Lucie Grebíková1, Stuart G Whittington2, Julius G Vancso1. 1. Materials Science and Technology of Polymers, MESA+, Institute of Nanotechnology , University of Twente , P.O. Box 217, 7500 AE Enschede , The Netherlands. 2. Department of Chemistry , University of Toronto , 80 St. George Street , Toronto , Ontario M5S 3H6 , Canada.
Abstract
The adsorption-desorption behavior of polymer chains is at the heart of macromolecular surface science and technology. With the current developments in atomic force microscopy (AFM), it has now become possible to address the desorption problem from the perspective of a single macromolecule. Here, we report on desorption of single polymer chains on planar surfaces by AFM-based single molecule force spectroscopy (SMFS) as a function of the pulling angle with respect to the surface-normal direction. SMFS experiments were performed in water with various substrates using different polymers covalently attached to the AFM probe tip. End-grafting at the AFM tip was achieved by surface-initiated polymerization using initiator functionalized tips. We found that the desorption force increases with a decreasing pulling angle, i.e., an enhanced adhesion of the polymer chain was observed. The magnitude of the desorption force shows a weak angular dependence at pulling angles close to the surface normal. A significant increase of the force is observed at shallower pulling from a certain pulling angle. This behavior carries the signature of an adsorption-desorption transition. The angular dependence of the normalized desorption force exhibits a universal behavior. We compared and interpreted our results using theoretical predictions for single-chain adsorption-desorption transitions.
The adsorption-desorption behavior of polymer chains is at the heart of macromolecular surface science and technology. With the current developments in atomic force microscopy (AFM), it has now become possible to address the desorption problem from the perspective of a single macromolecule. Here, we report on desorption of single polymer chains on planar surfaces by AFM-based single molecule force spectroscopy (SMFS) as a function of the pulling angle with respect to the surface-normal direction. SMFS experiments were performed in water with various substrates using different polymers covalently attached to the AFM probe tip. End-grafting at the AFM tip was achieved by surface-initiated polymerization using initiator functionalized tips. We found that the desorption force increases with a decreasing pulling angle, i.e., an enhanced adhesion of the polymer chain was observed. The magnitude of the desorption force shows a weak angular dependence at pulling angles close to the surface normal. A significant increase of the force is observed at shallower pulling from a certain pulling angle. This behavior carries the signature of an adsorption-desorption transition. The angular dependence of the normalized desorption force exhibits a universal behavior. We compared and interpreted our results using theoretical predictions for single-chain adsorption-desorption transitions.
Polymer adhesion is
a complex phenomenon that has high relevance
in theoretical and experimental areas of research. Our ability to
delineate the properties that determine polymer adhesion at the molecular
level is crucial for the further development of innovative designs
of surgical adhesives[1] and polymer coatings
with antifouling properties,[2−4] biosensors,[5,6] low-friction
low-wear surfaces,[7] and polymer-based technological
glues.[1,8]Macromolecular adhesion and desorption
have been studied traditionally
at the ensemble level of thermodynamics. These studies average over
large populations of polymers. Therefore, individual characteristics
of single-chain adsorption are hidden. With the current developments
in atomic force microscopy (AFM) single molecule force spectroscopy
(SMFS) techniques, it has now become possible not only to study polymer
adhesion at the true molecular level but also to tackle directional
dependence of the desorption problem (i.e., failure of physical adhesion).
It has hitherto remained a grand challenge to address the adsorption/desorption
of single polymer chains as a function of the pulling angle with respect
to the plane of adsorption. Polymer adsorption–desorption has
been investigated at the substrate surface-normal direction while
the surface adsorbed chains were either picked up and stretched by
the tip probe or firmly attached to the tip. In both cases, the force
exerted on the cantilever was measured as a function of the distance
from the surface.[9−14] With the advances in the new generation of AFM instruments, it is
now possible to control the tip position and direction with high precision
with respect to the surface in three spatial directions, utilizing
independent feedback control in the three Cartesian directions. The
adsorbed polymer can then be pulled at different positions along the
backbone (for this experiment, previous single-chain imaging is required)
or also from the surface in different directions. However, modifications
to the software of commercial instrumentation are still required for
the latter experiment. To date, few studies have made use of this
option. In our approach,[15] the AFM tip
is retracted perpendicularly to the surface with a constant velocity
while the substrate is simultaneously moved in x and y directions. With precise control of the velocities, the
desired pulling angle can be achieved. Kühner et al.[16] performed experiments with polymers pinned to
the AFM tip where the cantilever was retracted to a certain distance
away from the surface, then moved parallel to the surface, with numerous
repetitions of these steps, and so forth. A similar approach was employed
in stretching one single DNA duplex repeatedly at different horizontal
locations by controlling the motion of the XYZ stage under the AFM
tip.[17] These studies focused on force–extension
measurements of individual polymers by SMFS in order to derive parameters
relevant to polymer adhesion, conformation and mechanics.Several
theoretical and computational studies[18−22] have addressed the polymer desorption from a substrate
by varying the angle of pulling forces with respect to the substrate
surface normal, in the range from substrate perpendicular to parallel
pull. These models predict that, depending on the angle, the polymer
favors either adsorption or desorption. To date, only some studies
attempted to vary force directions and pull polymer chains laterally
over the substrate to enhance polymer adsorption or to study single-polymer
friction.[7,16,23,24] Thus, designed experiments to probe how the polymer
behaves in the intermediate region remain essentially unexplored.
