Abstract

We report a case of guide wire fracture during percutaneous transluminal coronary angioplasty, using one of the newer catheter systems. The broken free end of the guide wire remained within the coronary tree, and surgical removal was necessary. Entrapment and overcoiling of the guide wire can cause fracture; in our case excessive bending in a tortuous coronary tree resulted in this unusual complication. An excessive tensile load to the guide wire may result from anatomical peculiarities and can cause wire fracture.

Entities:  

Mesh:

Year:  1988        PMID: 2972384     DOI: 10.1002/ccd.1810150208

Source DB:  PubMed          Journal:  Cathet Cardiovasc Diagn        ISSN: 0098-6569


  5 in total

1.  Surgical management of entrapped percutaneous transluminal coronary angioplasty hardware.

Authors:  Te-Ming Tommy Chang; Daniel Pellegrini; Alexey Ostrovsky; Albert G Marrangoni
Journal:  Tex Heart Inst J       Date:  2002

Review 2.  Intravascular foreign bodies: danger of unretrieved fragmented medical devices.

Authors:  Minori Tateishi; Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2009-06-18       Impact factor: 1.731

3.  Retained jailed wire: a case report and literature review.

Authors:  Masoud Pourmoghaddas; Omid Hashemi Fard
Journal:  ARYA Atheroscler       Date:  2011

Review 4.  Management of retained intervention guide-wire: a literature review.

Authors:  Abdulrahman M Al-Moghairi; Hussein S Al-Amri
Journal:  Curr Cardiol Rev       Date:  2013-08

5.  Broken guidewire - A tale of three cases.

Authors:  Goutam Datta
Journal:  Indian Heart J       Date:  2016-01-14
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.