Pulmonary deposition of fluticasone propionate/formoterol in healthy volunteers, asthmatics and COPD patients with a novel breath-triggered inhaler.

Dominik Kappeler1, Knut Sommerer2, Claudius Kietzig3, Bärbel Huber4, Jo Woodward5, Mark Lomax6, Prashant Dalvi7.   

Abstract

INTRODUCTION: A combination of fluticasone propionate/formoterol fumarate (FP/FORM) has been incorporated within a novel, breath-triggered device, named K-haler®. This low resistance device requires a gentle inspiratory effort to actuate it, triggering at an inspiratory flow rate of approximately 30 L/min; thus avoiding the need for coordination of inhalation with manual canister depression. The aim of the study was to evaluate total and regional pulmonary deposition of FP/FORM when administered via the K-haler device.
MATERIALS AND METHODS: Twelve healthy subjects, 12 asthmatics, and 12 COPD patients each received a single dose of 2 puffs 99mtechnetium-labelled FP/FORM 125/5 μg. A gamma camera was used to obtain anterior and posterior two-dimensional images of drug deposition. Prior transmission scans (using a99mtechnetium flood source) allowed the definition of regions of interest and calculation of attenuation correction factors. Image analysis was performed per standardised methods.
RESULTS: Of 36 subjects, 35 provided evaluable post-dose scintigraphic data. Mean subject ages were 35.7 (healthy), 44.5 (asthma) and 61.7 years (COPD); mean FEV1% predicted values were 109.8%, 77.4% and 43.2%, respectively. Mean pulmonary deposition was 26.6% (healthy), 44.7% (asthma), 39.0% (COPD) of the delivered dose. The respective mean penetration indices (peripheral:central ratio normalised to a transmission lung scan) were 0.44, 0.31 and 0.30.
CONCLUSION: FP/FORM administration via the K-haler device resulted in high lung deposition in patients with obstructive lung disease but somewhat lesser deposition in healthy subjects. Regional deposition data demonstrated drug deposition in both the central and peripheral regions in all subject populations. EUDRACT NUMBER: 2015-000744-42.
Copyright © 2018 The Authors. Published by Elsevier Ltd.. All rights reserved.

Entities:  

Keywords:  Deposition; Fluticasone propionate; Formoterol fumarate; Radiolabelled; Scintigraphy

Mesh:

Substances:

Year:  2018        PMID: 29724381     DOI: 10.1016/j.rmed.2018.03.029

Source DB:  PubMed          Journal:  Respir Med        ISSN: 0954-6111            Impact factor:   3.415


  4 in total

Review 1.  Lung Deposition and Inspiratory Flow Rate in Patients with Chronic Obstructive Pulmonary Disease Using Different Inhalation Devices: A Systematic Literature Review and Expert Opinion.

Authors:  Adolfo Baloira; Araceli Abad; Antonia Fuster; Juan Luis García Rivero; Patricia García-Sidro; Eduardo Márquez-Martín; Marta Palop; Néstor Soler; J L Velasco; Fernando González-Torralba
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2021-04-19

2.  Differential Performance and Lung Deposition of Levofloxacin with Different Nebulisers Used in Cystic Fibrosis.

Authors:  Carsten Schwarz; Claudio Procaccianti; Laura Costa; Riccardo Brini; Richard Friend; Grazia Caivano; Hosein Sadafi; Charles Mussche; Nicolas Schwenck; Michael Hahn; Xabier Murgia; Federico Bianco
Journal:  Int J Mol Sci       Date:  2022-08-24       Impact factor: 6.208

3.  A Generative Adversarial Network (GAN) Technique for Internet of Medical Things Data.

Authors:  Ivan Vaccari; Vanessa Orani; Alessia Paglialonga; Enrico Cambiaso; Maurizio Mongelli
Journal:  Sensors (Basel)       Date:  2021-05-27       Impact factor: 3.576

Review 4.  Asthma management with breath-triggered inhalers: innovation through design.

Authors:  Mário Morais-Almeida; Helena Pité; João Cardoso; Rui Costa; Carlos Robalo Cordeiro; Eurico Silva; Ana Todo-Bom; Cláudia Vicente; José Agostinho Marques
Journal:  Asthma Res Pract       Date:  2020-06-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.