Our goal in this study is to gain a microscopic understanding of the
hitherto experimentally unaddressed adsorption–desorption transitions,
i.e., how a polymer responds to an external force applied at various
angles.A theoretical model for pulling an adsorbed polymer
at an angle
has been considered by Orlandini and Whittington[19] and by Osborn and Prellberg.[21] Although the theoretical approaches used in the two papers are different,
the underlying model and the results are identical. The model used
is a partially directed walk on the simple cubic lattice. Consider
the simple cubic lattice and attach the obvious coordinate system
(x, y, z), so that
each vertex of the lattice has integer coordinates. A partially directed
walk can take steps in the positive x-direction,
in the positive y-direction, and in both the positive
and negative z-directions, subjected to the walk
not revisiting any already visited vertices of the lattice. The walk
starts at the origin and is confined to the half-space z ≥ 0, so that the plane z = 0 represents
an impenetrable surface at which adsorption can occur. The walk receives
weights for each edge in z = 0 and is subjected to
force at the vertex of degree 1 that is not fixed at the origin. The
model can be solved completely using generating function techniques
and the thermodynamics (in the infinite walk limit) can be extracted
from the singularity structure of the generating function. The principle
result is the angular dependence of the critical force required for
desorption as a function of a parameter, T, that
plays the role of a reduced temperature.In this paper, we chose
two polymer systems that differ in hydrophobicity
and their conformational states in water (used as medium), namely,
poly(2-hydroxyethyl methacrylate) (PHEMA) and poly(methyl methacrylate)
(PMMA). Detailed pulling angle dependence of the adsorption was studied
for PHEMA on various substrates including mica, silica and highly
ordered pyrolytic graphite (HOPG). PHEMA and PMMA were synthesized
and end-grafted from an AFM tip via surface-initiated, quasi-living
atom transfer radical polymerization (SI-ATRP).[25] This approach allowed us to obtain end-grafted chains fastened
to the tip with controlled chain length.
Results and Discussion
The adhesion of PHEMA and PMMA end-grafted from the AFM tip was
investigated by SMFS in aqueous solution on three different substrates,
namely, mica, silica, and HOPG. PHEMA and PMMA were grown from initiator
functionalized AFM tips by ATRP with low enough grafting density to
ensure the presence of individual polymer chains at relatively large
distances, thus facilitating single-chain contact with the substrate.
In the first series of experiments, the polymer chain was desorbed
by pulling the polymer decorated AFM tip away perpendicularly from
the surface. Subsequently, the angle-dependent single polymer chain
adhesion experiments were carried out where the force response of
the polymer chain was measured while it was pulled from the surface
along different directions (Figure ).
Figure 1
Schematic diagram of the experiment. The polymer is end-grafted
to the AFM tip, adsorbed on the substrate first, and then pulled (under
various angles) with respect to the surface. The desorption force F is geometrically defined by the measured F and the applied pulling angle θ. D and D represent
the desorption length and the measured length in the z direction,
respectively. Angle-dependent measurements are performed by controlling
the polar angle θ, not considering the azimuthal angle φ.
A, Last adhering segment in the scenario depicted; B, chain end.
Schematic diagram of the experiment. The polymer is end-grafted
to the AFM tip, adsorbed on the substrate first, and then pulled (under
various angles) with respect to the surface. The desorption force F is geometrically defined by the measured F and the applied pulling angle θ. D and D represent
the desorption length and the measured length in the z direction,
respectively. Angle-dependent measurements are performed by controlling
the polar angle θ, not considering the azimuthal angle φ.
A, Last adhering segment in the scenario depicted; B, chain end.
Pulling in the Surface-Normal Direction
Typical experimental
results for end-grafted PHEMA and PMMA are shown in Figure . In the first series of experiments,
a series of approach–retraction force profiles was recorded
as the polymer was pulled in the surface-normal direction. The force
profiles varied from one approach–retraction cycle to another,
and single molecule events were observed in approximately 20% of the
force spectroscopy experiments for PHEMA and in less than 1% for PMMA.
