area (5.99 ± 1.21 mm2 vs. 5.17 ± 0.87 mm2 ; SMD = 0.46; 95% CI 0.19-0.73; P = 0.001) compared with shorter duration stent inflation. This difference remained significant in sensitivity analyses that excluded studies with very prolonged inflation duration or multiple stent balloon inflations.
CONCLUSION: Despite differences in duration and methodology, prolonged stent inflation is associated with improved expansion compared with shorter duration. The effect of this optimization recommends randomized trials to determine whether long-term outcomes can be improved by this simple technical modification.
© 2018 Wiley Periodicals, Inc.

Entities:  

Keywords:  percutaneous coronary intervention; stent; stent deployment

Mesh:

Year:  2018        PMID: 29726622     DOI: 10.1002/ccd.27641

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  1 in total

1.  Impact of the Balloon Inflation Time and Pattern on the Coronary Stent Expansion.

Authors:  Jarosław Skowroński; Rafał Wolny; Jan Jastrzębski; Paweł Tyczyński; Karol Szlazak; Jerzy Pręgowski; Gary S Mintz; Karolina Liżewska; Wojciech Świeszkowski; Zbigniew Chmielak; Adam Witkowski
Journal:  J Interv Cardiol       Date:  2019-03-21       Impact factor: 2.279

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.