Literature DB >> 29716359

A hollow coaxial cable Fabry-Pérot resonator for liquid dielectric constant measurement.

Chen Zhu1, Yiyang Zhuang1, Yizheng Chen1, Jie Huang1.   

Abstract

We report, for the first time, a low-cost and robust homemade hollow coaxial cable Fabry-Pérot resonator (HCC-FPR) for measuring liquid dielectric constant. In the HCC design, the traditional dielectric insulating layer is replaced by air. A metal disk is welded onto the end of the HCC serving as a highly reflective reflector, and an open cavity is engineered on the HCC. After the open cavity is filled with the liquid analyte (e.g., water), the air-liquid interface acts as a highly reflective reflector due to large impedance mismatch. As a result, an HCC-FPR is formed by the two highly reflective reflectors, i.e., the air-liquid interface and the metal disk. We measured the room temperature dielectric constant for ethanol/water mixtures with different concentrations using this homemade HCC-FPR. Monitoring the evaporation of ethanol in ethanol/water mixtures was also conducted to demonstrate the ability of the sensor for continuously monitoring the change in dielectric constant. The results revealed that the HCC-FPR could be a promising evaporation rate detection platform with high performance. Due to its great advantages, such as high robustness, simple configuration, and ease of fabrication, the novel HCC-FPR based liquid dielectric constant sensor is believed to be of high interest in various fields.

Entities:  

Year:  2018        PMID: 29716359     DOI: 10.1063/1.5021684

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  1 in total

1.  Microwave device inspired by fiber-optic extrinsic Fabry-Perot interferometer: a novel ultra-sensitive sensing platform.

Authors:  Chen Zhu; Rex E Gerald; Jie Huang
Journal:  J Lightwave Technol       Date:  2020-08-20       Impact factor: 4.142

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.