PMMA shows a low probability of events compared to PHEMA, which results
from the hydrophobicity and tight precipitated coil conformation of
the polymer chain in water and weak interaction with the substrates.
The rest of the force curves show no single-polymer events at all
(unspecific adhesion force signature could only be observed). When
the polymer was bridging the tip and the substrate, one could distinguish
two different types of events: the successive unbinding of individual
segments, all bound to the surface in the so-called “train”
conformation, until the last segment (chain end; B in Figure ) becomes desorbed, and continuous
increase in the restoring force with distance (i.e., polymer elongation).
In most cases, the force profiles on retraction exhibited plateaus
of constant force with increasing extension, as shown in Figure (top left). Such
desorption behavior has been observed and is well established in previous
work.[4,9,11] Here, the
plateaus of constant force reflect the desorption of successive chain
segments, whereas the complete desorption of the entire polymer chain
results in a sudden drop of the adhesive force. The desorption force
is independent of the chain length as would be expected for the “train”
conformation of the molecule on the surface. We confine ourselves
in this analysis to “train-like” adsorption.
Figure 2
Desorption
of end-grafted polymers from different substrates in
ultrapure water. Typical retraction force curves for (a) PHEMA and
(b) PMMA, as captured for the polymer desorbing in the surface-normal
direction. Histograms of the desorption forces for (c) PHEMA and (d)
PMMA recorded on mica, silica, and HOPG, from top to bottom. The vertical
axes correspond to the number of counts, normalized to the maximum.
The black lines show the Gaussian fit of the desorption forces.
Desorption
of end-grafted polymers from different substrates in
ultrapure water. Typical retraction force curves for (a) PHEMA and
(b) PMMA, as captured for the polymer desorbing in the surface-normal
direction. Histograms of the desorption forces for (c) PHEMA and (d)
PMMA recorded on mica, silica, and HOPG, from top to bottom. The vertical
axes correspond to the number of counts, normalized to the maximum.
The black lines show the Gaussian fit of the desorption forces.The measurements are performed
in thermodynamic quasi-equilibrium,
meaning that the bonds involved in the adsorption process dissociate
and reassociate on a much faster time scale than the experimental
desorption process.[26] This assumption is
supported experimentally by the finding that the desorption forces
are independent of the bond loading rate used in the experiments,
as we reported elsewhere.[15] In the case
of PHEMA, only a very small percentage of the total force events (<0.2%)
exhibited sharp sawtooth-like patterns in the force curves reflecting
the stretching of sections of the individual polymer chains. In this
case, the polymer-surface bonds were stable and relaxed slowly (the
chain was pinned, and one could observe a nonlinear attractive peak
in the retraction curve). The polymer stretching was much more likely
for PMMA. We shall not address stretching of isolated single chains
in this study; stretching of this type has been well established[10,12,27,28] in the literature. Please note that because of the “unspecific”
interaction of the silicon nitride tips with the substrate, force–extension
curves occasionally show a high unspecific adhesion peak at short
tip–substrate distances. However, in our measurements, the
polymer chains exhibit sufficiently large contour lengths so that
the unspecific adhesion peak does not affect the measured distributions
of the desorption force. An example of the scatter plot of desorption
force versus contour length for PHEMA on silica can be found in the Supporting Information (Figure S1).The
difference in the desorption forces of end-grafted PHEMA and
PMMA caused by the interaction with various substrates was investigated
on mica, silica, and HOPG. To this end, for each force curve obtained,
we measured the magnitude of the average desorption force from individual
plateaus. Histograms of the desorption forces in Figure c,d show that the distribution
of the desorption forces is approximately Gaussian. (The fit parameters
are shown in the Supporting Information.) The desorption force represents a direct measure of the interaction
between the polymer and the surface in the solvent in question. The
nature of the polymer and the surface substantially influences the
magnitude of the desorption force, as can clearly be seen in Figure . The values of the
average desorption forces and the corresponding standard deviations
are summarized in Figure . The desorption force increases with increasing hydrophobicity
of the substrate, i.e., with increasing adsorption strength. This
trend is more pronounced for PHEMA and is very weak for PMMA. Depending
on the substrate in question, the desorption forces lie in the range
of 55–100 pN for PHEMA and 48–60 pN for PMMA (Table ). These findings
compare well to the values reported for PHEMA previously[29] and the desorption of other synthetic polymers
or single stranded DNA.[11,13,30−32] PHEMA and PMMA are neutral polymers composed of a
hydrophobic backbone, having a hydrophilic ester group in their repeating
unit, carrying a dipole due to the carbonyl group and featuring different
side groups. The free hydroxyl groups of PHEMA make the polymer hydrophilic
and lead to a relatively high degree of hydration.[33] The side groups of PMMA are composed of only short hydrophobic
methyl groups that make the chain hydrophobic.
