| Literature DB >> 29713778 |
Paul D Griffiths1, Deborah Jarvis2, Leanne Armstrong2, Daniel J A Connolly3, Pauline Bayliss3, Julie Cook3, Anthony R Hart4, Elizabeth Pilling3, Tamanna Williams3, Martyn N J Paley2.
Abstract
OBJECTIVES: MR imaging of neonates is difficult for many reasons and a major factor is safe transport to the MR facilities. In this article we describe the use of a small, investigational 3-T MR customised for brain imaging and sited on a neonatal unit of a tertiary centre in the UK, which is in contrast to a 300-m journey to the whole-body MR scanner used at present for clinical cases.Entities:
Keywords: Brain; Magnetic resonance angiography; Magnetic resonance imaging; Magnetic resonance spectroscopy; Neonate
Mesh:
Year: 2018 PMID: 29713778 PMCID: PMC6132701 DOI: 10.1007/s00330-018-5357-7
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1The GE Healthcare investigational 3-T neonatal MR system installed at Sheffield Teaching Hospital and a plan of the MR suite showing the proximity to the neonatal intensive care unit and delivery suite. The scanner control, scan room and equipment room occupy a floor space of 6.0 m × 6.5 m
Fig. 2Preparation of a baby for MR scanning on the GE Healthcare investigational 3-T neonatal MR scanner illustrated by a ‘mock’ case using a mannequin and research staff (see text for full details). The baby is changed into clothes with no ferro-magnetic components on the ward and an axillary temperature probe is attached (a). After transfer to the atrium in the MR suite the baby is re-checked for absence of ferro-magnetic items and earplugs, ear defenders and a vital sign monitor are applied (b). The baby is placed in a transport sling (c) and transferred on to the scan cradle of the tabletop, which has been detached from the MR scanner (d). The table and baby are pushed into the scanner room (e), the table is docked with a scanner and the cradle manually slid into the bore of the MR scanner (f). Close-up images of the scanner/table docking mechanism are shown in g to i
Summary of optimised MR sequence data used in the present study
| Sequences for vascular imaging | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| 2D T2 ssFSE | 3D FSPGR T1 Volume | 2D Propeller DWI | 2D GE | 2D T2 FSE | 3D SWAN (susceptibility-weighted imaging) | Single-voxel spectroscopy | 3D MRA (time of flight) | 3D MRV (phase contrast) | |
| Repetition time (TR) | 908 (min) | 10 | 6262 | 600 | 7000 | 57.5 | 1500 | 21 | 35 |
| Time to echo (TE) | 140 | Minimum full | 60 | 15 | 124 | 23 | 144 | 3.4 | 5 |
| Flip angle | 90 | 8 | 110 | 15 | 142 | 25 | 90 | 20 | 10 |
| Bandwidth (KHz) | 31.25 | 15.6 | 50 | 15.6 | 15.6 | 31.3 | 27.8 | 21 | |
| Prep time | - | 900 | - | - | - | - | - | - | - |
| Echo train length | - | 20 | - | 16 | 5 | - | - | - | |
| NEX (signal averages) | 0.55 | 0.75 | 2 | 2 | 2 | 0.7 | 1 | 0.63 | |
| Slice thickness/slice gap (mm) | 3/0 | 1.0/0 | 4/0 | 3.0/0.3 | 3/0 | 3/0 | 15 | 1/0 | 2.4/0 |
| Field of view (cm) | 16 × 14.4 | 18 × 12.6 | 16 × 16 | 16 × 12 | 16 × 14.4 | 16 × 11.2 | 1.5 × 1.5 | 20 × 14 | 20 × 13 |
| Freq/phase matrix | 256/192 | 256/192 | 128/128 | 156/192 | 256/192 | 256/192 | 384/224 | 320/192 | |
| Interpolated matrix and slice thickness | 512/384/0.5 | 512/384/1.5 | 256/384/1.5 | 384/224/0.5 | 320/384/1.2 | ||||
| b Value | - | - | 1000 | - | - | - | - | - | - |
| Scan time (s) | 29 | 264 | 192 | 180 | 231 | 173 | 140 | 240 | 364 |
| Additional | Flow comp. | Flow comp. | Flow comp. | VENQ = 10 | |||||
Fig. 3Representative MR images from a normal baby born at term showing the range of routine sequences acquired in cases 20-49 in this study
Fig. 4Representative images from a normal baby born at term showing MR arteriography and MR venography image sequences optimised in cases 13-18 in this study
Imaging quality assessments made by the paediatric neuroradiology experts
| Poor | Average | Good | Percentage of ‘average’ or ‘good’ cases | |
|---|---|---|---|---|
| Overall image quality | 5 | 13 | 34 | 90% |
| Image contrast | 4 | 8 | 40 | 92% |
| Presence of artefacts | 13 | 20 | 19 | 75% |
| Signal-to-noise ratio (SNR) | 7 | 7 | 38 | 87% |
| Tissue contrast | 2 | 9 | 41 | 96% |
| B0 inhomogeneity leading to fat/water separation | 0 | 22 | 30 | 100% |
Fig. 5Representative axial gradient echo T2* images from case A described in the text showing small areas of haemosiderin staining in the caudate-thalamic notches bilaterally consistent with previous haemorrhage into the remnants of the germinal matrix (arrowed, c)
Fig. 6Representative axial FSE T2* images from case B described in the text showing a small volume of intraventricular blood in the gravity dependant parts of the occipital horns of the lateral ventricles (arrowed, a) and a tubular structure running through the right frontal lobe and terminating in the superior sagittal sinus (block arrows, b-d) in keeping with a developmental venous anomaly