Literature DB >> 2971235

Guide wire angioplasty balloon catheter: preliminary report.

C J Tegtmeyer1.   

Abstract

A new balloon catheter has been developed for angioplasty. The balloon is more flexible and its walls are thinner than those of conventional catheters. It folds down better yet withstands 8-10 atm of pressure. This advance has been married to new guide wire technology, producing a low-profile angioplasty balloon on a 0.038-inch guide wire. It compared favorably with a standard 7-F angioplasty catheter when tested at 4 atm in a rabbit and has been successfully used to dilate ten peripheral stenoses in four patients.

Entities:  

Mesh:

Year:  1988        PMID: 2971235     DOI: 10.1148/radiology.169.1.2971235

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105