Figure 3
Desorption of end-grafted
PHEMA and PMMA from different substrates
in ultrapure water. Average desorption forces for PHEMA and PMMA are
shown, recorded for mica, silica and HOPG with the corresponding standard
deviations.
Table 1
Statistical
Analysis of the Adhesion
Data of PHEMA and PMMA on Various Surfaces with the Corresponding
Standard Deviations
Desorption
Force (pN)
Adsorption
Energy per Monomer (kT)
Surface
Water Contact Angle (deg)
PHEMA
PMMA
PHEMA
PMMA
Mica
≤5
55.7 ± 3.0
48.4 ± 1.0
3.7
3.2
Silica
18 ± 5
67.8 ± 5.5
54.2 ± 2.1
4.5
3.6
HOPG
71 ± 5
97.3 ± 7.0
58.9 ± 5.4
6.4
3.9
Desorption of end-grafted
PHEMA and PMMA from different substrates
in ultrapure water. Average desorption forces for PHEMA and PMMA are
shown, recorded for mica, silica and HOPG with the corresponding standard
deviations.The PHEMA chains swell
in aqueous solution with a hydration layer
formed via hydrogen bonding with water molecules.[34] The strong increasing trend observed for PHEMA adhesion
can be explained by hydrophobic interactions and a high van der Waals
attractive interaction between the hydrophobic backbone of the polymer
and the substrate. An adsorption energy per monomer ranging from 3.7
to 6.7 kT is observed, where kT refers
to the thermal energy (see Table ). Similar adsorption energies have been reported for
synthetic polymers and polyelectrolytes.[35] Lower adsorption energies obtained for more hydrophilic surfaces
can be explained by the adsorption resistance at the hydrophilic surface
that results from strongly bound interfacial water due to water–surface
hydrogen bonding, suppressing direct polymer–surface interactions.[4] PMMA exhibits only a weak variation in the adsorption
energy on different substrates in aqueous solution. Unlike PHEMA,
water is a nonsolvent for PMMA, so the surface-tethered chains are
in a collapsed globule state, excluding the solvent.[36] In water, the PMMA chains form self-assembled domains in
a pearl-necklace conformation.[37] We observed
an increased adhesion on more hydrophobic substrates, i.e., HOPG,
with a high frequency of stretching events that can be rationalized
by interaction forces of a hydrophobic character.Adhesion energies
obtained experimentally were compared to the
theoretical values for the polymer interacting with various substrates
across a medium using the Lifshitz theory.[38] The calculation procedure is described in the Supporting Information. The trend and variation of the magnitude
of the calculated values of adhesion energies for various substrates
(as presented in Table ) corroborate the measured adsorption energies per segment (Table ), i.e., the adhesion
increases in the order of mica–silica–HOPG. The absolute
values differ as the calculations neglect contributions from other
short-range forces.
Table 2
Theoretical Values
for Macroscopic
Phases 1 and 2 Interacting Across a Medium 3
Phase
Dielectric Constant ε[38,39]
Refractive Index n(38,40,41)
Hamaker Constant[38]A132 (10–20 J)
Adhesion Energy W[38] (kT)
Mica
4
1.56
0.9
1.5
Silica
11
1.45
1.6
2.6
HOPG
10000
1.5
2.8
4.5
Water
80
1.33
Polymer
2.6
1.49
In the desorption experiments,
we occasionally observed plateaus
of constant force with several steps in the force profiles. Such “stapled”
plateaus can reflect desorption involving loops within individual
polymer molecules or contributions from several polymer chains pulled
off the substrate. In such cases, molecules of different lengths are
simultaneously adsorbed on the surface, and the steps represent their
desorption, one after the other, until the last polymer is fully detached.
The last step then reflects the desorption of a single polymer from
the substrate (see Figure S2 Supporting Information). The probability of multiple steps in the desorption force varied
with the substrate. The steps were observed in 1.2% of all cases for
mica, in 7% for silica, and in 88.1% for HOPG. This trend can be rationalized
by different polymer-surface interactions and possible loop formation
with increased hydrophobicity of the substrate.
