| Literature DB >> 29724977 |
Cemal Kemaloğlu1, Refik Emre Altekin, Ömer Bayezid.
Abstract
As the number of implanted left ventricular assist devices (LVADs) used increases, the frequency of chronic complications encountered also increases. The pause of blood flow in the outflow graft is a rare, but fatal, complication. The aim of this article was to present the case of a patient in whom HeartWare outflow graft occlusion was removed by balloon angioplasty and to examine the treatment modalities of HeartWare outflow graft occlusions that have been percutaneously performed to date. The literature was searched for percutaneous interventions on outflow grafts of the left ventricular assist devices. The results of six patients who underwent interventions on outflow grafts were analyzed. Three of six patients with HeartWare outflow graft stenosis were treated with covered stents, while the remaining three were treated with bare metal stents. All procedures were applied successfully. Percutaneous interventions can be performed with appropriate equipment in patients with HeartWare outflow graft stenosis or total occlusion.Entities:
Mesh:
Year: 2018 PMID: 29724977 PMCID: PMC6280264 DOI: 10.14744/AnatolJCardiol.2018.76736
Source DB: PubMed Journal: Anatol J Cardiol ISSN: 2149-2263 Impact factor: 1.596
Figure 1HVAD parameters and flow pulsatility over the last 14 days
Figure 2Multislice CT view; a totally occluded outflow graft is seen and is adherent to sternal sutures (arrow)
Figure 3Angiogram; stenosis was detected from the ostium of the outflow graft to the middle part
Figure 4Angiogram; balloon dilatations
Figure 5Embolectomy material
Figure 6First-month CT view (outflow graft is seen)
Interventions to stenotic outflow grafts: a summary
| Reference | Pre-Interventional flow and power (L/W) | Pre-Interventional rate (RPM) | Post-Interventional flow and power (L/W) | Post-Interventional rate (RPM) | Intervention | Cerebral protection | Device On/Off | Stenosis severity |
|---|---|---|---|---|---|---|---|---|
| 2 | 2.2/4.9 | 3600 | 5/6 | 3200 | Covered stent implantation | None | Off | Unknown |
| 3 | 2.5/Unknown | 2900 | 4/Unknown | 2350 | Covered stent implantation | None | On | 90% |
| 4 | 1/2 | 2700 | 6/4.9 | 2700 | Bare-metal stent implantation | None | On | Nearly occluded |
| 4 | 3/2.5 | 3000 | 6/5.9 | 3000 | Bare-metal stent implantation | None | On | Nearly occluded |
| 4 | 2/1 | 2500 | 4/3 | 2500 | Bare- metal stent implantation | Balloon blockade | On | occluded |
| 5 | 8/Unknown | Unknown | Unknown | Unknown | Covered stent implantation | Balloon blockade + embolic filter | Off | Unknown |
| 0 | 2500 | 3.7/2.9 | 2500 | Balloon dilatation only | Manual carotid occlusion | Off | Total occlusion |
L - Liters; W - Watts; RPM - Revolutions per minute
Our patient