Angle-Dependent
Pulling
We investigated angle-dependent
adhesion of tip end-grafted PHEMA while desorbed from various substrates
at different pulling angles in ultrapure water by SMFS. We aimed to
measure the magnitude and control the direction of forces acting on
the polymers using a custom-written software, in combination with
a commercial AFM instrument, as reported in detail elsewhere.[15] The experimental strategy is illustrated in Figure . The AFM tip with
end-grafted polymer approached the surface, the polymer chain was
adsorbed on the surface and subsequently pulled away at different
angles with respect to the surface. In our experiments, we only varied
polar angles, as the chain orientation on the surface was not known,
and the variation in the azimuthal angles could therefore not be controlled
and is not expected to provide meaningful information. Furthermore,
by considering the FJC model,[42] the desorption
of the polymer segment by segment with a constant force suggests that
only the orientation of two neighboring points at the time is relevant,
and the remaining segments of the polymer chain are attached to the
surface. During the experiments, the force measured through the deflection
of the AFM cantilever, F, represents the vertical component of the force applied to the polymer
molecule, F. The desorption force acting along the
contour of the polymer was then determined from the measured F (where F = F sin θ) using
the pulling geometry as given in Figure . Similarly, the measured extension, D, is only the projection of
the distance between the two anchor points.The other components
of the force vector (e.g., lateral) can cause deflection of the AFM
cantilever (twisting the cantilever). Several mechanisms will contribute
to the measured desorption force: adhesion, friction, and elasticity
of the polymer as well. We have recorded the lateral deflection signal
during the angle-dependent experiments to verify if there is a measurable
lateral torsion of the cantilever. The lateral deflection signal did
not display any detectable change while desorbing a polymer chain
at various angles.[15]Figure a shows
typical force profiles of PHEMA obtained in angle-dependent adhesion
measurements upon retraction at various pulling angles on silica in
ultrapure water. The pulling angles were varied in the range between
30° and 90°. The z-component of the desorption length decreases
with a decreasing pulling angle as the polymer is continuously desorbed
from the surface (Figure b). These findings correlate well with the geometrical prediction
of the experiment. While desorbing a polymer chain, we consider a
mechanism where a polymer chain can be peeled from the surface without
changing the binding sites of the monomers. Within the FJC model we
assume that only two neighboring connected segments are relevant for
polymer pulling. However, due to variable surface conformations and
the change of force attack at shallow angles, an error between contour
length and experimentally measured length arises (Figure b). If we pull a chain perpendicular
to the substrate and if no displacement has taken place, then the
desorption (contour) length Z can be estimated as , where L is contour length
and Ree end-to-end distance. The deviation
between contour length and pulling length would be approximately ≈1.5%.
We neglect this error and assume that the pull length (Z travel) for perpendicular pull is the same as the contour length. Figure b shows that the
deviation is most noticeable for the shallowest pulling angle. We
have estimated the deviation between contour length and experimentally
measured length for 30° and 90° pull to be approximately
15%, taking into account similar polymer chains[43] imaged on the same substrate (estimated deviation here
13%), end-to-end distance, contour length and experimental geometry.
Details about the error estimation are provided in the Supporting Information.
Figure 4
Angle-dependent desorption
of end-grafted PHEMA from silica in
ultrapure water. (a) Force profiles upon retraction exhibiting plateaus
of constant force observed for pulling angles in the range between
30° amd 90°. (b) Average desorption length plotted as a
function of pulling angle with corresponding standard deviations.
Angle-dependent desorption
of end-grafted PHEMA from silica in
ultrapure water. (a) Force profiles upon retraction exhibiting plateaus
of constant force observed for pulling angles in the range between
30° amd 90°. (b) Average desorption length plotted as a
function of pulling angle with corresponding standard deviations.We state that over the pulling
experimental time scales chain lateral
diffusion can be neglected. We base this statement on the fact that
single chains can be visualized in a given conformation at substrates
by tapping mode imaging[13,43] over a time scale of
several minutes. A single pulling measurement in our case takes typically
4 s. Obviously, if nonmoving chains can be imaged over several minutes,
then we can assume that lateral diffusion would be negligible within
the time scale of our experiment. We can also rule out the option
that the XY offset of the actually desorbing segment closely follows
the tip motion, i.e., in this case the pulling length would be equal
to the contour length. This cannot be justified by the data (Figure b). Qualitatively,
the same findings have been observed for the angle-dependent desorption
of PHEMA from other substrates, as well. Thus, we are left with two
options: (a) no slide, (b) some slide but variation at XY offset closely
following the tip motion is not fulfilled.The average desorption
forces of end-grafted PHEMA as a function
of the pulling angle measured on various substrates in ultrapure water
are summarized in Figure . All surfaces exhibit the same behavior; the desorption force
increases monotonically as the pulling angle decreases. The distribution
of the average desorption forces for the individual pulling angles
was relatively narrow and approximately Gaussian (example shown in Supporting Information for HOPG, Figure S3).
Desorption measurements at a particular pulling angle represent a
statistical result arising from many possible chain conformations
on the substrate but our finding supports the assumption that the
orientation of the chain on the surface (i.e., variation in the azimuthal
angle) has a negligible effect on the polymer desorption on homogeneous,
unstructured substrates. We note that the change in the azimuthal
angle might play a significant role on polymer desorption from heterogeneous,
structured (“stripped”) or anisotropic substrates. The
importance of pulling angle for such surfaces has been demonstrated
theoretically.[22]
Figure 5
Angle-dependent adhesion
of various systems. The average desorption
forces with corresponding standard deviations recorded for different
pulling angles on HOPG in ultrapure water compared to (a) end-grafted
PHEMA desorbed from mica, (b) end-grafted PHEMA desorbed from silica,
and (c) end-grafted PHEMA and PMMA. The dashed lines serve only to
guide the eye. The desorption force acting along the contour of the
polymer F was then determined from F (where F = F sin θ) using the pulling
geometry as given in Figure .
Angle-dependent adhesion
of various systems. The average desorption
forces with corresponding standard deviations recorded for different
pulling angles on HOPG in ultrapure water compared to (a) end-grafted
PHEMA desorbed from mica, (b) end-grafted PHEMA desorbed from silica,
and (c) end-grafted PHEMA and PMMA. The dashed lines serve only to
guide the eye. The desorption force acting along the contour of the
polymer F was then determined from F (where F = F sin θ) using the pulling
geometry as given in Figure .The magnitude of the desorption
force depends on the strength of
the polymer–surface interactions in question; for all surfaces,
decreasing the pulling angle resulted in enhanced adhesion of the
polymer chain, as predicted theoretically.[18,21,24] As shown in Figure a, we display in detail the angle-dependent
desorption of PHEMA on two substrates, namely, on mica and HOPG. The
average desorption force is approximately constant at higher pulling
angles, but by decreasing the angle beyond 50°, the desorption
force abruptly increases. This we interpret as the observation of
the fingerprints of the theoretically predicted adsorption–desorption
transition with varying pulling angles. The theoretical predictions[19,21] rely on partially directed walk models in three dimensions of a
polymer tethered at one end to a surface and being pulled at fixed
angles away from the point of tethering. At a given temperature, pulling
angle variations lead to changes in the interaction force. The transition
angle would not change when the interaction energy is varied; however,
the temperature dependence of the force magnitude would vary. The
lattice model for polymer adhesion considered by Orlandini and Whittington,[19] and by Osborn and Prellberg,[21] predicts that at first, for pulling directions close to
normal to the surface, there is only a small dependence on pulling
angle. However, the dependence becomes more marked as the angle decreases.
At very large forces (and so at very small angles where the required
force is large), there is a problem with the model. One would expect
that the walk would eventually follow the pulling direction but this
is not entirely possible for a lattice model. This means that the
model overestimates the force required for desorption at small angles.
This only becomes a major issue at angles less than about 30°
and, at larger angles, there is very good qualitative agreement between
theory and experiment. As Figure a,b shows, the transition angle does not change with
the variation of the substrate; only the magnitude of the measured
force varies as the result of the difference in the interaction energy
of PHEMA with the various substrates.Figure c shows
the average desorption force vs pulling angle for PHEMA and PMMA.
For the two polymers, we observed qualitatively the same trend, i.e.,
an enhancement of the polymer adhesion with decreasing pulling angle.
The magnitude of the force varies because of different polymer–surface
interactions.Figure offers
a semiquantitative comparison between the theory and the experiments.
The shape of the force dependence on the pulling angle is the same,
with only a scaling factor difference. The theory is based on a partially
directed walk on a lattice and there is no “chemistry”
in the model, i.e., all surfaces and all polymers are treated as being
identical, as described in the Introduction. However, it is interesting to note that the experimental data all
more or less collapsed to a single curve when normalized by F(90°). This suggests that the angular term is largely
independent of the system, showing a universal behavior. Theoretical
force values were determined from the critical force-temperature curves
for various values of the angle at which the force is applied. We
have compared values obtained for the temperature range of 2.4–2.8,
i.e., below the critical adsorption temperature of about 3.03 (the
temperature at which desorption occurs without a force).[21] At a particular temperature T, the force required for the polymer desorption changes, depending
strongly on T at each angle. However, the ratio of F(θ)/F(90°) does not depend
much on the temperature. Dividing by the force at 90° value significantly
removes the temperature dependence. (We note that all our studies
were performed at isothermal conditions at room temperature.)
Figure 6
Desorption
force at pulling angle θ normalized by force at
90° as a function of the pulling angle. Experimental data obtained
for the desorption of PHEMA and PMMA on various substrates are compared
to theoretical critical force curves versus pulling angle for given
range of temperatures as based on partially directed walk models of
adsorbed polymers.[19]
Desorption
force at pulling angle θ normalized by force at
90° as a function of the pulling angle. Experimental data obtained
for the desorption of PHEMA and PMMA on various substrates are compared
to theoretical critical force curves versus pulling angle for given
range of temperatures as based on partially directed walk models of
adsorbed polymers.[19]
Conclusion
In conclusion, we addressed the single polymer
chain response to
external force applied at various angles in aqueous solution. The
experiments performed with different polymers (PHEMA and PMMA) and
on various substrates demonstrated an enhancement of the polymer adhesion
while decreasing the pulling angle, in full agreement with theoretical
predictions.[19,21] A clear trend for the adsorption–desorption
transition angle was found experimentally by performing direction-dependent
force spectroscopy experiments. The dependence of the desorption force
on the pulling angle significantly increased at smaller angles and
showed with decreasing angle the adsorption–desorption transition,
in qualitative agreement with theoretical predictions.[19,21] The angular dependence of the normalized desorption force exhibits
a universal behavior. A semiquantitative comparison between the theory
and the experiments has been offered. The force dependence on the
pulling angle shows the same shape, with only a scaling factor difference.
Experimental Section
Materials
3-Aminopropyl
trimethoxysilane (APTMS, 97%),
triethylamine (TEA, 99.5%), copper(II) bromide (CuBr2,
99%), 2,2′-bipyridine (bipy, 98%), 2-bromoisobutyryl bromide
(BIBB, 98%), 1-bromocarbonyl-1-methylethyl acetate (BMA, 96%), N,N,N′,N′,N′-pentamethyldiethylenetriamine
(PMDETA, 98%), and l-ascorbic acid (LAA) were purchased from
Sigma-Aldrich and were used as received. Toluene (AR) and ethanol
(absolute) were purchased from Biosolve and Merck, respectively. All
water used in the experiments was ultrapure (Millipore Milli-Q grade).
2-Hydroxyethyl methacrylate (HEMA, 97%) and methyl methacrylate (MMA,
99%) were purchased from Sigma-Aldrich and purified from inhibitors
by passing through a basic alumina column. Copper(I) chloride (CuCl,
98%) was purchased from Sigma-Aldrich, purified by stirring in glacial
acetic acid for several hours, filtered, washed with absolute ethanol,
and vacuum-dried. The native oxidized silicon wafer surfaces were
purchased from Okmetic (Vantaa, Finland). The silicon cantilevers
were obtained from Nanosensors (Neuchâtel, Switzerland).
Synthesis of Polymer Chains by SI-ATRP
The synthesis
of the PHEMA and PMMA polymer chains was achieved through classical
ATRP and activators regenerated by electron transfer (ARGET) ATRP,
respectively. Polymer brushes and single polymer chains were end-grafted
from reference substrates of planar silicon and AFM tip, respectively.
The SI-ATRP of PHEMA and PMMA on silicon surfaces and AFM tips was
performed simultaneously in the same batch. Each polymerization step
was verified on the planar substrates. Briefly, the synthesis consisted
of four steps: the pretreatment/cleaning of the surfaces, the amino
functionalization via silanization of the substrates, the coupling
with active and inactive ATRP-initiators via an acylation reaction
and the controlled radical polymerization (CRP) of the monomers. The
native oxidized silicon wafer surfaces were cleaned with piranha solution,
and the cantilevers were plasma cleaned with an SPI Plasma Prep II
instrument (SPI Supplies, West Chester, USA) in an oxygen-enriched
atmosphere prior to use. Subsequently, the planar reference surfaces
and cantilevers were silanized for 12 h in the gas phase in an evacuated
container in the presence of APTMS and washed with toluene and ethanol.
To control the polymer grafting density, the cantilevers with amino-functionalities
were coupled with active and inactive ATRP initiator precursors: BIBB
and 1-bromocarbonyl-1-methylethyl acetate (BMA), respectively. The
cantilevers were immersed in toluene with TEA, to which BIBB and BMA
were added in the desired molar ratio. The surfaces were thoroughly
washed with toluene, ethanol, and water and dried under the stream
of nitrogen. PHEMA polymer chains were synthesized via SI-ATRP, while
the synthesis of PMMA polymer chains was achieved via SI-ARGET-ATRP.
The PHEMA and the PMMA-modified samples were washed with water then
ethanol and dried in a stream of nitrogen. Further details on the
preparation of the surfaces can be found in the Supporting Information.
Film Characterization
Surface functionalization and
the resulting polymer films of the reference planar silicon substrates
were characterized by FTIR spectroscopy, contact angle measurements,
and AFM imaging to ensure that each step of the polymerization of
HEMA and MMA on the AFM cantilever was successful. FTIR spectra were
obtained using a Vertex 70v spectrometer (Bruker) with a nitrogen-cooled
cryogenic detector. Background spectra were obtained using a cleaned
silicon oxide substrate. Contact angle measurements were performed
with an optical contact angle device (OCA 15, Dataphysics, Germany)
equipped with an electronic syringe unit and connected to a charge-coupled
device (CCD) camera. Values of the static contact angles of ultrapure
water were measured using the sessile drop mode. The dry thickness
of the polymer brush on the planar substrate was obtained by measuring
the step height of scratches gently made on the silicon. A Dimension
D3100 AFM equipped with a hybrid scanner and a NanoScope IVa controller
(Digital Instruments, Veeco-Bruker, Santa Barbara, CA) were employed,
operated in intermittent-contact mode in air using cantilevers (Nanoworld,
Neuchâtel, Switzerland) with a force constant of approximately
42 N/m and a resonance frequency of 320 kHz at a scan rate of 1 Hz.
Angle-Dependent SFMS
SMFS measurements were carried
out in ultrapure water (Millipore Milli-Q grade, pH 7) in contact
mode with a commercial AFM instrument Molecular Force Probe 3D (MFP-3D)
(Asylum Research, Santa Barbara, CA). The cantilevers (PointProbe
Plus Contact Mode, Nanosensors, Neuchâtel, Switzerland) with
the end-grafted polymer chain, prepared as described above, had a
nominal tip radius smaller than 7 nm and a resonance frequency of
10–15 kHz in water. The SMFS experiments were performed with
different polymers (PHEMA and PMMA) on various substrates, namely,
mica, silica, and highly ordered pyrolytic graphite (HOPG). High-grade
mica and HOPG were obtained from Plano (Wetzlar, Germany) and cleaved
with adhesive tape in air prior to use. The native oxidized silicon
wafer surfaces were plasma cleaned with an SPI Plasma Prep II instrument
(SPI Supplies, West Chester, USA) in an oxygen-enriched atmosphere
for 20 min prior to use.In SFMS, a series of several thousand
approach–retraction force curves was recorded with a sampling
rate of 2 kHz, whereby the deflection of the cantilever and the vertical
piezo displacement were acquired as a function of time. After the
surface was contacted, the AFM tip with end-grafted polymer was retracted
under different pulling angles relative to the substrate, as shown
in Figure . The pulling
angle control was programmed based on Igor Pro and Asylums Research’s
built-in functions. In detail, the tip was retracted perpendicularly
to the surface with a constant velocity while the substrate was simultaneously
moved in the x and y directions. By precisely controlling the velocities
in the Cartesian directions, the desired pulling angle was set. The
experimental strategy for the directional force experiments is addressed
in more detail elsewhere.[15] To preserve
the functionalized tip, the maximum applied force upon contact with
the surface was kept smaller than 4 nN. The cantilever spring constants
were in the range of 0.12–0.28 N/m as measured by thermal fluctuations
in air.[44] The retraction velocity used
was 250 nm/s. Retraction force curves were very similar when the velocity
was varied in the range of 25–18 000 nm/s, and the recorded
desorption force was the same for all velocities within the experimental
error (data not shown). Changing the dwell time toward the surface
did not have any effect on the recorded force response. All AFM experiments
were carried out at room temperature of 20 ± 1 °C. The resulting
single molecule force profiles were analyzed with the program Igor
Pro (Wavemetrics) with custom procedures. Plateau force and length
were extracted for each force curve. The values of each measurement
were plotted in a histogram and fitted with Gaussian distributions.
The maximum of the Gaussian functions was extracted and the standard
deviation was plotted as error.
Authors: Xiaoying Zhu; Shifeng Guo; Dominik Jańczewski; Fernando Jose Parra Velandia; Serena Lay-Ming Teo; G Julius Vancso Journal: Langmuir Date: 2013-12-19 Impact factor: 3.882
Authors: Ferdinand Kühner; Matthias Erdmann; Lars Sonnenberg; Andreas Serr; Julia Morfill; Hermann E Gaub Journal: Langmuir Date: 2006-12-19 Impact factor: 3.882
Authors: Suresh Manohar; Amber R Mantz; Kevin E Bancroft; Chung-Yuen Hui; Anand Jagota; Dmitri V Vezenov Journal: Nano Lett Date: 2008-12 Impact factor: 11.189
Authors: Wanhao Cai; Jakob T Bullerjahn; Max Lallemang; Klaus Kroy; Bizan N Balzer; Thorsten Hugel Journal: Chem Sci Date: 2022-04-26 Impact factor: 9.969